      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72751",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including:  AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.",
      "recall_number": "Z-0597-2016",
      "product_description": "Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.",
      "product_quantity": "73 units",
      "reason_for_recall": "Potential post-processing software issue when using tabcard \"4D\" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.",
      "recall_initiation_date": "20151118",
      "center_classification_date": "20160106",
      "termination_date": "20161205",
      "report_date": "20160113",
      "code_info": "Software Version: SW VD-10E"
    }
  ]
}