{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "North Industrial Park",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81521",
      "recalling_firm": "Mazor Robotics Ltd",
      "address_1": "5 Shacham Street",
      "address_2": "P.O. Box 3104",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Affected devices were distributed in the US to four different medical facilities in four states",
      "recall_number": "Z-0609-2019",
      "product_description": "Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038).  Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of, high contrast objects.",
      "product_quantity": "5 units",
      "reason_for_recall": "In some situations the carousel position pin (locker) may cause the carousel upper plate to pop out and the carousel may then inadvertently move.",
      "recall_initiation_date": "20141126",
      "center_classification_date": "20181214",
      "termination_date": "20190703",
      "report_date": "20181226",
      "code_info": "Affected products: all Carousels under Level 01 were affected and therefore are applicable to this issue.      The related lots include: WP065611, WP066389, WP066411, WP066412, WP066420.      Explanation of the coding system: a part number (e.g. MAS1025-01) is composed of a generic part number related to the part (e.g. MAS1025) and the level of the part (e.g. 01).",
      "more_code_info": ""
    }
  ]
}