{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plano",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81461",
      "recalling_firm": "St. Jude Medical, Inc.",
      "address_1": "6901 Preston Rd",
      "address_2": "N/A",
      "postal_code": "75024-2508",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, MD, MN, MO, MT, NC, NJ, NY, OH, PA, and WA. International distribution in the countries of  Argentina, Australia, Austria, Belgium, Brazil, Colombia, Czech Republic, Finland, France, Germany, Hungary, Italy, Mexico, Poland, Russian Federation, Saudi Arabia, South Korea, Spain, Sweden, Taiwan, Turkey, and United Kingdom",
      "recall_number": "Z-0593-2019",
      "product_description": "St. Jude Medical InfinityTM DBS System 8CH Directional Lead, 30 cm, 0.5 REF 6170 - STERILE EO Rx ONLY  CE 0086 0123  St. Jude Medical  Plano, TX    St. Jude Medical InfinityTM DBS System 8CH Directional Lead, 40 cm, 0.5 REF 6172 STERILE EO Rx ONLY  St. Jude Medical  Plano, TX    Instructions For Use: Lead and Extension Kits  for Deep Brain Stimulation Systems Clinician's Manual  ST. JUDE MEDICAL - Product Usage: St. Jude Medical Deep Brain Stimulation (DBS)leads are intended to deliver stimulation to target areas in the brain. DBS extensions are intended to connect the leads to implantable pulse generators (IPGs). St. Jude Medical DBS leads are designed for introduction into the brain using standard stereotactic neurosurgical techniques. DBS system delivers electrical stimulation to a precisely targeted area in the brain. Leads are implanted in the brain and are connected to extensions, which are passed under the skin and are connected to the neurostimulator. Leads for the St. Jude Medical Infinity DBS System feature electrodes on a stiff distal end with an inactive lead tip. The proximal end of the lead contains contact bands that correspond with each of the distal electrodes and an inactive band that functions as a contact for a setscrew when connecting to a compatible extension. The 8-channel leads contain cylindrical and segmented electrodes. The segmented electrodes can be activated independently to focus stimulation in one direction to help target desired neurological structures. As stated in Clinician manual (ARTEN600008305 Rev A), leads materials which are intended to come into contact with tissue are Platinum-iridium and polycarbonate urethane. In addition, drawing for the Stim Tip, 1-3-3-1 Directional DBS lead requires that the electrode material shall be 90/10 Platinum Iridium.",
      "product_quantity": "N/A",
      "reason_for_recall": "The most proximal unsegmented electrode of the Deep Brain Stimulation leads, may be constructed with MP35N instead of the required 90/10 platinum-iridium.",
      "recall_initiation_date": "20181011",
      "center_classification_date": "20181210",
      "termination_date": "20220913",
      "report_date": "20181219",
      "code_info": "For Model 6170ANS:  UDI/GTIN - 05415067020635 / 05415067030306  US:  Serial Numbers    Lot Number  17277045             6351121    OUS:  17145460 6244627  17149644 6262443  17149652 6262443  17149625 6262443  17149630 6262443  17149631 6262443  17149647 6262443  17149650 6262443  17149653 6262443  17168532 6272675  17168533 6272675  17168534 6272675  17168541 6272675  17145439 6244627  17145459 6244627  17145461 6244627  17145463 6244627  17113441 6212647  17113442 6212647  17113445 6212647  17145451 6244627  17113436 6212647  17145454 6244627  17145458 6244627  17145438 6244627  17145446 6244627  17145448 6244627  17145453 6244627  17145457 6244627  17145462 6244627  17166635 6272679  17113438 6212647  