  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86896",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US nationwide, ARGENTINA, AUSTRALIA, Brazil,  CANADA, CHILE, CHINA, COSTA RICA, EMEA, HONG KONG, INDIA, JAPAN, KOREA, MALAYSIA, NEW ZEALAND, SINGAPORE, and TAIWAN",
      "recall_number": "Z-0627-2021",
      "product_description": "RINGLOC ACETABULAR SYSTEM Self-tapping bone screws, Item numbers 103531, 103533, 103534, 103535 - Product Usage: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented; Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented; Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component); Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer.",
      "product_quantity": "7428",
      "reason_for_recall": "Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.",
      "recall_initiation_date": "20201125",
      "center_classification_date": "20201217",
      "termination_date": "20240508",
      "report_date": "20201223",
      "code_info": "Lots 259830  259860  259880  259890  259910  259920  260020  260140  260160"
    }
  ]
}