      "country": "United States",
      "city": "Dover",
      "address_1": "200 W Ohio Ave",
      "reason_for_recall": "Updated on-device label and 3 pages of an updated Operator/Service Manual.  Also, decrease from 3.8 meters (m) to 1.0 separation distance between the device and RF communication equipment.",
      "address_2": "",
      "product_quantity": "8,231 devices total",
      "code_info": "Lot No. Item No. 60400040400, Lot No. 62510548 (Serial Numbers: 2213JAAA, 2213JAAB, 2213JAAC, 2213JAAD, 2213JAAE, 2213JAAF, 2213JAAG, 2213JAAH, 2213JAAJ, 2213JAAK, 2213JAAL, 2213JAAM, 2213JAAN, 2213JAAP, 2213JAAQ, 2213JAAR, 2213JAAS, 2213JAAT, 2213JAAU, 2213JAAW, 2213JABA, 2213JABB, 2213JABC, 2213JABD, 2213JABE, 2213JABF, 2213JABG, 2213JABH);      Lot No. 62534445 (Serial Numbers: 2213JABJ, 2213JABK, 2213JABL, 2213JABM, 2213JABN, 2213JABP, 2213JABQ, 2213JABR, 2213JABS, 2213JABT, 2213JABU, 2213JABW, 2213JACA, 2213JACB, 2213JACC, 2213JACD, 2213JACE, 2213JACF, 2213JACG, 2213JACH);      Lot No. 62631977 (Serial Numbers: 2214BAAA, 2214BAAB, 2214BAAC, 2214BAAD, 2214BAAE, 2214BAAF, 2214BAAG, 2214BAAH, 2214BAAJ, 2214BAAK, 2214BAAL, 2214BAAM, 2214BAAN, 2214BAAP, 2214BAAQ,   2214BAAR, 2214BAAS, 2214BAAT, 2214BAAU, 2214BAAW);      Lot No. 62625059 (Serial Numbers: 2214BABA, 2214BABB, 2214BABC, 2214BABD, 2214BABE, 2214BABF, 2214BABG, 2214BABH, 2214BABJ, 2214BABK, 2214BABL, 2214BABM, 2214BABN, 2214BABP, 2214BABQ, 2214BABR, 2214BABS, 2214BABT, 2214BABU, 2214BABW);      Lot No. 62635789 (Serial Numbers: 2214CAAA, 2214CAAB, 2214CAAC, 2214CAAD, 2214CAAE, 2214CAAF, 2214CAAG, 2214CAAH, 2214CAAJ, 2214CAAK, 2214CAAL, 2214CAAM, 2214CAAV, 2214CAAP, 2214CAAQ, 2214CAAR, 2214CAAS, 2214CAAT, 2214CAAU, 2214CAAW);      Lot No. 62642067 (Serial Numbers: 2214DAAA, 2214DAAB, 2214DAAC, 2214DAAD, 2214DAAE, 2214DAAF, 2214DAAG, 2214DAAH, 2214DAAJ, 2214DAAK, 2214DAAL, 2214DAAM, 2214DAAN, 2214DAAP, 2214DAAQ, 2214DAAR, 2214DAAS, 2214DAAT, 2214DAAU, 2214DAAW);      Lot No. 62642068 (Serial Numbers: 2214EAAA, 2214EAAB, 2214EAAC, 2214EAAD, 2014EAAE, 2214EAAF, 2214EAAG, 2214EAAH, 2214EAAJ, 2214EAAK, 2214EAAL, 2214EAAM, 2214EAAN, 2214EAAP, 2214EAAQ, 2214EAAR, 2214EAAS, 2214EAAT, 2214EAAU, 2214EAAW);      Lot No. 62675190 (Serial Numbers: 2214DABA);      Lot No. 62732714 (Serial Numbers: 2214FAAA, 2214FAAB, 2214FAAC, 2214FAAD);      Lot No. 62727347 (Serial Numbers: 2214FAAE, 2214FAAF, 2214FAAG, 2214FAAH);      Lot No. 62727348 (Serial Numbers: 2214FAAJ, 2214FAAK, 2214FAAL);      Lot No. 62745916 (Serial Numbers: 2214FAAM, 2214FAAN);      Lot No. 62723747 (Serial Numbers: 2214FAAP, 2214FAAQ, 2214FAAR, 2214FAAS, 2214FAAT, 2214FAAU, 2214FAAW, 2214FABA, 2214FABB, 2214FABC, 2214FABD, 2214FABE, 2214FABF, 2214FABG, 2214FABH, 2214FABJ, 2214FABK, 2214FABL, 2214FABM, 2214FABN);      Lot No. 62723748 (Serial Numbers: 2214FABP, 2214FABQ, 2214FABR, 2214FABS, 2214FABT, 2214FABU, 2214FABW, 2214FACA, 2214FACB, 2214FACC, 2214FACD, 2214FACE, 2214FACF, 2214FACG, 2214FACH, 2214FACJ, 2214FACK, 2214FACL, 2214FACM, 2214FACN);      Lot No. 62758708 (Serial Numbers: 2214GAAA, 2214GAAB, 2214GAAC, 2214GAAD);      Lot No. 62758711 (Serial Numbers: 2214GAAE);      Lot No. 62760907 (Serial Numbers: 2214GAAF, 2214GAAG);      Lot No. 63773338 (Serial Numbers: 2214GAAH, 2214GAAJ, 2214GAAK,   2214GAAL);      Lot No. 62773820 (Serial Numbers: 2214GAAM, 2214GAAN);      Lot No. 62773463 (Serial Numbers: 2214GAAP);      Lot No. 62773471 (Serial Numbers: 2214GAAQ, 2214GAAR, 2214GAAS);      Lot No. 62773478 (Serial Numbers: 2214GAAT, 2214GAAU, 2214GAAW, 2214GABA);      Lot No. 62773485 (Serial Numbers: 2214GABB, 2214GABC);      Lot No. 62748403 (Serial Numbers: 2214GABD, 2214GABE, 2214GABF, 2214GABG, 2214GABH, 2214GABJ, 2214GABK, 2214GABL, 2214GABM, 2214GABN, 2214GABP, 2214GABQ, 2214GABR, 2214GABS, 2214GABT, 2214GABU, 2214GABW, 2214GACA, 2214GACB, 2214GACC);      Lot No. 62779664 (Serial Numbers: 2214GACD, 2214GACE);      Lot No. 62797832 (Serial Numbers: 2214HAAA, 2214HAAB, 2214HAAC);      Lot No. 62797833 (Serial Numbers: 2214HAAD, 2214HAAE, 2214HAAF, 2214HAAG, 2214HAAH);      Lot No. 62799449 (Serial Numbers: 2214HAAJ, 2214HAAK, 2214HAAL, 2214HAAM, 2214HAAN, 2214HAAP, 2214HAAQ, 2214HAAR, 