  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "2 Dundee Park Dr",
      "reason_for_recall": "The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard.  As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.",
      "address_2": "",
      "product_quantity": "37 total devices",
      "code_info": "Serial Numbers: 650074 650085  650075 650086  650076 650087  650079 650088  650080 650089  650081 650090  650082 650091  650083 650092  650094  650095  650096  650098  650099  650100  650104  650055  650058  650062  650066  650067  650069  650070  650073",
      "center_classification_date": "20191203",
      "distribution_pattern": "The products were distributed to the following US states:  KY, MO, NY, OH, PA",
      "state": "MA",
      "product_description": "EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1",
      "report_date": "20191211",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "NDDd Medical Technologies, Inc.",
      "recall_number": "Z-0597-2020",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "83992",
      "termination_date": "20200505",
      "more_code_info": "",
      "recall_initiation_date": "20170815",
      "postal_code": "01810-3735",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}