17149635 6262443  17149641 6262443  17149655 6262443  17166627 6272679  17149629 6262443  17149633 6262443  17149637 6262443  17149645 6262443  17149646 6262443  17149626 6262443  17149654 6262443  17145447 6244627  17145456 6244627  17215431 6299732  17215434 6299732  17215455 6299732  17113440 6212647  17113447 6212647  17149624 6262443  17149640 6262443  17166633 6272679  17149634 6262443  17215427 6299732  17215438 6299732  17215440 6299732  17215444 6299732  17145441 6244627  17145444 6244627  17149628 6262443  17149649 6262443  17166624 6272679  17166631 6272679  17166634 6272679  17166640 6272679  17166623 6272679  17149642 6262443  17149658 6262443  17166625 6272679  17166629 6272679  17166638 6272679  17166641 6272679  17168535 6272675  17233826 6320021  17233834 6320021  17233846 6320021  17233856 6320021  17168524 6272675  17168525 6272675  17168526 6272675  17168529 6272675  17168531 6272675  17168538 6272675  17168543 6272675  17168545 6272675  17168547 6272675  17168528 6272675  17168537 6272675  17168546 6272675  17168548 6272675  17233830 6320021  17233838 6320021  17233848 6320021  17233850 6320021  17233854 6320021  17233855 6320021  17145437 6244627  17233835 6320021  17168521 6272675  17113434 6212647  17149638 6262443  17166622 6272679  17145440 6244627  17145455 6244627  17149632 6262443  17215428 6299732  17215429 6299732  17215433 6299732  17215435 6299732  17215437 6299732  17215439 6299732  17215441 6299732  17215443 6299732  17215446 6299732  17215447 6299732  17215448 6299732  17215449 6299732  17215450 6299732  17215453 6299732  17233842 6320021  17233860 6320021  17233861 6320021  17233865 6320021  17168522 6272675  17168527 6272675  17168530 6272675  17233852 6320021  17233867 6320021  17166630 6272679  17168536 6272675  17168542 6272675  17215432 6299732  17233828 6320021  17233829 6320021  17233833 6320021  17233839 6320021  17233841 6320021  17233843 6320021  17233849 6320021  17233853 6320021  17233862 6320021  17233868 6320021  17262603 6320023  17262619 6320023  17277049 6351121  17277053 6351121  17277059 6351121  17233840 6320021  17262604 6320023  17262608 6320023  17262613 6320023  17262618 6320023  17262620 6320023  17262624 6320023  17262635 6320023  17145445 6244627  17233869 6320021  17262622 6320023  17262628 6320023  17149636 6262443  17233832 6320021  17113437 6212647  17113446 6212647  17233847 6320021  17233857 6320021  17233859 6320021  17233866 6320021  17262605 6320023  17262611 6320023  17262614 6320023  17262615 6320023  17262617 6320023  17262631 6320023  17262632 6320023  17277042 6351121  17277064 6351121  17277065 6351121  17166628 6272679  17215426 6299732  17215451 6299732  17277050 6351121  17233837 6320021  17233851 6320021  17215452 6299732  17277068 6351121  17168539 6272675  17168540 6272675  17233858 6320021  17233863 6320021  17233864 6320021  17145443 6244627  17215454 6299732  17262634 6320023  17262606 6320023  17262609 6320023  17262616 6320023  17262630 6320023  17277046 6351121  17277051 6351121  17149656 6262443  17215445 6299732  17277044 6351121  17277063 6351121  17215442 6299732  17277052 6351121  17277055 6351121  17277057 6351121  17277066 6351121  17309334 6371714  17309342 6371714  17309331 6371714  17233827 6320021  17262621 6320023  17149627 6262443  17149657 6262443  17233844 6320021  17233845 6320021  17262629 6320023  17277056 6351121  17309332 6371714  17309354 6371714  17309325 6371714  17309355 6371714  17309328 6371714  17277060 6351121  17277061 6351121  17309343 6371714  17309349 6371714  17113443 6212647  17145436 6244627  17145449 6244627  17262633 6320023  17309357 6371714  17309323 