{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Munich",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71924",
      "recalling_firm": "Brainlab AG",
      "address_1": "Olof-Palme-Strasse 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) including Puerto Rico; and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Latvia, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, and Vietnam.",
      "recall_number": "Z-0598-2016",
      "product_description": "Brainlab Cranial Navigation System: An Image Guided Surgery System / Stereotactic. Radiology Departments.    The BrainLAB Cranial IGS System is intended to be an intra-operative image guided localization system to enable minimally invasive surgery.",
      "product_quantity": "2528 units / 2119 consignees",
      "reason_for_recall": "Software Error: The effect of setup on overall navigation accuracy could potentially intensify small inaccuracies arising from individual steps of a complex navigation procedure that may cause an inaccurate display of instruments by the navigation system compared to the actual patient anatomy.",
      "recall_initiation_date": "20130422",
      "center_classification_date": "20160119",
      "termination_date": "20180417",
      "report_date": "20160127",
      "code_info": "1) Model 22035 , VVC NAVIGATION SOFTWARE CRANIAL 5.0 , Serial Numbers: 22035-103911-004 ,  22035-103953-001 ,  22035-103977-001 ,  22035-103982-002 ,  22035-103987-001 ,  22035-103988-001 ,  22035-104566-001 ;   2) Model 22040 , VV NAVIGATION SOFTWARE CRANIAL 4.0 , Serial Numbers:   22040/18820 ,  22040/21370 ,  22040/26909 ,  22040/36997 ,  22040-103675-006 ,  22040-103687-009 ,  22040-103710-001 ,  22040-103942-008 ,  22040-103974-011 ,  22040-104067-001 ,  22040-104072-001 ,  22040-104114-001 ,  22040-104148-001 ,  22040-104151-001 ,  22040-104505-001 ,  22040-104542-001 ,  22040-104545-001 ,  22040-104614-005 ,  22040-123772-001 ;   3) Model 22040-02 , VV NAV SW CRANIAL 7.8.1 (ANY PLATFORM) , Serial Number:  000000002600509001 ;   4) Model 22040-11 , UPDATE VV NAV SW CRANIAL 7.9.1 FROM 7.9 , Serial Numbers:   000000007665811001 ,  000000007801111004 ,  000000007801111005 ,  000000007801111006 ,  000000007801111007 ,  000000007801111008 ,  000000007801111009 ,  000000007808711004 ,  000000007808711007 ,  000000007808711008 ,  000000007837611010 ,  000000007847411007 ,  000000007865211004 ,  000000007865211005 ,  000000007942211004 ,  000000007942211008 ,  000000007942211010 ,  000000007950411001 ,  000000007950411004 ,  000000007950411007 ,  000000007950411010 ,  000000007956611005 ,  000000007956611010 ,  000000007956611017 ,  000000007965011001 ,  000000007965011006 ,  000000007965011007 ,  000000007980111001 ,  000000007980111002 ,  000000007980111003 ,  000000007980111004 ,  000000007980111005 ,  000000007980111006 ,  000000007980111007 ,  000000008000111003 ,  000000008000111006 ,  000000008000111007 ,  000000008012411001 ,  000000008012411003 ,  000000008012411008 ,  000000008026011002 ,  000000008026011003 ,  000000008036811002 ,  000000008036811006 ,  000000008036811007 ,  000000008036811008 ,  000000008036911002 ,  000000008036911004 ,  000000008036911005 ,  000000008046111002 ,  000000008046111005 ,  000000008073511003 ,  000000008073511007 ,  000000008086811001 ,  000000008106711001 ,  000000008114111004 ,  000000008129711004 ,  000000008129711005 ,  000000008129711006 ,  000000008133311001 ,  000000008133311002 ,  000000008143111002 ,  000000008143111003 ,  000000008143111004 ,  000000008157411005 ,  000000008157411007 ,  000000008167811005 ,  000000008167811008 ,  000000008167811012 ,  000000008167811014 ,  000000008167811015 ,  000000008167811016 ,  000000008180811001 ,  000000008180811008 ,  000000008180811009 ,  000000008180811010 ,  000000008180811011 ,  000000008181011002 ,  000000008208411002 ,  000000008230111001 ,  000000008230111002 ,  000000008230111009 ,  000000008230111013 ,  000000008230111015 ,  000000008230111022 ,  000000008230111024 ,  000000008230111025 ,  000000008230211001 ,  000000008230211002 ,  000000008230211008 ,  000000008230211017 ,  000000008230211020 ,  000000008230211025 ,  000000008230211030 ,  000000008230311006 ,  000000008230311014 ,  000000008230311015 ,  000000008230311018 ,  000000008230411003 ,  000000008230411005 ,  000000008230411009 ,  000000008230411010 ,  000000008230411012 ,  000000008230411016 ,  000000008230411021 ,  000000008230411024 ,  000000008230411029 ,  000000008230411030 ,  000000008230411031 ,  000000008230411032 ,  000000008242111002 ,  000000008242111003 ,  