      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89376",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was nationwide, including Puerto Rico and the Virgin Islands.  There was also government and military distribution.  Foreign distribution was made to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaidjan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Republic of Congo, Croatia, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Equatorial Guinea, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrhyzstan, Latvia, Lebanon, Libyan Arabic Jamahiriya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Mali, Malta, Mauritania, Mexico, Republic of Moldova, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Nordic-HQ, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Swaziland, Sweden,  Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Zambia, and Zimbabwe.",
      "recall_number": "Z-0592-2022",
      "product_description": "GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Pioneer System, nuclear magnetic resonance imaging system.",
      "product_quantity": "17,228 devices total in this field correction",
      "reason_for_recall": "The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.",
      "recall_initiation_date": "20211223",
      "center_classification_date": "20220209",
      "report_date": "20220216",
      "code_info": "All systems."
    }
  ]
}