6371714  17309326 6371714  17309336 6371714  17309352 6371714  17309329 6371714  17309340 6371714  17309345 6371714  17309346 6371714  17309348 6371714  17309333 6371714  17309358 6371714  17309327 6371714  17309338 6371714  17309351 6371714  17309359 6371714  17233836 6320021  17338051 6391795  17338055 6391795  17338061 6391795  17338065 6391795  17338066 6391795  17309324 6371714  17309330 6371714  17309363 6371714  17233831 6320021  17338056 6391795  17338062 6391795  17113435 6212647  17113444 6212647  17145435 6244627  17145442 6244627  17145450 6244627  17149639 6262443  17149643 6262443  17149648 6262443  17149651 6262443  17166626 6272679  17166636 6272679  17166637 6272679  17166642 6272679  17168523 6272675  17168544 6272675  17215430 6299732  17215436 6299732  17262607 6320023  17262610 6320023  17262612 6320023  17262623 6320023  17262625 6320023  17262626 6320023  17262627 6320023  17277043 6351121  17277047 6351121  17277048 6351121  17277054 6351121  17277058 6351121  17277062 6351121  17277069 6351121  17309335 6371714  17309341 6371714  17309344 6371714  17309350 6371714  17309356 6371714  17309362 6371714  17338050 6391795  17338052 6391795  17338053 6391795  17338054 6391795  17338057 6391795  17338058 6391795  17338059 6391795  17338060 6391795  17338064 6391795    For Model 6172ANS:  UDI/GTIN - 05415067020680 / 05415067030320    US:  Serial Numbers   Lot Number  17101268 6212648  17101271 6212648  17101279 6212648  17101281 6212648  17101282 6212648  17101285 6212648  17101288 6212648  17101292 6212648  17101294 6212648  17101297 6212648  17101298 6212648  17101301 6212648  17101302 6212648  17101315 6212648  17101316 6212648  17101318 6212648  17101322 6212648  17101307 6212648  17101296 6212648  17101293 6212648  17101304 6212648  17145561 6244642  17145539 6244642  17145542 6244642  17145549 6244642  17158024 6262459  17158016 6262459  17158029 6262459  17174775 6272706  17174765 6272706  17174762 6272706  17174784 6272706  17174785 6272706  17174758 6272706  17205551 6299729  17205550 6299729  17205560 6299729  17174789 6272706  17205558 6299729  17205559 6299729  17205574 6299729  17231245 6310618  17231254 6310618  17231409 6320018  17231413 6320018  17231419 6320018  17252228 6320020  17252230 6320020  17252237 6320020  17252250 6320020  17252251 6320020  17252256 6320020  17252226 6320020  17252236 6320020  17252252 6320020  17264770 6339283  17264774 6339283  17264775 6339283  17252215 6320020  17252212 6320020  17252208 6320020  17252222 6320020  17252214 6320020  17264765 6339283  17252247 6320020  17279683 6351127  17279697 6351127  17279698 6351127  17279699 6351127  17327582 6381181  17327588 6381181  17380720 6413167  17380718 6413167  17380727 6413167  17380732 6413167  17380706 6413167  17380831 6413195  17380820 6413195  17380825 6413195  17380827 6413195  17380834 6413195  17380839 6413195  17380840 6413195  17401440 6434604  17401441 6434604  17380712 6413167  17401437 6434604  17401438 6434604  17401443 6434604  17401446 6434604  17401447 6434604  17401450 6434604  17401456 6434604  17401459 6434604  17401458 6434604  17434708 6456727  17434714 6456727  17454542 6478084  17454555 6478084  17437569 6467125  17437576 6467125  17437579 6467125  17437581 6467125  17437583 6467125  17536074 6523556  17536081 6523556  17536084 6523556  17555101 6523557  17437556 6467125  17437561 6467125  17437572 6467125  17437580 6467125  17437590 6467125  17437591 6467125  17437592 6467125  17437594 6467125  17437595 6467125  17454531 6478084  17454539 6478084  17454544 6478084  17454547 6478084  17454549 6478084  17327580 6381181  17454538 6478084  17454548 