  "results": [
    {
      "country": "United States",
      "city": "Deerfield",
      "address_1": "1 Baxter Pkwy",
      "reason_for_recall": "Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill syringes with the HEPA filtration disabled.",
      "address_2": "",
      "product_quantity": "28 devices",
      "code_info": "Lot # 123, 125R, 127, 129, 145, 146R, 147R, 148, 215R, 500, 501, 502, 503, 504, 505, 506, 508, 510, 511R, 512, 514, 515, 516R, 517, 518, 519, and 520.",
      "center_classification_date": "20131229",
      "distribution_pattern": "Nationwide Distribution including the states of CA, CO, CT, FL, IL, LA, MA, MD, MI, NC, NE, NY, OH, OR, PA, TN, TX, and WV.",
      "state": "IL",
      "product_description": "IntelliFill i.v. Pharmacy Compounding Device (PCD).    Intended to fill syringes from 0.5mL up to 11.5mL using mixing instructions from its formulary to fill the orders.",
      "report_date": "20140108",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Baxter Healthcare Corp.",
      "recall_number": "Z-0607-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66979",
      "termination_date": "20150323",
      "more_code_info": "",
      "recall_initiation_date": "20131108",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}