  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "40 Liberty Blvd",
      "reason_for_recall": "The possibility exists for the monitor display of Artis one systems with a DP Transceiver below revision 03 to experience blinking and/or a blue/black screen",
      "address_2": "",
      "product_quantity": "207 worldwide (13 distributed in US)",
      "code_info": "Serial # 82044; 82105; 82070; 82054; 82040;82038;82065  82108; 82010; 82039; 82109  82102  82066",
      "center_classification_date": "20180214",
      "distribution_pattern": "US Distribution to the states of :  AR, MO, NY, IN, AR, TX , NC, PA, MI, LA, WA, CA and AL.",
      "state": "PA",
      "product_description": "Artis oneFloor-mounted system for uncompromised imaging; Artis one, Interventional fluoroscopic x-ray system, Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. X-ray generator POLYDOROS A100 Plus",
      "report_date": "20180221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, INC",
      "recall_number": "Z-0614-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78963",
      "termination_date": "20180821",
      "more_code_info": "",
      "recall_initiation_date": "20170822",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}