6478084  17454551 6478084  17454557 6478084  17434689 6456727  17437555 6467125  17437558 6467125  17437575 6467125  17437578 6467125  17437597 6467125  17437557 6467125  17457048 6478085  17457054 6478085  17457080 6478085  17457071 6478085  17457055 6478085  17457069 6478085  17555106 6523557  17555109 6523557  17231420 6320018  17457049 6478085  17457057 6478085  17457060 6478085  17457077 6478085  17327571 6381181  17457050 6478085  17457053 6478085  17457056 6478085  17457070 6478085  17555099 6523557  17078876 6197962  17078907 6197962  17101269 6212648  17101270 6212648  17101272 6212648  17101274 6212648  17101275 6212648  17101276 6212648  17101278 6212648  17101280 6212648  17101283 6212648  17101284 6212648  17101286 6212648  17101287 6212648  17101289 6212648  17101290 6212648  17101291 6212648  17101295 6212648  17101299 6212648  17101303 6212648  17101305 6212648  17101306 6212648  17101308 6212648  17101309 6212648  17101310 6212648  17101311 6212648  17101312 6212648  17101313 6212648  17101314 6212648  17101317 6212648  17101319 6212648  17101320 6212648  17101321 6212648  17101323 6212648  17101324 6212648  17145535 6244642  17145537 6244642  17145538 6244642  17145543 6244642  17145544 6244642  17145545 6244642  17145546 6244642  17145547 6244642  17145548 6244642  17145550 6244642  17145551 6244642  17145553 6244642  17145554 6244642  17145556 6244642  17145558 6244642  17145562 6244642  17158011 6262459  17158012 6262459  17158013 6262459  17158014 6262459  17158018 6262459  17158019 6262459  17158020 6262459  17158021 6262459  17158023 6262459  17158025 6262459  17158027 6262459  17158031 6262459  17158034 6262459  17158039 6262459  17158041 6262459  17174759 6272706  17174760 6272706  17174761 6272706  17174763 6272706  17174764 6272706  17174766 6272706  17174767 6272706  17174768 6272706  17174769 6272706  17174770 6272706  17174771 6272706  17174772 6272706  17174773 6272706  17174774 6272706  17174776 6272706  17174777 6272706  17174778 6272706  17174779 6272706  17174780 6272706  17174781 6272706  17174782 6272706  17174783 6272706  17174786 6272706  17174788 6272706  17205552 6299729  17205553 6299729  17205554 6299729  17205555 6299729  17205556 6299729  17205557 6299729  17205561 6299729  17205562 6299729  17205565 6299729  17205566 6299729  17205567 6299729  17205570 6299729  17205572 6299729  17205573 6299729  17205575 6299729  17205576 6299729  17205577 6299729  17205578 6299729  17205579 6299729  17231228 6310618  17231229 6310618  17231230 6310618  17231231 6310618  17231232 6310618  17231233 6310618  17231235 6310618  17231237 6310618  17231238 6310618  17231239 6310618  17231242 6310618  17231244 6310618  17231249 6310618  17231251 6310618  17231253 6310618  17231411 6320018  17231412 6320018  17231415 6320018  17231416 6320018  17231417 6320018  17231418 6320018  17231421 6320018  17231422 6320018  17231423 6320018  17231425 6320018  17231426 6320018  17231427 6320018  17231428 6320018  17231429 6320018  17231430 6320018  17231431 6320018  17252204 6320020  17252205 6320020  17252206 6320020  17252207 6320020  17252209 6320020  17252210 6320020  17252211 6320020  17252213 6320020  17252216 6320020  17252217 6320020  17252218 6320020  17252219 6320020  17252221 6320020  17252223 6320020  17252224 6320020  17252227 6320020  17252229 6320020  17252231 6320020  17252232 6320020  17252234 6320020  17252235 6320020  17252239 6320020  17252240 6320020  17252241 6320020  17252244 6320020  17252245 6320020  17252246 6320020  17252248 6320020  17252253 6320020  17252254 6320020  17252255 6320020  17264759 6339283  17264760 6339283  17264761 6339283  