  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0621-2015",
      "product_description": "PKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080230.   Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "40727  41812  42276  42321  43396  140312  143267  245172  246071  246923  247037  340652  340653  440289  441514  441892  443352  443353  444984  444985  51299  0543138D  0543898D  0640771D  0641429D  0644144D  0646013D  0647316D  0647549D  0740185D  0744066D  0744695D  0746332D  0840071D  0840907D  0842105D  0843936D  0844813D  0846982D  0846986D  0847650D  0847661D  0847671D  92509  0941061D  0941604D  0941726D  0944836D  0945121D  0945472D  0945725D  0945884D  0945983D  1040247D  1040397D  1040426D  1040985D  1042094D  1042466D  1042503D  1042504D  1042940D  1042942D  1044541D  1045186D  1045983D  1046856D  1140701D  1141744D  1142005D  1143438D  1143439D  1144077D  1144676D  1144677D  1145548D  1145549D  1146121D  1240550D  1242235D  1243162  1243162D  1243163  1243596  1340609  1340610  1340610D  1341875  1341876  1342868  1342868D  1342869  1345090  1345090D  1345799  1440293  181299  511298  521298"
    }
  ]
}