000000008242111005 ,  000000008242611001 ,  000000008242611002 ,  000000009168012001 ;   5) Model 22040B , VV NAVIGATION SOFTWARE CRANIAL7.5.0 , Serial Numbers:   05-48916-22040B ,  05-51803-22040B ,  05-52149-22040B ,  22040B/46911 ;   6) Model 22040C , VV NAVIGATION SOFTWARE CRANIAL 7.6.0 , Serial Numbers:   05-63659-22040C ,  05-63669-22040C ,  05-66754-22040C ,  06-79914-22040C ;   7) Model 22040D , VV NAVIGATION SW CRANIAL 7.7 , Serial Numbers:   06-84302-22040D ,  06-85635-22040D ,  06-88801-22040D ;   8) Model 22040E , VV NAVIGATION SW CRANIAL 7.8 , Serial Numbers:   06-101594-22040E ,  06-103819-22040E ,  06-108132-22040E ,  07-102189-22040E ,  07-106354-22040E ,  07-108464-22040E ,  07-109071-22040E ,  07-109217-22040E ,  07-110114-22040E ,  07-112255-22040E ,  07-112347-22040E ,  07-113858-22040E ,  07-114331-22040E ,  07-117343-22040E ,  07-118687-22040E ,  07-118695-22040E ,  07-119019-22040E ,  07-119599-22040E ,  07-120231-22040E ,  07-120842-22040E ,  07-122228-22040E ,  07-123321-22040E ,  07-124014-22040E ,  07-126723-22040E ,  07-128644-22040E ,  07-129197-22040E ,  07-129354-22040E ,  07-130356-22040E ;   9) Model 22040F , VV NAVIGATION SW CRANIAL 7.8.1 , Serial Numbers:   000000001242908001 ,  000000001275408001 ,  000000001305408001 ,  000000001346908001 ,  000000001386608001 ,  000000001402808001 ,  000000001506308001 ,  000000001516208001 ,  000000001548508001 ,  000000001611208001 ,  000000001644608001 ,  000000001716308001 ,  000000001730608001 ,  000000001832508001 ,  000000001832608001 ,  000000001867508001 ,  000000001869408001 ,  000000001924908001 ,  000000001943208001 ,  000000001944708001 ,  000000002054708001 ,  000000002082108001 ,  000000002179709001 ,  000000002352209001 ,  000000002401209001 ,  000000002540709001 ,  000000002540809001 ,  000000002565609001 ,  000000002565809001 ,  000000002610309001 ,  000000002921409001 ,  000000002945709001 ,  000000003057109001 ,  000000003171909001 ,  000000003331709001 ,  000000003433809001 ,  000000003446909001 ,  3540908001 ,  3557608001 ,  6039008001 ,  6110908001 ,  6303008001 ,  6406408001 ,  6594408001 ,  6655308001 ,  6757308001 ,  6983608001 ,  07-133796-22040F ,  07-134724-22040F ,  07-135062-22040F ,  07-135413-22040F ,  07-136068-22040F ,  07-136871-22040F ,  07-137864-22040F ,  07-138500-22040F ,  07-138954-22040F ,  07-138987-22040F ,  07-139009-22040F ,  07-139027-22040F ,  07-139037-22040F ,  07-139083-22040F ,  07-139412-22040F ,  07-139955-22040F ,  07-141837-22040F ,  07-144111-22040F ,  07-145211-22040F ,  07-146009-22040F ,  08-148434-22040F ,  08-149705-22040F ,  08-153066-22040F ,  08-154083-22040F ,  08-155770-22040F ,  08-156023-22040F ,  08-156226-22040F ,  08-156570-22040F ,  08-157148-22040F ,  08-157972-22040F ;   10) Model 22040H , VV NAVIGATION SW CRANIAL 7.9.1 , Serial Numbers:   000000004274913001 ,  000000008413811001 ;   11) Model 22043 , UPDATE VV NAV SWCRANIAL 6.0 TO 6.01 , Serial Numbers:   07-135238-22043 ,  22043/1194 ;   12) Model 22204-02 , KOLIBRI NAV SW CRAN 2.6.1 (ANY PLATFORM) , Serial Number:  000000003102809001 ;   13) Model 22204B , KOLIBRI NAVIGATION SW CRANIAL 2.5.0 , Serial Number:  06-86477-22204B ;   14) Model 22204C , KOLIBRI NAVIGATION SW CRANIAL 2.6.0 , Serial Numbers:   06-100679-22204C ,  06-104941-22204C ,  07-112890-22204C ,  07-114630-22204C ,  07-116967-22204C ,  07-123214-22204C ,  07-128684-22204C ,  07-129055-22204C ,  07-133191-22204C ,  07-133265-22204C ,  15) Model 22204D , KOLIBRI NAVIGATION SW CRANIAL 2.6.1 , Serial Numbers:   000000001120208001 ,  000000001125508001 ,  000000001178108001 ,  000000001291308001 ,  000000001343608001 ,  000000001362808001 ,  000000001671208001 ,  000000002019108001 ,  000000002108909001 ,  000000002219709001 ,  000000002308509001 ,  000000002494309001 ,  6259008001 ,  6381208001 ,  6662908001 ,  6898208001 ,  07-134523-22204D ,  07-135013-22204D ,  07-135592-22204D ,  07-141778-22204D ,  07-142020-22204D ,  07-149290-22204D ,  08-148598-22204D ,  08-152171-22204D ,  08-153392-22204D ,  08-153694-22204D ,  08-154014-22204D ,  08-154473-22204D ,  08-157303-22204D ,  08-157669-22204D ;   16) Model 22204E , KOLIBRI NAVIGATION SW CRANIAL 2.7 , Serial Number:  000000003211909001 ;   17) Model 22207 , CRANIAL ESSENTIAL NAV SW 1.0 , Serial Numbers:  000000002787409001 ,   000000002796409001 ,  000000002811509001 ,  000000002867009001 ,  000000002906409001 ,  000000002930409001 ,  000000003058009001 ,  000000003220409001 ,  000000003224109001 ,  000000003546309001 ,  000000003576409001 ,  000000003680909001 ,  000000003738509001 ,  