  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72652",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Drive",
      "address_2": "N/A",
      "postal_code": "07470",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "United States Nationwide Distribution",
      "recall_number": "Z-0589-2016",
      "product_description": "Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W    Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.",
      "product_quantity": "100 Units (distributed in United States)",
      "reason_for_recall": "Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.",
      "recall_initiation_date": "20151113",
      "center_classification_date": "20160105",
      "termination_date": "20170407",
      "report_date": "20160113",
      "code_info": "HCU 30 Base Unit 100 - 120 V, Model Number 704629/70103.4371  HCU 30 Base Unit 200 - 240 V, Model Number 704631/70103.4642  HCU 30 Base Unit Canadian/US, Model Number 705501/70103.4653  Heater Cooler Unit 30 with Serial Number <93101480    The affected power supply boards have serial numbers less than 107006001."
    }
  ]
}