  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mettawa",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89366",
      "recalling_firm": "Vyaire Medical",
      "address_1": "26125 N Riverwoods Blvd",
      "address_2": "N/A",
      "postal_code": "60045-3420",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, CT, DC, FL, GA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OK, PA, SD, TX, VA, WA, WI and WV, including Puerto Rico and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, France, Germany, Guatemala, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Namibia, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam, and Zimbabwe.",
      "recall_number": "Z-0622-2022",
      "product_description": "bellavista 1000 ventilator, Catalog #301.100.030 and bellavista 1000e ventilator, 301.100.130.    The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients, and optionally, infant and neonatal patients.",
      "product_quantity": "4189 units",
      "reason_for_recall": "Potential cessation in ventilation can occur under specific conditions.",
      "recall_initiation_date": "20211220",
      "center_classification_date": "20220216",
      "termination_date": "20240828",
      "report_date": "20220223",
      "code_info": "Hardware generation G6, software version V6.0.1600.0 or higher installed, all serial numbers.  UDI 07640149381115 (bellavista 1000 ventilator) and 07640149388879 (bellavista 1000e ventilator)."
    }
  ]
}