000000003982409001 ,  000000003990809001 ,  000000004154809001 ,  000000004494710001 ,  000000004535510001 ,  000000004622210001 ,  000000004636910001 ,  000000004804710001 ,  000000004910010001 ,  000000004923210001 ,  000000005074410001 ,  000000005078110001 ,  000000005100710001 ,  000000005153810001 ,  000000005353810001 ,  000000005354110001 ,  000000005427710001 ,  000000005453410001 ,  000000005607210001 ,  000000005651510001 ,  000000005668710001 ,  000000005710010001 ,  000000005736910001 ,  000000005838110001 ,  000000006077510001 ,  000000006148710001 ,  000000006179010001 ,  000000006306910001 ,  000000006339910001 ,  000000006363010001 ,  000000006726411001 ,  000000006949311001 ,  000000007053911001 ,  000000007076211001 ,  000000007146511001 ,  000000007172711001 ,  000000007302311001 ,  000000007418911001 ;   18) Model 22207A , CRANIAL ESSENTIAL NAVIGATION SW 1.0.1 , Serial Number:  000000007678411001 ;   19) Model 22208 , CRANIAL UNLIMITED NAV SW 1.0 , Serial Numbers:  000000002958709001 , 000000003668009001 ,  000000004574810001 ,  000000004694310001 ,  000000005096210001 ,  000000005278010001 ,  000000005728710001 ,  000000006119610002 ,  000000006119610003 ,  000000006119610004 ,  000000006119610005 ,  000000006271410001 ,  000000006654910001 ,  000000006694611001 ,  000000006786911001 ;   20) Model 22208-11 , UPDATE CRANIAL UNLIMITED 1.0.1 FROM 1.0 , Serial Numbers:  000000007719311001 , 000000007757911001 ,  000000007775111001 ,  000000007801311002 ,  000000007801311005 ,  000000007808811004 ,  000000007814811001 ,  000000007814811002 ,  000000007830611001 ,  000000007830611003 ,  000000007837811001 ,  000000007837811003 ,  000000007837811004 ,  000000007847611002 ,  000000007856311001 ,  000000007856311003 ,  000000007856311005 ,  000000007856311006 ,  000000007865011004 ,  000000007865011005 ,  000000007865011006 ,  000000007891211001 ,  000000007891211002 ,  000000007891211003 ,  000000007891211004 ,  000000007891211005 ,  000000007927911001 ,  000000007927911002 ,  000000007927911003 ,  000000007927911005 ,  000000007927911009 ,  000000007928011001 ,  000000007928011002 ,  000000007928011003 ,  000000007942511002 ,  000000007942511003 ,  000000007942511005 ,  000000007942511006 ,  000000007942511013 ,  000000007942511014 ,  000000007942511016 ,  000000007942511018 ,  000000007950711002 ,  000000007950711004 ,  000000007950711005 ,  000000007950711009 ,  000000007950711011 ,  000000007957111012 ,  000000007980311001 ,  000000007980311002 ,  000000007980311004 ,  000000007980311006 ,  000000007980311010 ,  000000007980311011 ,  000000007980311012 ,  000000008012711002 ,  000000008012711004 ,  000000008025711001 ,  000000008025711002 ,  000000008025711003 ,  000000008037311001 ,  000000008037411001 ,  000000008037411002 ,  000000008037411003 ,  000000008037411005 ,  000000008045711002 ,  000000008045711004 ,  000000008059111001 ,  000000008059211001 ,  000000008074511004 ,  000000008074511006 ,  000000008074511008 ,  000000008074511009 ,  000000008074511011 ,  000000008074511013 ,  000000008074511014 ,  000000008074511016 ,  000000008074611001 ,  000000008074611002 ,  000000008087011002 ,  000000008087011004 ,  000000008087011008 ,  000000008087011009 ,  000000008087011010 ,  000000008087011011 ,  000000008087211001 ,  000000008106811001 ,  000000008106811003 ,  000000008106811004 ,  000000008114411002 ,  000000008114411005 ,  000000008114411009 ,  000000008114411012 ,  000000008129911003 ,  000000008129911004 ,  000000008129911005 ,  000000008129911006 ,  000000008129911007 ,  000000008129911008 ,  000000008134911002 ,  000000008134911003 ,  000000008134911004 ,  000000008134911005 ,  000000008134911006 ,  000000008134911007 ,  000000008134911008 ,  000000008134911009 ,  000000008134911010 ,  000000008134911011 ,  000000008134911016 ,  000000008143611001 ,  000000008143611002 ,  000000008143611004 ,  000000008143611006 ,  000000008143611007 ,  000000008143611008 ,  000000008143611010 ,  000000008143611011 ,  000000008143611012 ,  000000008143611016 ,  000000008143611018 ,  000000008143611019 ,  000000008143611020 ,  000000008143611024 ,  000000008143611026 ,  000000008143611027 ,  000000008150011001 ,  000000008150011003 ,  000000008150011004 ,  000000008150011005 ,  000000008150011006 ,  000000008157811001 ,  000000008157811002 ,  000000008157811003 ,  000000008157811007 ,  000000008157811008 ,  000000008157811009 ,  000000008157811010 ,  000000008157811012 ,  000000008168311002 ,  000000008168311005 ,  000000008168311006 ,  000000008168311007 ,  000000008168311008 ,  000000008168311009 ,  000000008168311010 ,  000000008168311011 ,  000000008181511003 ,  000000008181511004 ,  000000008208911001 ,  000000008208911004 ,  000000008208911005 ,  000000008222511001 ,  000000008222511002 ,  000000008222511003 ,  000000008222511005 ,  000000008222511006 ,  000000008222511008 ,  000000008222511009 ,  000000008222511010 ,  000000008222511011 ,  000000008222511013 ,  000000008222511014 ,  000000008230711001 ,  000000008230711004 ,  000000008230711005 ,  000000008230711006 ,  000000008230711007 ,  000000008230711008 ,  000000008242711001 ,  000000008242711005 ,  000000008242711006 ,  000000008242711008 ,  000000008242711009 ,  000000008242711010 ,  000000008242711011 ,  000000008242711012 ,  000000008242711013 ,  000000008242711015 ,  000000008242711016 ,  000000008242711019 ,  000000008242711020 ,  000000008242711023 ,  000000008242711024 ,  000000008242711025 ,  000000008242711026 ,  000000008242711027 ,  000000008242711028 ,  000000008242711029 ,  000000008242711031 ,  000000008242711033 ,  000000008242711034 ,  000000008242711035 ,  000000008242711036 ,  000000008242711038 ,  000000008242711041 ,  000000008242711042 ,  000000008242711043 ,  000000008242711044 ,  000000008277511001 ,  000000008352911001 ,  000000009167712001 ,  000000009167812001 ,  000000009167912001 ,  000000009239412001 ,  000000009359612001 ,  000000009408712001 ,  000000009408812001 ;   21) Model 22208A , CRANIAL UNL NAVIGATION SW 1.0.1 , Serial Numbers:  000000001287412001 ,  000000003539513001 ,  000000004457713001 ,  000000008293111001 ,  000000008471411001 ,  000000008489611001 ;   22) Model 22214 , CRANIAL NAVIGATION SW 2.1 , Serial Numbers:  000000007467211001 ,  000000007489111001 ,  000000007494611001 ,  000000007495311001 ,  000000007505911001 ,  000000007516811001 ,  000000007516911001 ,  000000007545411001 ,  000000007572811001 ,  000000007576811001 ,  000000007578311001 ,  000000007604811001 ,  000000007605711001 ,  000000007605811001 ,  000000007605911001 ,  000000007683111001 ,  000000007695511001 ,  000000007735211001 ,  000000007773611001 ,  000000007777511001 ,  000000007802011001 ,  000000007840911001 ,  000000007872611001 ,  000000007899411001 ,  000000007936111001 ,  000000007954111001 ,  000000007960411001 ,  000000007973911001 ,  000000008063011001 ,  000000008067111001 ,  000000008100511001 ,  000000008111511001 ,  000000008165111001 ,  000000008189511001 ,  000000008210511001 ,  000000008245011001 ,  000000008245111001 ,  000000008260011001 ,  000000008290111001 ,  000000008370011001 ,  000000008431811001 ,  000000008441811001 ;   23) Model 22214-01 , UPDATE CRANIAL 2.1.1 FROM 2.1 , Serial Numbers:  000000008574211001 ,  000000008741911001 ,  000000008861011001 ,  000000008980412001 ,  000000009122512001 ,  000000009199812001 ,  000000009237012001 ,  000000009283712001 ,  000000009283812001 ,  000000009283912001 ,  000000009284012001 ,  000000009284112001 ,  000000009316312001 ,  000000009363312001 ,  000000009422712001 ,  000000009437312001 ,  000000009437412001 ,  000000009437512001 ,  000000009437612001 ,  000000009481412001 ,  000000009861512001 ;   24) Model 22214-01A , UPD CRANIAL 2.1.1 W/ DATA LOAD FROM 2.1 , Serial Numbers:  000000000239512001 , 000000000246112001 ,  000000000605512001 ,  000000000605814001 ,  000000000797312001 ,  000000000815712001 ,  000000001221412001 ,  000000001866612001 ,  000000001866712001 ,  000000002099012001 ,  000000002698513001 ,  000000004417413001 ,  000000009530014001 ;   25) Model 22214-01B , UPD CRANIAL 2.1.1 W/ DATA LOAD FROM 2.1 , Serial Numbers:  000000000051714001 ,  000000005104313001 ,  000000006146515001 ,  000000006146615001 ,  000000006146715001 ,  000000006146815001 ,  000000007622015001 ,  000000009289414001 ;   26) Model 22214A , CRANIAL NAVIGATION SW 2.1.1 , Serial Numbers:  000000000052612001 ,  000000000052712001 ,  000000000055912001 ,  000000000157012001 ,  000000008553411001 ,  000000008598111001 ,  000000008615711001 ,  000000008624811001 ,  000000008674211001 ,  000000008700211001 ,  000000008714811001 ,  000000008715911001 ,  000000008766711001 ,  000000008795711001 ,  000000008844011001 ,  000000008848411001 ,  000000008861911001 ,  000000008874011001 ,  000000008902511001 ,  000000008947212001 ,  000000008968312001 ,  000000009013812001 ,  000000009022712001 ,  000000009043912001 ,  000000009053712001 ,  000000009106912001 ,  000000009133112001 ,  000000009133612001 ,  000000009134012001 ,  000000009168512001 ,  000000009306012001 ,  000000009342012001 ,  000000009382912001 ,  000000009395512001 ,  000000009401412001 ,  