2214HAAS, 2214HAAT, 2214HAAU, 2214HAAW, 2214HABA, 2214HABB, 2214HABC, 2214HABD, 2214HABE, 2214HABF, 2214HABG, 2214HABH);      Lot No. 62799475 (Serial Numbers: 2214HABJ, 2214HABK, 2214HABL, 2214HABM, 2214HABN, 2214HABP, 2214HABQ, 2214HABR, 2214HABS, 2214HABT, 2214HABU, 2214HABW, 2214HACA, 2214HACB, 2214HACC, 2214HACD, 2214HACE, 2214HACF, 2214HACG, 2214HACH);      Lot No. 62824836 (Serial Numbers: 2214IAAA, 2214IAAB, 2214IAAC, 2214IAAD);      Lot No. 62838042 (Serial Numbers: 2214IAAE, 2214IAAF, 2214IAAG, 2214IAAH);      Lot No. 62812885 (Serial Numbers: 2214IAAJ, 2214IAAK, 2214IAAL, 2214IAAM, 2214IAAN, 2214IAAP, 2214IAAQ, 2214IAAR, 2214IAAS, 2214IAAT, 2214IAAU, 2214IAAW, 2214IABA, 2214IABB, 2214IABC, 2214IABD, 2214IABE, 2214IABF, 2214IABG, 2214IABH);      Lot No. 62812853 (Serial Numbers: 2214IABJ, 2214IABK, 2214IABL, 2214IABM, 2214IABN, 2214IABP, 2214IABQ, 2214IABR, 2214IABS, 2214IABT, 2214IABU, 2214IABW, 2214IACA, 2214IACB, 2214IACC, 2214IACD, 2214IACE, 2214IACF, 2214IACG, 2214IACH);      Lot No. 62846159 (Serial Numbers: 2214JAAA, 2214JAAB, 2214JAAC);      Lot No. 62844916 (Serial Numbers: 2214JAAD);      Lot No. 62869981 (Serial Numbers: 2214KAAE, 2214KAAF, 2214KAAG, 2214KAAH, 2214KAAJ);      Lot No. 62871639 (Serial Numbers: 2214JAAE, 2214JAAF, 2214JAAG, 2214JAAH);      Lot No. 62870928 (Serial Numbers: 2214KAAK, 2214KAAL, 2214KAAM, 2214KAAN, 2214KAAP, 2214KAAQ, 2214KAAR, 2214KAAS, 2214KAAT, 2214KAAU, 2214KAAW, 2214KABA, 2214KABB, 2214KABC, 2214KABD, 2214KABE, 2214KABF, 2214KABG, 2214KABH, 2214KABJ);      Lot No. 62877932 (Serial Numbers: 221411110001, 221411110002, 221411110003, 221411110004, 221411110005, 221411110006, 221411110007, 221411110008, 221411110009, 221411110010);      Lot No. 62885586 (Serial Numbers: 221411120001, 221411120002, 221411120003, 221411120004, 221411120005);      Lot No. 62885587 (Serial Numbers: 221411120006, 221411120007, 221411120008, 221411120009, 221411120010, 221411120011, 221411120012, 221411120013, 221411120014, 221411120015);      Lot No. 62895898 (Serial Numbers: 221411200001, 221411200002, 221411200003, 221411200004, 221411200005, 221411200006, 221411200007, 221411200008, 221411200009, 221411200010);      Lot No. 62905257 (Serial Numbers: 221411240001);      Lot No. 62904091 (Serial Numbers: 221411250001, 221411250002, 221411250003, 221411250004, 221411250005, 221411250006, 221411250007, 221411250008, 221411250009, 221411250010, 221411250011, 221411250012, 221411250013, 221411250014, 221411250015, 221411250016, 221411250017, 221411250018, 221411250019, 221411250020);      Lot No. 62904093 (Serial Numbers: 221412010001, 221412010002, 221412010003, 221412010004, 221412010005, 221412010006, 221412010007, 221412010008, 221412010009, 221412010010, 221412010011, 221412010012, 221412010013, 221412010014, 221412010015, 221412010016, 221412010017, 221412010018, 221412010019, 221412010020);      Lot No. 62879139 (Serial Numbers: 221412040001, 221412040002, 221412040003, 221412040004, 221412040005);      Lot No. 62905838 (Serial Numbers: 221412080001, 221412080002, 221412080003, 221412080004, 221412080005);      Lot No. 62895896 (Serial Numbers: 221412120001, 221412120002, 221412120003, 221412120004, 221412120005);      Lot No. 62904123 (Serial Numbers: 221412150001, 221412150002, 221412150003, 221412150004, 221412150005, 221412150006, 221412150007, 221412150008, 221412150009, 221412150010, 221412150011, 221412150012, 221412150013, 221412150014, 221412150015, 221412150016, 221412150017, 221412150018, 221412150019, 221412150020);      Lot No. 62932244 (Serial Numbers: 221412180001, 221412180002, 221412180003, 221412180004, 221412180005, 221412180006, 221412180007, 221412180008, 221412180009, 221412180010, 221412180011, 221412180012, 221412180013, 221412180014, 221412180015, 221412180016, 221412180017, 221412180018, 221412180019, 221412180020);      Lot No. 62935217 (Serial Numbers: 221412190001, 221412190002, 221412190003, 221412190004, 221412190005);      Lot No. 62937693 (Serial Numbers: 221501050001, 221501050002, 221501050003, 221501050004, 221501050005);      Lot No. 62945082 (Serial Numbers: 221501080001, 221501080002, 221501080003, 221501080004, 221501080005);      Lot