17264762 6339283  17264763 6339283  17264764 6339283  17264768 6339283  17264772 6339283  17264773 6339283  17264776 6339283  17264777 6339283  17264778 6339283  17264779 6339283  17279674 6351127  17279675 6351127  17279676 6351127  17279678 6351127  17279680 6351127  17279681 6351127  17279682 6351127  17279684 6351127  17279685 6351127  17279686 6351127  17279688 6351127  17279689 6351127  17279691 6351127  17279692 6351127  17279693 6351127  17279694 6351127  17279695 6351127  17279696 6351127  17279700 6351127  17290508 6359088  17327572 6381181  17327574 6381181  17327575 6381181  17327576 6381181  17327577 6381181  17327578 6381181  17327579 6381181  17327584 6381181  17327585 6381181  17327586 6381181  17327587 6381181  17327589 6381181  17327591 6381181  17327592 6381181  17327594 6381181  17327595 6381181  17380696 6413167  17380698 6413167  17380699 6413167  17380701 6413167  17380702 6413167  17380703 6413167  17380705 6413167  17380707 6413167  17380708 6413167  17380709 6413167  17380710 6413167  17380713 6413167  17380714 6413167  17380715 6413167  17380717 6413167  17380719 6413167  17380721 6413167  17380722 6413167  17380723 6413167  17380724 6413167  17380725 6413167  17380726 6413167  17380728 6413167  17380729 6413167  17380730 6413167  17380731 6413167  17380818 6413195  17380828 6413195  17380829 6413195  17380832 6413195  17380833 6413195  17380835 6413195  17380836 6413195  17380837 6413195  17380838 6413195  17380841 6413195  17380843 6413195  17380844 6413195  17380846 6413195  17380847 6413195  17380848 6413195  17380850 6413195  17380851 6413195  17380853 6413195  17401434 6434604  17401436 6434604  17401439 6434604  17401442 6434604  17401444 6434604  17401448 6434604  17401451 6434604  17401453 6434604  17401457 6434604  17401460 6434604  17401462 6434604  17401463 6434604  17434690 6456727  17434691 6456727  17434692 6456727  17434693 6456727  17434694 6456727  17434695 6456727  17434696 6456727  17434697 6456727  17434698 6456727  17434699 6456727  17434701 6456727  17434702 6456727  17434703 6456727  17434704 6456727  17434705 6456727  17434706 6456727  17434707 6456727  17434709 6456727  17434710 6456727  17434711 6456727  17434712 6456727  17434713 6456727  17434715 6456727  17434716 6456727  17434717 6456727  17434718 6456727  17434719 6456727  17437551 6467125  17437553 6467125  17437559 6467125  17437560 6467125  17437562 6467125  17437563 6467125  17437564 6467125  17437565 6467125  17437567 6467125  17437568 6467125  17437571 6467125  17437573 6467125  17437574 6467125  17437577 6467125  17437582 6467125  17437585 6467125  17437586 6467125  17437587 6467125  17437589 6467125  17437593 6467125  17437596 6467125  17454529 6478084  17454530 6478084  17454534 6478084  17454535 6478084  17454536 6478084  17454540 6478084  17454543 6478084  17454545 6478084  17454546 6478084  17454550 6478084  17454552 6478084  17454553 6478084  17454556 6478084  17454558 6478084  17454561 6478084  17454562 6478084  17457052 6478085  17457058 6478085  17457062 6478085  17457064 6478085  17457073 6478085  17457075 6478085  17457081 6478085  17536055 6523556  17536059 6523556  17536061 6523556  17536064 6523556  17536067 6523556  17536069 6523556  17536070 6523556  17536072 6523556  17536073 6523556  17536075 6523556  17536078 6523556  17536082 6523556  17555077 6523557  17555080 6523557  17555082 6523557  17555084 6523557  17555085 6523557  17555086 6523557  17555088 6523557  17555089 6523557  17555090 6523557  17555091 6523557  17555093 6523557  17555097 6523557  17555098 6523557  17555100 6523557  17555103 6523557    OUS:  17103771 6212651  17103774 6212651  17103778 