  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63616",
      "recalling_firm": "Mindray DS USA, Inc. d.b.a. Mindray North America",
      "address_1": "800 MacArthur Blvd.",
      "address_2": "N/A",
      "postal_code": "07430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including OH and GA",
      "recall_number": "Z-0630-2013",
      "product_description": "DPM Central Station  Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430    The Central Monitoring Station (CMS) network transfers information between DOM Central Station and other network devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and or WLAN connections. The DPM Central Station is used for remote monitor management, storing, printing, reviewing, or processing of information from net worked devices, and it is operated by medical personnel in hospitals or medical institutions.",
      "product_quantity": "two units",
      "reason_for_recall": "There is a software anomaly with the DPM Central Monitoring System where the trend data of one patient maybe replaced with the trend data of a second patient.",
      "recall_initiation_date": "20110512",
      "center_classification_date": "20130103",
      "termination_date": "20130219",
      "report_date": "20130109",
      "code_info": "510 K K080192  P/N 300BF-PA200002"
    }
  ]
}