No. 62945086 (Serial Numbers: 201501080006, 201501080007, 201501080008);      Lot No. 62932650 (Serial Numbers: 221503090001, 221503090002, 221503090003, 221503090004, 221503090005, 221503090006, 221503090007, 221503090008, 221503090009, 221503090010, 221503090011, 221503090012, 221503090013, 221503090014, 221503090015, 221503090016, 221503090017, 221503090018, 221503090019, 22150309002);      Lot No. 62961804 (Serial Numbers: 221503120001, 221503120002, 221503120003, 221503120004, 221503120005, 221503120006, 221503120007, 221503120008, 221503120009, 221503120010, 221503120011, 221503120012, 221503120013, 221503120014, 221503120015, 221503120016, 221503120017, 221503120018, 221503120019, 221503120020);      Lot No. 62976480 (Serial Numbers: 221503190001, 221503190002, 221503190003, 221503190004, 221503190005, 221503190006, 221503190007, 221503190008, 221503190009, 221503190010, 221503190011, 221503190012, 221503190013, 221503190014, 221503190015, 221503190016, 221503190017, 221503190018, 221503190019, 221503190020);      Lot No. 62976481 (Serial Numbers: 221503250001, 221503250002, 221503250003, 221503250004, 221503250005, 221503250006, 221503250007, 221503250008, 221503250009, 221503250010, 221503250011, 221503250012, 221503250013, 221503250014, 221503250015, 221503250016, 221503250017, 221503250018, 221503250019, 221503250020);      Lot No. 63026460 (Serial Numbers: 221503270001, 221503270002, 221503270003, 221503270004, 221503270005);      Lot No. 63007852 (Serial Numbers: 221504080001, 221504080002, 221504080003, 221504080004, 221504080005, 221504080006, 221504080007, 221504080008, 221504080009, 221504080010, 221504080011, 221504080012, 221504080013, 221504080014, 221504080015, 221504080016, 221504080017, 221504080018, 221504080019, 221504080020    Lot No. 63042181 (Serial Numbers: 221504140001, 221504140002, 221504140003, 221504140004, 221504140005);      Lot No. 63042182 (Serial Numbers: 221504140006, 221504140007, 221504140008, 221504140009, 221504140010);      Lot No. 63007853 (Serial Numbers: 221504140011, 221504140012, 221504140013, 221504140014, 221504140015, 221504140016, 221504140017, 221504140018, 221504140019, 221504140020, 221504140021, 221504140022, 221504140023, 221504140024, 221504140025, 221504140026, 221504140027, 221504140028, 221504140029, 221504140030);      Lot No. 63042179 (Serial Numbers: 221504160001, 221504160002, 221504160003);      Lot No. 63042184 (Serial Numbers: 221504220001, 221504220002, 221504220003, 221504220004, 221504220005);      Lot No. 63037252 (Serial Numbers: 221504250001, 221504250002, 221504250003, 221504250004, 221504250005, 221504250006, 221504250007, 221504250008, 221504250009, 221504250010, 221504250011, 221504250012, 221504250013, 221504250014, 221504250015, 221504250016, 221504250017, 221504250018, 221504250019, 221504250020);      Lot No. 63042180 (Serial Numbers: 221504280001, 221504280002, 221504280003, 221504280004, 221504280005, 221504280006, 221504280007, 221504280008, 221504280009, 221504280010, 221504280011, 221504280012, 221504280013, 221504280014, 221504280015, 221504280016, 221504280017, 221504280018, 221504280019, 221504280020);      Lot No. 63042183 (Serial Numbers: 221505110001, 221505110002, 221505110003, 221505110004, 221505110005);      Lot No. 63037254 (Serial Numbers: 221505120001, 221505120002, 221505120003, 221505120004, 221505120005, 221505120006, 221505120007, 221505120008, 221505120009, 221505120010, 221505120011, 221505120012, 221505120013, 221505120014, 221505120015, 221505120016, 221505120017, 221505120018, 221505120019, 221505120020);      Lot No. 63092209 (Serial Numbers: 221506030001, 221506030002, 221506030003);      Lot No. 63092210 (Serial Numbers: 221506050001, 221506050002, 221506050003, 221506050004, 221506050005);      Lot No. 63072301 (Serial Numbers: 221506100001, 221506100002, 221506100003, 221506100004, 221506100005, 221506100006, 221506100007, 221506100008, 221506100009, 221506100010, 221506100011, 221506100012, 221506100013, 221506100014, 221506100015, 221506100016, 221506100017, 221506100018, 221506100019, 221506100020);      Lot No. 63103587 (Serial Numbers: 221506130001, 221506130002, 221506130003, 221506130004, 