  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Westlake",
      "state": "OH",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "37530",
      "recalling_firm": "Radiometer America Inc",
      "address_1": "810 Sharon Dr",
      "address_2": "N/A",
      "postal_code": "44145-1598",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to KY, NC, and NY.",
      "recall_number": "Z-0608-2013",
      "product_description": "Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042.     The solution is used for quality check of blood gas analyzers.",
      "product_quantity": "840 Ampules",
      "reason_for_recall": "The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.",
      "recall_initiation_date": "20060606",
      "center_classification_date": "20121221",
      "termination_date": "20130205",
      "report_date": "20130102",
      "code_info": "Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008."
    }
  ]
}