  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Duderstadt",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72960",
      "recalling_firm": "Otto Bock Healthcare GmbH",
      "address_1": "Max-Nader-Str. #15",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-0608-2016",
      "product_description": "Ottobock Modular Polycentric EBS Knee Joint 3R60, 3R60=ST, 3R60=KD, 3R60=HD.  Used for the prosthetic fitting of amputations of the lower. limb.",
      "product_quantity": "40 units",
      "reason_for_recall": "Otto Bock HealthCare GmbH is recalling Modular Polycentric EBS Knee Joint that were serviced between 27 April 2015 and 17 November 2015 due to a service error where the wrong Loctite was  used during repair.",
      "recall_initiation_date": "20160104",
      "center_classification_date": "20160107",
      "termination_date": "20160502",
      "report_date": "20160113",
      "code_info": "All serviced between 27 April 2015 and 17 November 2015"
    }
  ]
}