221506130005, 221506130006, 221506130007, 221506130008, 221506130009, 221506130010, 221506130011, 221506130012, 221506130013, 221506130014, 221506130015);      Lot No. 63072324 (Serial Numbers: 221506170001, 221506170002, 221506170003, 221506170004, 221506170005, 221506170006, 221506170007, 221506170008, 221506170009, 221506170010, 221506170011, 221506170012, 221506170013, 221506170014, 221506170015, 221506170016, 221506170017, 221506170018, 221506170019, 221506170020);      Lot No. 63103164 (Serial Numbers: 221506170021, 221506170022, 221506170023);      Lot No. 63072352 (Serial Numbers: 221506200001, 221506200002, 221506200003, 221506200004, 221506200005, 221506200006, 221506200007, 221506200008, 221506200009, 221506200010, 221506200011, 221506200012, 221506200013, 221506200014, 221506200015, 221506200016, 221506200017, 221506200018, 221506200019, 221506200020);      Lot No. 63096468 (Serial Numbers: 221507150001, 221507150002, 221507150003, 221507150004, 221507150005, 221507150006, 221507150007, 221507150008, 221507150009, 221507150010, 221507150011, 221507150012, 221507150013, 221507150014, 221507150015, 221507150016, 221507150017, 221507150018, 221507150019, 221507150020);      Lot No. 63109655 (Serial Numbers: 221507160001, 221507160002, 221507160003);      Lot No. 63103590 (Serial Numbers: 221507210001, 221507210002, 221507210003, 221507210004, 221507210005, 221507210006, 221507210007, 221507210008, 221507210009, 221507210010, 221507210011, 221507210012, 221507210013, 221507210014, 221507210015);      Lot No. 63109651 (Serial Numbers: 221507280001, 221507280002, 221507280003, 221507280004, 221507280005, 221507280006, 221507280007, 221507280008, 221507280009, 221507280010);      Lot No. 63116357 (Serial Numbers: 221507290001, 221507290002, 221507290003, 221507290004, 221507290005, 221507290006, 221507290007, 221507290008, 221507290009, 221507290010, 221507290011, 221507290012, 221507290013, 221507290014, 221507290015, 221507290016, 221507290017, 221507290018, 221507290019, 221507290020);      Lot No. 63123305 (Serial Numbers: 221508040001, 221508040002, 221508040003, 221508040004, 221508040005, 221508040006, 221508040007, 221508040008, 221508040009, 221508040010, 221508040011, 221508040012, 221508040013, 221508040014, 221508040015, 221508040016, 221508040017, 221508040018, 221508040019, 221508040020);      Lot No. 63138553 (Serial Numbers: 221508260001, 221508260002, 221508260003, 221508260004, 221508260005, 221508260006, 221508260007, 221508260008, 221508260009, 221508260010, 221508260011, 221508260012, 221508260013, 221508260014, 221508260015, 221508260016, 221508260017, 221508260018, 221508260019, 221508260020);      Lot No. 63138631 (Serial Numbers: 221509090001, 221509090002, 221509090003, 221509090004, 221509090005, 221509090006, 221509090007, 221509090008, 221509090009, 221509090010, 221509090011, 221509090012, 221509090013, 221509090014, 221509090015, 221509090016, 221509090017, 221509090018, 221509090019, 221509090020);      Lot No. 63171089 (Serial Numbers: 221509240001, 221509240002, 221509240003, 221509240004, 221509240005, 221509240006, 221509240007, 221509240008, 221509240009, 221509240010, 221509240011, 221509240012, 221509240013, 221509240014, 221509240015, 221509240016, 221509240017, 221509240018, 221509240019, 221509240020);      Lot No. 63202480 (Serial Numbers: 221510120001, 221510120002, 221510120003, 221510120004, 221510120005);      Lot No. 63202478 (Serial Numbers: 221510220001, 221510220002, 221510220003, 221510220004, 221510220005, 221510220006, 221510220007, 221510220008, 221510220009, 221510220010, 221510220011, 221510220012, 221510220013, 221510220014, 221510220015, 221510220016, 221510220017, 221510220018, 221510220019, 221510220020);      Lot No. 63234563 (Serial Numbers: 221511230001, 221511230002, 221511230003, 221511230004, 221511230005);      Lot No. 63234565 (Serial Numbers: 221512020001, 221512020002, 221512020003, 221512020004, 221512020005);      Lot No. 63212407 (Serial Numbers: 221512070001, 221512070002, 221512070003, 221512070004, 221512070005, 221512070006, 221512070007, 221512070008, 221512070009, 221512070010, 221512070011, 221512070012, 221512070013, 