000000009461012001 ,  000000009505812001 ,  000000009551212001 ,  000000009559212001 ,  000000009573412001 ,  000000009647912001 ,  000000009654412001 ,  000000009654612001 ,  000000009745412001 ,  000000009763112001 ,  000000009787112001 ,  000000009788212001 ,  000000009816112001 ,  000000009816212001 ,  000000009819512001 ,  000000009867212001 ,  000000009885512001 ,  000000009926512001 ,  000000009944112001 ,  000000009951112001 ,  000000009964612001 ;   27) Model 22214B , CRANIAL NAVIGATION SW 2.1.1 , Serial Numbers:  000000000240112001 ,  000000000240212001 ,  000000000254712001 ,  000000000353512001 ,  000000000385712001 ,  000000000385812001 ,  000000000405512001 ,  000000000406612001 ,  000000000496312001 ,  000000000503212001 ,  000000000520312001 ,  000000000546512001 ,  000000000574812001 ,  000000000587912001 ,  000000000615712001 ,  000000000623212001 ,  000000000681112001 ,  000000000709312001 ,  000000000709412001 ,  000000000712112001 ,  000000000725312001 ,  000000000726312001 ,  000000000760512001 ,  000000000792812001 ,  000000000813012001 ,  000000000848912001 ,  000000000871612001 ,  000000000959612001 ,  000000001014212001 ,  000000001132612001 ,  000000001168512001 ,  000000001196612001 ,  000000001213712001 ,  000000001235712001 ,  000000001324112001 ,  000000001341012001 ,  000000001354912001 ,  000000001362412001 ,  000000001382112001 ,  000000001384512001 ,  000000001634612001 ,  000000001634712001 ,  000000001635012001 ,  000000001645912001 ,  000000001695412001 ,  000000001705512001 ,  000000001752412001 ,  000000001782812001 ,  000000001783612001 ,  000000001804712001 ,  000000001823412001 ,  000000001850912001 ,  000000001873512001 ,  000000001882512001 ,  000000001936612001 ,  000000001938712001 ,  000000001938812001 ,  000000002045412001 ,  000000002059512001 ,  000000002104112001 ,  000000002179412001 ,  000000002240312001 ,  000000002323212001 ,  000000002346312001 ,  000000002360212001 ,  000000002368012001 ,  000000002370912001 ,  000000002371012001 ,  000000002380812001 ,  000000002401612001 ,  000000002421712001 ,  000000002422812001 ,  000000002424312001 ,  000000002433512001 ,  000000002444812001 ,  000000002508012001 ,  000000002580412001 ,  000000002610013001 ,  000000002615112001 ,  000000002662913001 ,  000000002682613001 ,  000000002705213001 ,  000000002714613001 ,  000000002788513001 ,  000000002815713001 ,  000000002967013001 ,  000000002988213001 ,  000000003009213001 ,  000000003053413001 ,  000000003059013001 ,  000000003067613001 ,  000000003071813001 ,  000000003158813001 ,  000000003223813001 ,  000000003273213001 ,  000000003286913001 ,  000000003297813001 ,  000000003300613001 ,  000000003302213001 ,  000000003332213001 ,  000000003356613001 ,  000000003379513001 ,  000000003382713001 ,  000000003409613001 ,  000000003465613001 ,  000000003489313001 ,  000000003633013001 ,  000000003633113001 ,  000000003643713001 ;   28) Model 22214C , CRANIAL NAVIGATION SW 2.1 , Serial Numbers:  000000000034814001 ,  000000000034914001 ,  000000000059214001 ,  000000000266114001 ,  000000001118214001 ,  000000001320614001 ,  000000001496814001 ,  000000001855614001 ,  000000001933214001 ,  000000001949014001 ,  000000001950314001 ,  000000002549514001 ,  000000003648014001 ,  000000003671614001 ,  000000003816414001 ,  000000003818813001 ,  000000003818913001 ,  000000003819013001 ,  000000003848013001 ,  000000003877714001 ,  000000003896213001 ,  000000003953413001 ,  000000003954013001 ,  000000003978813001 ,  000000003978913001 ,  000000003979013001 ,  000000003979113001 ,  000000003979213001 ,  000000004058613001 ,  000000004058713001 ,  000000004062414001 ,  000000004069313001 ,  000000004069413001 ,  000000004089813001 ,  000000004153413001 ,  000000004161613001 ,  000000004334713001 ,  000000004406013001 ,  000000004439913001 ,  000000004468913001 ,  000000004469013001 ,  000000004469113001 ,  000000004486313001 ,  000000004486413001 ,  000000004486513001 ,  000000004486613001 ,  000000004486713001 ,  000000004504313001 ,  000000004538314001 ,  000000004545113001 ,  000000004612413001 ,  000000004663813001 ,  000000004663913001 ,  000000004709113001 ,  000000004783814001 ,  000000004859613001 ,  000000004859713001 ,  000000004887513001 ,  000000004988313001 ,  000000005006513001 ,  000000005066313001 ,  000000005153913001 ,  000000005173713001 ,  000000005173813001 ,  000000005173913001 ,  000000005274015001 ,  000000005275513001 ,  000000005429413001 ,  000000005447613001 ,  000000005504713001 ,  000000005555513001 ,  000000005571213001 ,  000000005614113001 ,  000000005614213001 ,  000000005632713001 ,  000000005655613001 ,  000000005734813001 ,  000000005749813001 ,  000000005783213001 ,  000000005783313001 ,  000000005809913001 ,  000000005843413001 ,  000000005858813001 ,  000000005933613001 ,  000000005933713001 ,  000000006082113001 ,  000000006082213001 ,  000000006089613001 ,  000000006110313001 ,  000000006133613001 ,  000000006260615001 ,  000000006264013001 ,  000000006388913001 ,  000000006389013001 ,  000000006389113001 ,  000000006389213001 ,  000000006389313001 ,  000000006389413001 ,  000000006389513001 ,  000000006421115001 ,  000000006501613001 ,  000000006642913001 ,  000000006678013001 ,  000000006727313001 ,  000000006787613001 ,  000000006876713001 ,  000000006914715001 ,  000000006963413001 ,  000000006963513001 ,  000000006965115001 ,  000000006986013001 ,  000000006986113001 ,  000000007070313001 ,  000000007070413001 ,  000000007104213001 ,  000000007122413001 ,  000000007133613001 ,  000000007133713001 ,  000000007152313001 ,  000000007183813001 ,  000000007222213001 ,  000000007275913001 ,  000000007276013001 ,  000000007298113001 ,  000000007335813001 ,  000000007386913001 ,  000000007505713001 ,  000000007582914001 ,  000000007597214001 ,  000000007615214001 ,  000000007679014001 ,  000000007695914001 ,  000000007739514001 ,  000000007748814001 ,  000000007778815001 ,  000000007778915001 ,  000000007779015001 ,  000000007779115001 ,  000000007779215001 ,  000000007845614001 ,  000000007855114001 ,  000000007864215001 ,  000000007864315001 ,  000000007864415001 ,  000000007864515001 ,  000000007864615001 ,  000000007864715001 ,  000000007887514001 ,  000000007887614001 ,  000000007931114001 ,  000000007952814001 ,  000000008108414001 ,  000000008146114001 ,  000000008169214001 ,  000000008214914001 ,  000000008252814001 ,  000000008252914001 ,  000000008253014001 ,  000000008253114001 ,  000000008270014001 ,  000000008288414001 ,  000000008288514001 ,  000000008345715001 ,  000000008352914001 ,  000000008387414001 ,  000000008412014001 ,  000000008442114001 ,  000000008442214001 ,  000000008451814001 ,  000000008465514001 ,  000000008528214001 ,  000000008569214001 ,  000000008576714001 ,  000000008679414001 ,  000000008795114001 ,  000000008921914001 ,  000000008930314001 ,  000000008952114001 ,  000000008968714001 ,  000000008968814001 ,  000000008968914001 ,  000000008984714001 ,  000000009086514001 ,  000000009148214001 ,  000000009192514001 ,  000000009212114001 ,  000000009379914001 ,  000000009440114001 ,  000000009468614001 ,  000000009506614001 ,  000000009533714001 ,  000000009659414001 ,  000000009688014001 ,  000000009780614001 ,  000000009780714001 ,  000000009990514001 ,  000000009990614001 ;   29) Model 71000 , VV SOFTWARE UPGRADE CRANIAL 3.X TO 6.01 , Serial Numbers:  71000/21628 ,  71000-103744-001 ,  71000-103745-002 ,  05-49480-71005 ;   30) Model 71006 , UPG VV CRANIAL 6.01 TO CRANIAL 7.0.X , Serial Numbers:  000000002899209001 ,  000000003143809001 ,  000000004297109001 ,  6032908001 ,  05-50560-71006 ,  05-52155-71006 ,  08-148612-71006 ;   31) Model 71008 , UPG VV CRANIAL 7.0 TO CRANIAL 7.5 , Serial Number:  000000004384709001 ;   32) Model 71011 , UPG CRA ESSENTIAL 1.0 TO UNLIMIMITED 1.0 , Serial Numbers:  000000004150309001 ,  000000004441110001 ;   33) Model 71011A , UPGRADE CRANIAL ESS 1.0.1 TO UNL 1.0.1 , Serial Numbers:  000000007678511001 ,  000000008548411001 ;   34) Model 71012 , UPGRADE VV CRANIAL 7.9 TO UNLIMITED 1.0 , Serial Number:  000000004337709001 ;   35) Model 71012A , UPGRADE VV CRANIAL 7.9.1 TO UNL 1.0.1 , Serial Numbers:  000000005629513001 ,  000000005629613001 ,  000000005629713001 ,  000000005629913001 ,  000000005676213001 ,  000000005676313001 ,  000000005769113001 ,  000000005809113001 ;   36) Model 71013 , UPG KOLIBRI CRANIAL 2.7 TO ESSENTIAL 1.0 , Serial Number:  000000002811109001 ;   37) Model 71017 , UPG TO CRAN 2.x FROM CRAN UNLIMITED 1.0 , Serial Numbers:  000000000005412001 ,   000000000060612001 ,  000000000108312001 ,  000000000108412001 ,  000000000108512001 ,  000000000251912001 ,  000000000316012001 ,  000000000388112001 ,  000000000721512001 ,  000000000722012001 ,  000000001002812001 ,  000000001015412001 ,  000000001114112001 ,  000000001114212001 ,  000000001279512001 ,  000000001291612001 ,  000000001311912001 ,  000000001336712001 ,  000000001358812001 ,  000000001511812001 ,  000000001652012001 ,  000000001757512001 ,  000000002034512001 ,  000000002777513001 ,  000000002788013001 ,  000000002911513001 ,  000000002958813001 ,  000000003056913001 ,  