      "country": "United States",
      "city": "Marquette",
      "address_1": "375 River Park Cir",
      "reason_for_recall": "As the result of a recent internal review of regulatory documents, we have determined that the 10 mm x 12 mm Webless C-Plus implants were incorrectly documented as meeting the requirement for labeling as a partial Vertebral Body Replacement Device (VBR).  While the use of this device as a partial VBR is unlikely due to the size of the device, use of this device as a partial VBR could result in a s",
      "address_2": "",
      "product_quantity": "1906",
      "code_info": "part numbers: 30-CW-1012-5, 30-CW-1012-6, 30-CW-1012-7, 30-CW-1012-8, 30-CW-1012-9, 30-CW-1 012-10, 30-CW-1 012-11.",
      "center_classification_date": "20140103",
      "distribution_pattern": "Nationwide Distribution - USA including FL, MO, NY, CA, MI,and IL.",
      "state": "MI",
      "product_description": "C-Plus\" 10mm x 12mm Webless implants    Pioneer Surgical's C-Plus \" 10mm x 12mm Webless implants are cervical intervertebral fusion devices that are intended to stabilize a cervical spinal segment to promote fusion in order to restrict motion and decrease pain. They are intended for use with supplemental fixation and autograft. These implants are manufactured from PEEK (Polyaryletheretherketone) polymer with tantalum marker pins for radiographic visualization.",
      "report_date": "20140115",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "RTI Surgical, Inc.",
      "recall_number": "Z-0627-2014",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "67093",
      "termination_date": "20140226",
      "more_code_info": "",
      "recall_initiation_date": "20131007",
      "postal_code": "49855-1781",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}