221512070014, 221512070015, 221512070016, 221512070017, 221512070018, 221512070019, 221512070020);      Lot No. 63245752 (Serial Numbers: 221512160001, 221512160002, 221512160003, 221512160004, 221512160005, 221512160006, 221512160007);      Lot No. 63253256 (Serial Numbers: 221512180001, 221512180002, 221512180003, 221512180004, 221512180005);      Lot No. 63247425 (Serial Numbers: 221601060001, 221601060002, 221601060003, 221601060004, 221601060005, 221601060006, 221601060007, 221601060008, 221601060009, 221601060010, 221601060011, 221601060012, 221601060013, 221601060014, 221601060015, 221601060016, 221601060017, 221601060018, 221601060019, 221601060020);      Lot No. 63526287 (Serial Numbers: 221612140001, 221612140002, 221612140003, 221612140004, 221612140005, 221612140006, 221612140007, 221612140008, 221612140009, 221612140010, 221612140011, 221612140012, 221612140013, 221612140014, 221612140015, 221612140016, 221612140017, 221612140018, 221612140019, 221612140020);      Lot No. 63531691 (Serial Numbers: 221612150001, 221612150002, 221612150003, 221612150004, 221612150005, 221612150006, 221612150007);      Lot No. 63526321 (Serial Numbers: 221612170001, 221612170002, 221612170003, 221612170004, 221612170005, 221612170006, 221612170007, 221612170008, 221612170009, 221612170010, 221612170011, 221612170012, 221612170013, 221612170014, 221612170015, 221612170016, 221612170017, 221612170018, 221612170019, 221612170020);      Lot No. 63261584 (Serial Numbers: 221601070001, 221601070002, 221601070003, 221601070004, 221601070005);      Lot No. 63264374 (Serial Numbers: 221601110001, 221601110002, 221601110003, 221601110004, 221601110005, 221601110006, 221601110007, 221601110008, 221601110009, 221601110010, 221601110011, 221601110012, 221601110013, 221601110014, 221601110015, 221601110016, 221601110017, 221601110018, 221601110019, 221601110020);      Lot No. 63265280 (Serial Numbers: 221601140001, 221601140002, 221601140003, 221601140004, 221601140005, 221601140006, 221601140007);      Lot No. 63257987 (Serial Numbers: 221601180001, 221601180002, 221601180003, 221601180004, 221601180005, 221601180006, 221601180007, 221601180008, 221601180009, 221601180010, 221601180011, 221601180012, 221601180013, 221601180014, 221601180015, 221601180016, 221601180017, 221601180018, 221601180019, 221601180020);      Lot No. 63271202 (Serial Numbers: 221601220001, 221601220002, 221601220003, 221601220004, 221601220005);      Lot No. 63258039 (Serial Numbers: 221602060001, 221602060002, 221602060003, 221602060004, 221602060005, 221602060006, 221602060007, 221602060008, 221602060009, 221602060010, 221602060011, 221602060012, 221602060013, 221602060014, 221602060015, 221602060016, 221602060017, 221602060018, 221602060019, 221602060020);      Lot No. 63278388 (Serial Numbers: 221602150001, 221602150002, 221602150003, 221602150004, 221602150005, 221602150006, 221602150007, 221602150008, 221602150009, 221602150010, 221602150011, 221602150012, 221602150013, 221602150014, 221602150015, 221602150016, 221602150017, 221602150018, 221602150019, 221602150020);      Lot No. 63278436 (Serial Numbers: 221602240001, 221602240002, 221602240003, 221602240004, 221602240005, 221602240006, 221602240007, 221602240008, 221602240009, 221602240010, 221602240011, 221602240012, 221602240013, 221602240014, 221602240015, 221602240016, 221602240017, 221602240018, 221602240019, 221602240020);      Lot No. 63314121 (Serial Numbers: 221603090001, 221603090002, 221603090003);      Lot No. 63314118 (Serial Numbers: 221603110001, 221603110002, 221603110003);      Lot No. 63324625 (Serial Numbers: 221603220001, 221603220002, 221603220003);      Lot No. 63336959 (Serial Numbers: 221603300001, 221603300002, 221603300003, 221603300004, 221603300005, 221603300006, 221603300007, 221603300008, 221603300009, 221603300010);      Lot No. 63314042 (Serial Numbers: 221603300011, 221603300012, 221603300013, 221603300014, 221603300015, 221603300016, 221603300017, 221603300018, 221603300019, 221603300020, 221603300021, 221603300022, 221603300023, 221603300024, 221603300025, 221603300026, 221603300027, 221603300028, 221603300029, 221603300030);      Lot No. 63341344 (Serial Numbers: 