      "status": "Terminated",
      "city": "North Shore City",
      "state": "N/A",
      "country": "New Zealand",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63794",
      "recalling_firm": "Fisher & Paykel Healthcare, Ltd.",
      "address_1": "15 Marcel Place",
      "address_2": "East Tamaki",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - USA including IL, IN, AR and GA.",
      "recall_number": "Z-0621-2013",
      "product_description": "Fisher & Paykel Healthcare reusable breathing circuit.  Non-heated breathing circuit, 1.8m or 6 feet in length.    Oxygen therapy delivery system for adult patient.",
      "product_quantity": "200 units total, 26 units US",
      "reason_for_recall": "FPH has become aware that one batch of tubes supplied to FPH may contain hole damage. These tubes were incorporated by FPH into certain 900MR068 reusable circuits with lot numbers 110810 and 111020. If this hole damage is not detected during the standard leak test before patient use, it could potentially result in a gas leak in the breathing system, which may lead to a loss of pressure for the int",
      "recall_initiation_date": "20121119",
      "center_classification_date": "20130103",
      "termination_date": "20130711",
      "report_date": "20130109",
      "code_info": "Lots 110810 and 111020"
    }
  ]
}