  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72869",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "N/A",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of  Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.",
      "recall_number": "Z-0604-2016",
      "product_description": "Oncentra External Beam - VMAT - Radiation therapy planning system    Product Usage:  The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.",
      "product_quantity": "154",
      "reason_for_recall": "When using the option \"Tumor Overlap Fraction\" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume.  This could result in open MLC, and open jaws in areas away from the target volume.",
      "recall_initiation_date": "20151218",
      "center_classification_date": "20160106",
      "termination_date": "20210915",
      "report_date": "20160113",
      "code_info": "Versions 3.3 and higher"
    }
  ]
}