221604040001, 221604040002, 221604040003, 221604040004, 221604040005);      Lot No. 63341346 (Serial Numbers: 221604050001, 221604050002, 221604050003, 221604050004, 221604050005, 221604050006, 221604050007, 221604050008, 221604050009, 221604050010);      Lot No. 63314117 (Serial Numbers: 221604220001, 221604220002, 221604220003, 221604220004, 221604220005, 221604220006, 221604220007, 221604220008, 221604220009, 221604220010, 221604220011, 221604220012, 221604220013, 221604220014, 221604220015, 221604220016, 221604220017, 221604220018, 221604220019, 221604220020);      Lot No. 63352349 (Serial Numbers: 221604250001, 221604250002, 221604250003);      Lot No. 63339440 (Serial Numbers: 221605020001, 221605020002, 221605020003, 221605020004, 221605020005, 221605020006, 221605020007, 221605020008, 221605020009, 221605020010, 221605020011, 221605020012, 221605020013, 221605020014, 221605020015, 221605020016, 221605020017, 221605020018, 221605020019, 221605020020);      Lot No. 63343366 (Serial Numbers: 221605060001, 221605060002, 221605060003, 221605060004, 221605060005, 221605060006, 221605060007, 221605060008, 221605060009, 221605060010, 221605060011, 221605060012, 221605060013, 221605060014, 221605060015, 221605060016, 221605060017, 221605060018, 221605060019, 221605060020, 221605060021, 221605060022);      Lot No. 63352347 (Serial Numbers: 221605110001, 221605110002, 221605110003, 221605110004, 221605110005);      Lot No. 63339497 (Serial Numbers: 221605120001, 221605120002, 221605120003, 221605120004, 221605120005, 221605120006, 221605120007, 221605120008, 221605120009, 221605120010, 221605120011, 221605120012, 221605120013, 221605120014, 221605120015, 221605120016, 221605120017, 221605120018, 221605120019, 221605120020);      Lot No. 63378855 (Serial Numbers: 221606020001, 221606020002, 221606020003);      Lot No. 63369605 (Serial Numbers: 221606110001, 221606110002, 221606110003, 221606110004, 221606110005, 221606110006, 221606110007, 221606110008, 221606110009, 221606110010, 221606110011, 221606110012, 221606110013, 221606110014, 221606110015, 221606110016, 221606110017, 221606110018, 221606110019, 221606110020);      Lot No. 63403010 (Serial Numbers: 221606110021, 221606110022, 221606110023);      Lot No. 63403012 (Serial Numbers: 221606150001, 221606150002, 221606150003, 221606150004, 221606150005, 221606150006, 221606150007, 221606150008);      Lot No. 63403026 (Serial Numbers: 221606210001, 221606210002, 221606210003, 221606210004, 221606210005, 221606210006, 221606210007, 221606210008, 221606210009, 221606210010, 221606210011, 221606210012, 221606210013, 221606210014, 221606210015, 221606210016, 221606210017, 221606210018, 221606210019, 221606210020);      Lot No. 63403009 (Serial Numbers: 221606220001, 221606220002, 221606220003);      Lot No. 63403024 (Serial Numbers: 221606230001, 221606230002, 221606230003, 221606230004, 221606230005, 221606230006, 221606230007, 221606230008, 221606230009, 221606230010, 221606230011, 221606230012, 221606230013, 221606230014, 221606230015, 221606230016, 221606230017, 221606230018, 221606230019, 221606230020);      Lot No. 63403025 (Serial Numbers: 221606240001, 221606240002, 221606240003, 221606240004, 221606240005, 221606240006, 221606240007, 221606240008, 221606240009, 221606240010);      Lot No. 63403017 (Serial Numbers: 221606240011, 221606240012, 221606240013, 221606240014, 221606240015);      Lot No. 63394148 (Serial Numbers: 221606290001, 221606290002, 221606290003, 221606290004, 221606290005, 221606290006, 221606290007, 221606290008, 221606290009, 221606290010, 221606290011, 221606290012, 221606290013, 221606290014, 221606290015, 221606290016, 221606290017, 221606290018, 221606290019, 221606290020);      Lot No. 63410582 (Serial Numbers: 221607050001, 221607050002, 221607050003, 221607050004, 221607050005, 221607050006, 221607050007, 221607050008, 221607050009, 221607050010, 221607050011, 221607050012, 221607050013, 221607050014, 221607050015);      Lot No. 63422767 (Serial Numbers: 221607080001, 221607080002, 221607080003, 221607080004, 221607080005, 221607080006, 221607080007, 221607080008, 221607080009, 221607080010, 221607080011, 221607080012, 221607080013, 