  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84236",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0608-2020",
      "product_description": "HeartStart XL+ Defibrillator/Monitor, Model 861290",
      "product_quantity": "24,738 units",
      "reason_for_recall": "Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.",
      "recall_initiation_date": "20191003",
      "center_classification_date": "20191205",
      "termination_date": "20230419",
      "report_date": "20191211",
      "code_info": "All Units, Model number 861290",