{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84212",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to US states of  AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.",
      "recall_number": "Z-0590-2020",
      "product_description": "CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577     For diagnostic and interventional vascular procedures.",
      "product_quantity": "58",
      "reason_for_recall": "A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).",
      "recall_initiation_date": "20191108",
      "center_classification_date": "20191209",
      "termination_date": "20210201",
      "report_date": "20191218",
      "code_info": "Lot Numbers 9950878, 9953506  UDI: (01)00827002515772(17)221508(10)9950878  UDI: (01)00827002515772(17)221608(10)9953506",
      "more_code_info": ""
    }
  ]
}