221607080014, 221607080015, 221607080016, 221607080017, 221607080018, 221607080019, 221607080020);      Lot No. 63410583 (Serial Numbers: 221607210001, 221607210002, 221607210003, 221607210004, 221607210005, 221607210006, 221607210007, 221607210008, 221607210009, 221607210010);      Lot No. 63410585 (Serial Numbers: 221607220001, 221607220002, 221607220003, 221607220004, 221607220005, 221607220006, 221607220007, 221607220008, 221607220009, 221607220010);      Lot No. 63394176 (Serial Numbers: 221607260001, 221607260002, 221607260003, 221607260004, 221607260005, 221607260006, 221607260007, 221607260008, 221607260009, 221607260010, 221607260011, 221607260012, 221607260013, 221607260014, 221607260015, 221607260016, 221607260017, 221607260018, 221607260019, 221607260020);      Lot No. 63430499 (Serial Numbers: 221608080001, 221608080002, 221608080003, 221608080004, 221608080005, 221608080006, 221608080007, 221608080008, 221608080009, 221608080010, 221608080011, 221608080012, 221608080013, 221608080014, 221608080015, 221608080016, 221608080017, 221608080018, 221608080019, 221608080020);      Lot No. 63437710 (Serial Numbers: 221608120001, 221608120002, 221608120003, 221608120004, 221608120005);      Lot No. 63436338 (Serial Numbers: 221608160001, 221608160002, 221608160003, 221608160004, 221608160005);      Lot No. 63445193 (Serial Numbers: 221608170001, 221608170002, 221608170003, 221608170004, 221608170005, 221608170006, 221608170007, 221608170008, 221608170009, 221608170010, 221608170011, 221608170012, 221608170013, 221608170014, 221608170015, 221608170016, 221608170017, 221608170018, 221608170019, 221608170020);      Lot No. 63430540 (Serial Numbers: 221608190001, 221608190002, 221608190003, 221608190004, 221608190005, 221608190006, 221608190007, 221608190008, 221608190009, 221608190010, 221608190011, 221608190012, 221608190013, 221608190014, 221608190015, 221608190016, 221608190017, 221608190018, 221608190019, 221608190020);      Lot No. 63445194 (Serial Numbers: 221608230001, 221608230002, 221608230003, 221608230004, 221608230005, 221608230006, 221608230007, 221608230008, 221608230009, 221608230010, 221608230011, 221608230012, 221608230013, 221608230014, 221608230015, 221608230016, 221608230017, 221608230018, 221608230019, 221608230020);      Lot No. 63455066 (Serial Numbers: 221609010001, 221609010002, 221609010003, 221609010004, 221609010005, 221609010006, 221609010007, 221609010008, 221609010009, 221609010010, 221609010011, 221609010012, 221609010013, 221609010014, 221609010015, 221609010016, 221609010017, 221609010018, 221609010019, 221609010020);      Lot No. 63474281 (Serial Numbers: 221609080001, 221609080002, 221609080003, 221609080004, 221609080005, 221609080006, 221609080007, 221609080008, 221609080009, 221609080010, 221609080011, 221609080012, 221609080013, 221609080014, 221609080015, 221609080016, 221609080017, 221609080018, 221609080019, 221609080020);      Lot No. 63474283 (Serial Numbers: 221609080021, 221609080022, 221609080023, 221609080024, 221609080025, 221609080026);      Lot No. 63474284 (Serial Numbers: 221609150001, 221609150002, 221609150003, 221609150004, 221609150005, 221609150006, 221609150007, 221609150008);      Lot No. 63474287 (Serial Numbers: 221609210001, 221609210002, 221609210003, 221609210004, 221609210005, 221609210006, 221609210007, 221609210008, 221609210009, 221609210010);      Lot No. 63474285 (Serial Numbers: 221610010001, 221610010002, 221610010003, 221610010004, 221610010005, 221610010006, 221610010007, 221610010008, 221610010009, 221610010010, 221610010011, 221610010012, 221610010013, 221610010014, 221610010015);      Lot No. 63455098 (Serial Numbers: 221610050001, 221610050002, 221610050003, 221610050004, 221610050005, 221610050006, 221610050007, 221610050008, 221610050009, 221610050010, 221610050011, 221610050012, 221610050013, 221610050014, 221610050015, 221610050016, 221610050017, 221610050018, 221610050019, 221610050020);      Lot No. 63474282 (Serial Numbers: 221610110001, 221610110002, 221610110003, 221610110004, 221610110005, 221610110006, 221610110007, 221610110008, 221610110009, 221610110010, 221610110011, 221610110012, 221610110013, 221610110014, 