  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "9115 Hague Rd",
      "reason_for_recall": "Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature.  Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may unexpectedly change. This creates a risk the app might not transfer the blood glucose result or the user might not correctly input numerical values for carbohydrate used for bolus advice.",
      "address_2": "",
      "product_quantity": "67,040 unique users of version 2.1.0",
      "code_info": "Software Versions: 1.2.0, 1.2.2, 1.2.3, 2.0.0, 2.0.1, and 2.1.0",
      "center_classification_date": "20180215",
      "distribution_pattern": "US Nationwide",
      "state": "IN",
      "product_description": "Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017",
      "report_date": "20180221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Diabetes Care, Inc.",
      "recall_number": "Z-0625-2018",
      "initial_firm_notification": "Other",
      "product_type": "Devices",
      "event_id": "79001",
      "termination_date": "20180727",
      "more_code_info": "",
      "recall_initiation_date": "20170608",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}