  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "North Haven",
      "address_1": "60 Middletown Ave",
      "reason_for_recall": "The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging  from the needle shaft post-ablation. This can result in the tip remaining in the patient.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "GTIN Number - 10884521706583  Lot Numbers-    S7HG002X  S7HG003X  S7HG005KX  S7HG006KX  S7HG022LX  S7HG023LX  S7HG024LX  S7HG025LX  S7HG029LX  S7HG040X  S7HG041X  S7HG045X  S7HG049X  S7JG007X  S7JG011X  S7JG033LX  S7JG034X  S7JG035X  S7JG041LX  S7JG042LX  S7JG043LX  S7JG044LX  S7JG045LX  S7JG046LX  S7JG047LX  S7KG023LX  S7KG029X  S7KG033X  S7KG039X  S7KG040X  S7KG042X  S7LG001X  S7LG002X  S7LG006X  S7LG011X  S7LG012X  S7MG008LX  S7MG009LX  S7MG011X  S7MG012X  S7MG014X  S7MG018X  S7MG020X  S7MG021X  S7MG022PX  S7MG023PX  S7MG024X  S7MG025X  S7MG026X  S8AG001X  S8AG002X  S8AG011X  S8AG013X  S8CG004X  S8CG010X  S8CG017X  S8CG019PX  S8CG022PX  S8CG025X  S8DG014LX  S8DG015LX  S8DG016LX  S8DG017LX  S8DG018LX  S8DG024X  S8DG027X  S8EG005X  S8EG009X  S8EG010X  S8EG011X",
      "center_classification_date": "20181218",
      "distribution_pattern": "US Nationwide",
      "state": "CT",
      "product_description": "Covidien Emprint Short Percutaneous Antenna with Thermosphere Technology, Material CA15L2 intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.",
      "report_date": "20181226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Covidien LLC",
      "recall_number": "Z-0626-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81595",
      "more_code_info": "",
      "recall_initiation_date": "20181115",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}