  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Plainsboro",
      "address_1": "311 Enterprise Dr",
      "reason_for_recall": "Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type packages were found prior to use but after the sterile package was open.",
      "address_2": "",
      "product_quantity": "78 packages",
      "code_info": "1094085, 1101479, 1101647, 1102403, 1102807, 1104174, 1110043, 1120817, 1130797, 1131331, 1131660",
      "center_classification_date": "20140103",
      "distribution_pattern": "Worldwide Distribution - USA including FL, NE, WI, WA, CT, ND, AL, NY, NH, OR, OK, MN, and TX.  Internationally to Belgium, Czech Republic, Germany, France, Great Britian, Italy, Poland, and Portugal.",
      "state": "NJ",
      "product_description": "Integra Large Rickham -Style Reservoir  Rx Only Sterile for Single Use Only  Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA  NL 850-1132    Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart of the peritoneum.",
      "report_date": "20140115",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Integra LifeSciences Corp.",
      "recall_number": "Z-0622-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67025",
      "termination_date": "20150213",
      "more_code_info": "",
      "recall_initiation_date": "20131113",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}