000000003067913001 ,  000000003094513001 ,  000000003151913001 ,  000000003187813001 ,  000000003242113001 ,  000000003538413001 ,  000000007545311001 ,  000000007587511001 ,  000000007678011001 ,  000000007725111001 ,  000000007888511001 ,  000000008050311001 ,  000000008084411001 ,  000000008231511001 ,  000000008264411001 ,  000000008283511001 ,  000000008377711001 ,  000000008444611001 ,  000000008529511001 ,  000000008579211001 ,  000000008579311001 ,  000000008579411001 ,  000000008676811001 ,  000000008676911002 ,  000000008855711001 ,  000000009001312001 ,  000000009002812001 ,  000000009079812001 ,  000000009146112001 ,  000000009162712001 ,  000000009168712001 ,  000000009187212001 ,  000000009208512001 ,  000000009411312001 ,  000000009470312001 ,  000000009470412001 ,  000000009484312001 ,  000000009599812001 ,  000000009632012001 ,  000000009652512001 ,  000000009760312001 ,  000000009840212001 ,  000000009931412001 ,  000000009931512001 ;   38) Model 71017A , UPG TO CRAN 2.x FROM CRAN UNLIMITED 1.0 , Serial Numbers:  000000001088214001 ,  000000002036614001 ,  000000002632314001 ,  000000002855614001 ,  000000003113714001 ,  000000003113814001 ,  000000003531414001 ,  000000003603714001 ,  000000003753014001 ,  000000003753114001 ,  000000003924713001 ,  000000003979114001 ,  000000004041513001 ,  000000004111013001 ,  000000004152813001 ,  000000004152913001 ,  000000004321813001 ,  000000004321913001 ,  000000004322013001 ,  000000004519013001 ,  000000004591513001 ,  000000004813913001 ,  000000004866313001 ,  000000004866413001 ,  000000005172113001 ,  000000005650913001 ,  000000006367713001 ,  000000006367813001 ,  000000006662513001 ,  000000006752013001 ,  000000006894213001 ,  000000006894313001 ,  000000006909613001 ,  000000007311115001 ,  000000007365613001 ,  000000007432713001 ,  000000007514713001 ,  000000007624313001 ,  000000007657214001 ,  000000007704815001 ,  000000007765214001 ,  000000008248514001 ,  000000008508414001 ,  000000009319814001 ,  000000009378914001 ,  000000009466914001 ;   39) Model 71051 , UPD VV CRANIAL 7.5.X TO 7.6.X , Serial Number:  000000002382609001 ;   40) Model 71053 , UPD VV CRANIAL 7.6.X TO CRANIAL7.7.X , Serial Number:  06-88103-71053 ;   41) Model 71056 , UPD VV CRANIAL 7.7.X TO CRANIAL7.8.X , Serial Numbers:  000000001425108001 ,  000000001462008001 ,  000000003671509001 ,  000000009258012001 ;   42) Model 71063 , UPG KOL CRANIAL 2.5.X TO CRANIAL 2.6.X , Serial Number:  07-111143-71063 ;   43) Model 71067 , UPDATE VV NAV SW CRANIAL 7.8 TO 7.9 , Serial Numbers:  000000003270709001 ,  000000003428909001 ,  000000004577510001 ;   44) Model 71067A , UPDATE VV NAV SW CRANIAL 7.8 TO 7.9.1 , Serial Numbers:  000000000253112001 ,  000000000452712001 ,  000000001446312001 ,  000000001533312001 ,  000000001708612001 ,  000000002492312001 ,  000000004691813001 ,  000000009147012001 ,  000000009537312001 ,  000000009641812001 ,  000000009653312001 ;   45) Model 71069 , UPD KOLIBRI NAV SW CRANIAL 2.6 TO 2.7 , Serial Numbers:  000000003189109001 ,   000000003189209001 ,  000000003200109001 ,  000000003288809001 ,  000000003498009001 ,  000000004323109001 ,  000000004576710001 ,  000000006333310001 ,  000000006333410001 ;   46) Model 71075 , UPG CRA ESS 1.0 FROM ANY PREV. KOL SW , Serial Number:  000000005026310001 ;   47) Model 71076 , UPG VV CRA 7.0 FROM ANY PREV. VV CRA SW , Serial Number:  000000007742711001 ;   48) Model 71077 , UPG CRA UNL 1.0 FROM VV CRA 7.X , Serial Numbers:  000000005437910001 ,  000000005587010001 ,  000000006351510001 ;   49) Model 71077A , UPGRADE CRANIAL UNL 1.0.1 FROM VV 7.X , Serial Numbers:  000000002989913001 ,   000000002990013001 ,  000000007676711001 ,  000000008849411001 ,  000000009150612001 ;   50) Model 71086 , UPG TO CRANIAL NAV SW 2.X FROM ANY PREV , Serial Numbers:  000000000110212001 ,  000000000268812001 ,  000000000268912001 ,  000000000269112001 ,  000000000306812001 ,  000000000317512001 ,  000000000420912001 ,  000000000564112001 ,  000000000627612001 ,  000000000776012001 ,  000000000837012001 ,  000000001030812001 ,  000000001157712001 ,  000000001489112001 ,  000000001510112001 ,  000000001694212001 ,  000000002957213001 ,  000000004002213001 ,  000000007487311001 ,  000000007494511001 ,  000000007561011001 ,  000000007584811001 ,  000000007754211001 ,  000000007766011001 ,  000000007832511001 ,  000000007896911001 ,  000000008147311001 ,  000000008292611001 ,  000000008312911001 ,  000000008537511001 ,  000000008610611001 ,  000000008623211001 ,  000000008626411001 ,  000000008655511001 ,  