{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86832",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavagen 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0589-2021",
      "product_description": "RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0)",
      "product_quantity": "124",
      "reason_for_recall": "Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.",
      "recall_initiation_date": "20201111",
      "center_classification_date": "20201215",
      "termination_date": "20230509",
      "report_date": "20201223",
      "code_info": "Software version numbers:  4.0.0.14 ,  4.0.3.4 ,  4.5.1.14 ,  4.7.2.5 ,  4.7.3.13 ,  4.7.4.4 ,  4.7.5.4 ,  4.7.6.7 ,  4.9.0.42 ,;  5.0.1.11 ,  5.0.2.35 ,  5.0.3.17 ,  6.0.0.24 ,  6.1.1.2 , 6.2.0.7 ,  6.3.0.6 ,  7.0.0.19 ,  8.0.0.61 ,  8.0.1.10 ,  8.1.0.47 ,  8.1.1.8   8.1.2.5 ,  9.0.0.113 , 9.1.0.933 ,  9.2.0.483 ,  10.0.0.1154"
    }
  ]
}