  "results": [
    {
      "status": "Terminated",
      "city": "Christchurch",
      "state": "",
      "country": "New Zealand",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84217",
      "recalling_firm": "Enztec Limited",
      "address_1": "26 Dakota Crescent, Sockburn",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "CA",
      "recall_number": "Z-0612-2020",
      "product_description": "REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE",
      "product_quantity": "312 Femoral Impactors",
      "reason_for_recall": "The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument.  This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.",
      "recall_initiation_date": "20190816",
      "center_classification_date": "20191206",
      "termination_date": "20230302",
      "report_date": "20210203",
      "code_info": "Lot number Distribution Date:  161471 08-Jun-16,  161472 08-Jun-16,  161473 08-Jun-16,  161474 08-Jun-16,  161475 08-Jun-16,  161476 08-Jun-16,  161477 08-Jun-16,  161756 29-Jun-16,  161757 29-Jun-16,  161758 29-Jun-16,  161759 29-Jun-16,  161760 29-Jun-16,  161761 29-Jun-16,  161762 29-Jun-16,  161822 22-Jul-16,  161823 22-Jul-16,  161824 22-Jul-16,  161825 22-Jul-16,  161826 22-Jul-16,  161827 22-Jul-16,  161828 22-Jul-16,  170760 24-Mar-17,  170761 24-Mar-17,  170762 24-Mar-17,  170763 24-Mar-17,  170764 24-Mar-17,  170765 24-Mar-17,  170766 24-Mar-17,  170767 24-Mar-17,  170768 24-Mar-17,  171989 23-Jun-17  ,",
      "more_code_info": ""
    }
  ]
}