  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Komorniki",
      "state": "N/A",
      "country": "Poland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81562",
      "recalling_firm": "ARJOHUNTLEIGH POLSKA Sp. z.o.o.",
      "address_1": "Ui. Ks. Piotra Wawrzyniaka 2",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US nationwide distribution in the state of  WV.    OUS:  Information not provided due to the product being manufactured OUS and imported to US",
      "recall_number": "Z-0589-2019",
      "product_description": "ARJOHUNTLEIGH GETINGE GROUP  Flites  Passive Clip Sling - Product Usage: The Arjo Passive Clip Sling is a product intended for assisted transfer of patient/resident with limited ability to move. Designed to be used for one patient during hospital stay and assists the caregiver with everyday patient handling routines, such as bed and chair related transfers.  REF MFA1000M-L-L1  RM0255_7",
      "product_quantity": "10",
      "reason_for_recall": "Inspection process at the manufacturer found nonwoven material was used in specific lots of product that may break wile in use during the patient/resident transfer.  Patient/resident might fall on the floor.  IN consequence, the patient might sustain a major trauma, head injury - damage to tissue that is irreversible or may lead to death.",
      "recall_initiation_date": "20181022",
      "center_classification_date": "20181207",
      "termination_date": "20210519",
      "report_date": "20181219",
      "code_info": "Model - MFA1000M-L-L1  Lot: DHD1301721  UPN 40679252100332"
    }
  ]
}