000000008796311001 ,  000000008891611001 ,  000000009025612001 ,  000000009232112001 ,  000000009301712001 ,  000000009388712001 ,  000000009537112001 ,  000000009737712001 ,  000000009737712002 ,  000000009737712003 ,  000000009856112001 ,  000000009900812001 ,  000000009904912001 ,  000000009988512001 ,  000000009991312001 ,  000000009991412001 ;   51) Model 71086A , UPG TO CRANIAL NAV SW 2.X FROM ANY PREV , Serial Numbers:  000000002036914001 ,  000000002344114001 ,  000000004470113001 ,  000000004507013001 ,  000000004507113001 ,  000000004918215001 ,  000000005094113001 ,  000000005658713001 ,  000000006162913001 ,  000000007326413001 ,  000000007513413001 ,  000000007870114001 ,  000000008813214001 ,  000000009479914001 ;   52) Model 71088 , UPGR TO CRANIAL NAV SW 2.X 11-02-18JUG2 , Serial Numbers:  000000007578911001 ,  000000007626611001 ;   53) Model 71093 , GEN UPG CRANIAL NAV SW FOR CURVE , Serial Numbers:  000000000350812001 ,  000000000619012001 ,  000000000645412001 ,  000000000676712001 ,  000000000957112001 ,  000000001091112001 ,  000000001726912001 ,  000000001727012001 ,  000000002541512001 ,  000000002583812001 ,  000000002765713001 ,  000000003236413001 ,  000000003237213001 ,  000000009412112001 ,  000000009626512001 ,  000000009669712001 ,  000000009700112001 ;   54) Model 71093A , GEN UPG CRANIAL NAV SW FOR CURVE/KICK , Serial Numbers:  000000000494614001 ,  000000001210114001 ,  000000001239214001 ,  000000002580914001 ,  000000003754914001 ,  000000003828313001 ,  000000004239713001 ,  000000004315913001 ,  000000004591713001 ,  000000004606413001 ,  000000004785513001 ,  000000004881313001 ,  000000005430113001 ,  000000005602113001 ,  000000005602213001 ,  000000005658813001 ,  000000005658913001 ,  000000005669013001 ,  000000005982813001 ,  00000",
      "more_code_info": "0006044113001 ,  000000006044213001 ,  000000006395013001 ,  000000006395113001 ,  000000006501913001 ,  000000006643113001 ,  000000006738213001 ,  000000006878013001 ,  000000007095113001 ,  000000007095213001 ,  000000007223613001 ,  000000007248815001 ,  000000007377813001 ,  000000007487813001 ,  000000007641214001 ,  000000007641314001 ,  000000007819514001 ,  000000007844914001 ,  000000007931414001 ,  000000008021914001 ,  000000008260014001 ,  000000008296214001 ,  000000008319314001 ,  000000008454614001 ,  000000008500114001 ,  000000008795614001 ,  000000008860614001 ,  000000008860714001 ,  000000008878214001 ,  000000009480014001 ;   55) Model 71096 , UPG TO CRAN 2.x CURVE FROM CRAN UNL 1.0 , Serial Numbers:  000000000239112001 ,   000000000239212001 ,  000000000239312001 ,  000000000330112001 ,  000000000789212001 ,  000000000789312001 ,  000000000998012001 ,  000000001081912001 ,  000000001186712001 ,  000000001231712001 ,  000000001352512001 ,  000000001620512001 ,  000000001706812001 ,  000000001749112001 ,  000000001895012001 ,  000000002056312001 ,  000000002098912001 ,  000000002235012001 ,  000000002235112001 ,  000000002360412001 ,  000000002696813001 ,  000000002777313001 ,  000000002777413001 ,  000000002878213001 ,  000000003057013001 ,  000000003261813001 ,  000000003314513001 ,  000000003450613001 ,  000000003450713001 ,  000000003450813001 ,  000000003602013001 ,  000000003609213001 ,  000000003609313001 ,  000000003609413001 ,  000000003666113001 ,  000000009232212001 ,  000000009389912001 ,  000000009530112001 ,  000000009764012001 ;   54) Model 71096A , UPG TO CRAN 2.x CURVE FROM CRAN UNL 1.0 , Serial Numbers:  000000001634614001 , 000000002290314001 ,  000000003924813001 ,  000000003960613001 ,  000000003960713001 ,  000000004303913001 ,  000000004533313001 ,  000000005397113001 ,  000000005503813001 ,  000000005535413001 ,  000000006656213001 ,  000000006894413001 ,  000000006894513001 ,  000000006941613001 ,  000000007196713001 ,  000000007365713001 ,  000000007595914001 ,  000000007927114001 ,  000000008479114001 ,  000000008543914001 ,  000000009189414001 ,  000000009189514001 ,  000000009485314001 ,  000000009564714001 ,  000000009564814001 ,  000000009908214001 ,  000000009987114001 ;   55) Model 71097 , UPG TO CRAN 2.X CURVE FROM CRAN 2.X STD , Serial Numbers:  000000001620412001 ,  000000001662512001 ,  000000003087513001 ,  000000003756313001 ,  000000008548011001 ;   56) Model 71097A , UPG TO CRAN 2.X CURVE FROM CRAN 2.X STD , Serial Numbers:  000000001887414001 ,  000000002536114001 ,  000000004403613001 ,  000000005779713001 ,  000000006658513001 ,  000000007168513001 ,  000000007181513001 ,  000000007418613001 ,  000000007707214001 ,  000000008465014001"
    }
  ]
}