221610110015, 221610110016, 221610110017, 221610110018, 221610110019, 221610110020);      Lot No. 63484747 (Serial Numbers: 221610140001, 221610140002, 221610140003, 221610140004, 221610140005);      Lot No. 63482404 (Serial Numbers: 221610180001, 221610180002, 221610180003, 221610180004, 221610180005, 221610180006, 221610180007, 221610180008, 221610180009, 221610180010, 221610180011, 221610180012, 221610180013, 221610180014, 221610180015, 221610180016, 221610180017, 221610180018, 221610180019, 221610180020);      Lot No. 63482405 (Serial Numbers: 221610200001, 221610200002, 221610200003, 221610200004, 221610200005, 221610200006, 221610200007, 221610200008, 221610200009, 221610200010, 221610200011, 221610200012, 221610200013, 221610200014, 221610200015, 221610200016, 221610200017, 221610200018, 221610200019, 221610200020);      Lot No. 63482408 (Serial Numbers: 221611210001, 221611210002, 221611210003, 221611210004, 221611210005, 221611210006, 221611210007, 221611210008, 221611210009, 221611210010, 221611210011,221611210012, 221611210013, 221611210014, 221611210015, 221611210016, 221611210017, 221611210018, 221611210019, 221611210020);      Lot No. 63501470 (Serial Numbers: 221611210021, 221611210022, 221611210023, 221611210024, 221611210025);      Lot No. 63501471 (Serial Numbers: 221611210021, 221611210022, 221611210023, 221611210024, 221611210025);      Lot No. 63501459 (Serial Numbers: 221611280001, 221611280002, 221611280003, 221611280004, 221611280005, 221611280006, 221611280007, 221611280008, 221611280009, 221611280010, 221611280011, 221611280012, 221611280013, 221611280014, 221611280015, 221611280016, 221611280017, 221611280018, 221611280019, 221611280020);      Lot No. 63501486 (Serial Numbers: 221612020001, 221612020002, 221612020003, 221612020004, 221612020005, 221612020006, 221612020007, 221612020008, 221612020009, 221612020010, 221612020011, 221612020012, 221612020013, 221612020014, 221612020015, 221612020016, 221612020017, 221612020018, 221612020019, 221612020020);      Lot No. 63501469 (Serial Numbers: 221612030001, 221612030002, 221612030003, 221612030004, 221612030005, 221612030006, 221612030007, 221612030008, 221612030009, 221612030010, 221612030011, 221612030012, 221612030013, 221612030014, 221612030015, 221612030016, 221612030017, 221612030018);      Lot No. 63531693 (Serial Numbers: 221612050001, 221612050002, 221612050003, 221612050004, 221612050005);      Lot No. 63501509 (Serial Numbers: 221612070001, 221612070002, 221612070003, 221612070004, 221612070005, 221612070006, 221612070007, 221612070008, 221612070009, 221612070010, 221612070011, 221612070012, 221612070013, 221612070014, 221612070015, 221612070016, 221612070017, 221612070018, 221612070019, 221612070020);      Lot No. 63501547 (Serial Numbers: 221612100001, 221612100002, 221612100003, 221612100004, 221612100005, 221612100006, 221612100007, 221612100008, 221612100009, 221612100010, 221612100011, 221612100012, 221612100013, 221612100014, 221612100015, 221612100016, 221612100017, 221612100018, 221612100019, 221612100020);      Lot No. 63501493 (Serial Numbers: 221612130001, 221612130002, 221612130003, 221612130004, 221612130005, 221612130006, 221612130007, 221612130008)",
      "center_classification_date": "20180214",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Guam; and countries to Australia, Brazil, Canada, Chile, China, Denmark, Dominican, El Salvador, England, Finland, France, Hong Kong, India, Italy, Japan, Malaysia, Mexico, Netherlands, Panama, Singapore, Sweden, Taiwan, and Thailand..",
      "state": "OH",
      "product_description": "A.T.S. 2200TS Tourniquet Systems, Personalized Pressure.  Zimmer Surgical, Inc. Dover, OH 44622     Intended to be used by qualified professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.",
      "report_date": "20180221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Surgical Inc",
      "recall_number": "Z-0613-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78759",
      "termination_date": "20190801",
      "more_code_info": "",
      "recall_initiation_date": "20170727",
      "postal_code": "44622-9642",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}