6212651  17103781 6212651  17103785 6212651  17103787 6212651  17103789 6212651  17086343 6197963  17086348 6197963  17086336 6197963  17086338 6197963  17086339 6197963  17086340 6197963  17086351 6197963  17086352 6197963  17103776 6212651  17145486 6244631  17145492 6244631  17145482 6244631  17145483 6244631  17145494 6244631  17145496 6244631  17174087 6272696  17174088 6272696  17174089 6272696  17174090 6272696  17174091 6272696  17174092 6272696  17174093 6272696  17174094 6272696  17174095 6272696  17174096 6272696  17174097 6272696  17174098 6272696  17174099 6272696  17174100 6272696  17174101 6272696  17174102 6272696  17174106 6272696  17174107 6272696  17174108 6272696  17174109 6272696  17174110 6272696  17174111 6272696  17174114 6272696  17174115 6272696  17174116 6272696  17174117 6272696  17086332 6197963  17086345 6197963  17103784 6212651  17103790 6212651  17103792 6212651  17103794 6212651  17103795 6212651  17103797 6212651  17086337 6197963  17086342 6197963  17086355 6197963  17086357 6197963  17086331 6197963  17086341 6197963  17086354 6197963  17103772 6212651  17103773 6212651  17103775 6212651  17103777 6212651  17103780 6212651  17103783 6212651  17103786 6212651  17103788 6212651  17103791 6212651  17103793 6212651  17103796 6212651  17103799 6212651  17145480 6244631  17145481 6244631  17145485 6244631  17145487 6244631  17145488 6244631  17145489 6244631  17145491 6244631  17145497 6244631  17145499 6244631  17145500 6244631  17145502 6244631  17145503 6244631  17145504 6244631  17145505 6244631  17145506 6244631  17145507 6244631  17086346 6197963  17086347 6197963  17086353 6197963  17103779 6212651  17103770 6212651  17103782 6212651  17103798 6212651  17228992 6310617  17228997 6310617  17229001 6310617  17229002 6310617  17229005 6310617  17229007 6310617  17229008 6310617  17229012 6310617  17229013 6310617  17229014 6310617  17229019 6310617  17229025 6310617  17229026 6310617  17229027 6310617  17242379 6320025  17242380 6320025  17242381 6320025  17242382 6320025  17242383 6320025  17242384 6320025  17242385 6320025  17242386 6320025  17242387 6320025  17242388 6320025  17242389 6320025  17242390 6320025  17242391 6320025  17242394 6320025  17145493 6244631  17279644 6351123  17279661 6351123  17086334 6197963  17086358 6197963  17279648 6351123  17279663 6351123  17279664 6351123  17279652 6351123  17279666 6351123  17279647 6351123  17279656 6351123  17279662 6351123  17279667 6351123  17279670 6351123  17316097 6371718  17316110 6371718  17324588 6381171  17324595 6381171  17324598 6381171  17324601 6381171  17324603 6381171  17324610 6381171  17324612 6381171  17324614 6381171  17324618 6381171  17324619 6381171  17256660 6320026  17228991 6310617  17228993 6310617  17290485 6359088  17290501 6359088  17290479 6359088  17290503 6359088  17289985 6359086  17289986 6359086  17289987 6359086  17289998 6359086  17279657 6351123  17279659 6351123  17279655 6351123  17316112 6371718  17319979 6371719  17316103 6371718  17256670 6320026  17086333 6197963  17103800 6212651  17289992 6359086  17290505 6359088  17290000 6359086  17290492 6359088  17289984 6359086  17289995 6359086  17290002 6359086  17290009 6359086  17319975 6371719  17319980 6371719  17319981 6371719  17319982 6371719  17319984 6371719  17319987 6371719  17319992 6371719  17319996 6371719  17319974 6371719  17319983 6371719  17319985 6371719  17256667 6320026  17279654 6351123  17316096 6371718  17316101 6371718  17316106 6371718  17228998 6310617  17228999 6310617  17290008 6359086  17086349 6197963  17316104 6371718  17078887 6197962  17078888 6197962  17316093 6371718  