  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sunnyvale",
      "address_1": "904 E Caribbean Dr",
      "reason_for_recall": "Cepheid has received reports that some of our customers have experienced a higher number of  temperature drift errors than would typically be expected when using this test and its intended use  specimen types. Investigation of the issue also revealed an occasional probability of false  negative results which remains consistent with package insert claims.",
      "address_2": "",
      "product_quantity": "3,586 kits (49,400 units)",
      "code_info": "Lot numbers: 25718, 25719, 25725, and 25727.",
      "center_classification_date": "20180212",
      "distribution_pattern": "Worldwide Distribution -- USA, Puerto Rico, and Haiti.",
      "state": "CA",
      "product_description": "Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075, Catalog GXCT/NG-10 Batch 1000056156 and Catalog GXCT/NG-10 Batch 1000056157.    Is a qualitative in vitro real-time PCR test.",
      "report_date": "20180221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Cepheid",
      "recall_number": "Z-0592-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78910",
      "termination_date": "20180213",
      "more_code_info": "",
      "recall_initiation_date": "20171009",
      "postal_code": "94089-1189",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}