  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72832",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AL, AZ, CA, CO, FL, GA, ID, MI, MN, MO, NV, NJ, OK, OR, PA, SD, TN, TX, WA, WI.    OUS: AUSTRALIA, GERMANY, SWITZERLAND, SWEDEN.",
      "recall_number": "Z-0625-2016",
      "product_description": "Zimmer,  Virage OCT Spinal Fixation System, Item Number: 07.01738.001, a posterior system intended to achieve an implant and construct of the Occipital-Cervical-Thoracic spine (Occiput-T3).     UDI Number 889024328952",
      "product_quantity": "372",
      "reason_for_recall": "Zimmer Biomet is initiating a recall of the Virage Occipital Eyelet, as it was found through product testing that the staking of the set screw was out of specification and can interfere with the insertion of the occipital rod when the set screw is staked too low.",
      "recall_initiation_date": "20151117",
      "center_classification_date": "20160109",
      "termination_date": "20160812",
      "report_date": "20160120",
      "code_info": "Lots: 62968065, 62947660, 62536230, 62745662, 62773482, 62758613, 62874869"
    }
  ]
}