17375742 6402167  17375745 6402167  17375746 6402167  17375754 6402167  17375756 6402167  17375758 6402167  17375760 6402167  17375763 6402167  17319976 6371719  17319989 6371719  17078926 6197962  17290010 6359086  17290498 6359088  17383884 6413196  17383911 6413196  17383925 6413196  17256669 6320026  17324613 6381171  17289996 6359086  17324591 6381171  17383879 6413196  17383914 6413196  17078868 6197962  17078934 6197962  17086344 6197963  17290004 6359086  17290007 6359086  17383912 6413196  17319993 6371719  17324590 6381171  17078901 6197962  17078924 6197962  17316111 6371718  17324585 6381171  17289990 6359086  17078885 6197962  17078927 6197962  17078882 6197962  17078889 6197962  17289989 6359086  17383894 6413196  17383906 6413196  17383915 6413196  17383923 6413196  17383886 6413196  17383897 6413196  17383917 6413196  17279649 6351123  17383901 6413196  17078892 6197962  17078910 6197962  17290006 6359086  17078883 6197962  17078912 6197962  17319990 6371719  17324596 6381171  17383896 6413196  17383921 6413196  17319977 6371719  17279651 6351123  17279653 6351123  17375751 6402167  17290482 6359088  17290491 6359088  17375738 6402167  17375759 6402167  17290481 6359088  17290500 6359088  17279668 6351123  17290003 6359086  17316115 6371718  17078886 6197962  17078909 6197962  17078925 6197962  17290478 6359088  17290506 6359088  17375740 6402167  17375741 6402167  17375752 6402167  17375755 6402167  17375761 6402167  17375762 6402167  17375765 6402167  17375768 6402167  17078880 6197962  17316098 6371718  17375749 6402167  17375757 6402167  17383877 6413196  17383898 6413196  17383903 6413196  17383905 6413196  17078872 6197962  17078875 6197962  17078878 6197962  17078891 6197962  17078893 6197962  17078894 6197962  17078899 6197962  17078906 6197962  17078921 6197962  17078928 6197962  17086335 6197963  17086350 6197963  17086356 6197963  17145484 6244631  17145490 6244631  17145498 6244631  17145501 6244631  17174112 6272696  17174113 6272696  17174118 6272696  17174119 6272696  17174120 6272696  17174121 6272696  17228994 6310617  17228995 6310617  17228996 6310617  17229000 6310617  17229003 6310617  17229004 6310617  17229006 6310617  17229009 6310617  17229010 6310617  17229011 6310617  17229015 6310617  17229016 6310617  17229017 6310617  17229018 6310617  17229020 6310617  17229021 6310617  17229023 6310617  17229024 6310617  17229028 6310617  17256658 6320026  17256659 6320026  17256661 6320026  17256662 6320026  17256663 6320026  17256664 6320026  17256665 6320026  17256666 6320026  17256668 6320026  17256671 6320026  17279646 6351123  17279650 6351123  17279658 6351123  17279660 6351123  17279665 6351123  17289991 6359086  17289994 6359086  17289997 6359086  17290011 6359086  17290012 6359086  17290013 6359086  17290487 6359088  17290496 6359088  17316095 6371718  17316100 6371718  17316102 6371718  17316108 6371718  17316109 6371718  17316113 6371718  17316119 6371718  17316121 6371718  17316124 6371718  17319978 6371719  17319986 6371719  17319988 6371719  17319991 6371719  17319995 6371719  17319997 6371719  17324586 6381171  17324589 6381171  17324592 6381171  17324593 6381171  17324594 6381171  17324597 6381171  17324599 6381171  17324600 6381171  17324604 6381171  17324606 6381171  17324607 6381171  17324608 6381171  17324609 6381171  17324611 6381171  17324615 6381171  17324616 6381171  17324617 6381171  17375739 6402167  17375744 6402167  17375750 6402167  17375753 6402167  17375766 6402167  17383878 6413196  17383881 6413196  17383887 6413196  17383895 6413196  17383907 6413196  17383908 6413196  17383910 6413196  17383919 6413196"
    }
  ]
}