  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81560",
      "recalling_firm": "COVIDIEN LLC",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and foreign distribution to Canada.",
      "recall_number": "Z-0597-2019",
      "product_description": "Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36\" (91 cm), Product Code 8888264911    Product Usage:  The Salem Sump Tubes are intended for decompression and drainage of stomach contents following injury, illness, or surgery in patients with intestinal obstruction, and to prevent the distention of the stomach.",
      "product_quantity": "1522",
      "reason_for_recall": "This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available.",
      "recall_initiation_date": "20181031",
      "center_classification_date": "20181211",
      "termination_date": "20200526",
      "report_date": "20181219",
      "code_info": "1812115564"
    }
  ]
}