{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Cloud",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89401",
      "recalling_firm": "Microbiologics Inc",
      "address_1": "200 Cooper Ave N",
      "address_2": "N/A",
      "postal_code": "56303-4440",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan,  Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.",
      "recall_number": "Z-0585-2022",
      "product_description": "KWIK-STIK, Quality control kit for culture media, sold as:     a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299    b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299",
      "product_quantity": "1,975 units",
      "reason_for_recall": "Tests returning incorrect susceptible results.  High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.",
      "recall_initiation_date": "20200218",
      "center_classification_date": "20220208",
      "termination_date": "20220711",
      "report_date": "20220216",
      "code_info": "a. Catalog Number 0959P, UDI: 20845357023210, Lot Number (Expiration Date): 959-94-3 (04/30/2020), 959-95-2 (09/30/2020), 959-96-1 (12/31/2020), 959-96-4 (12/31/2021), 959-97-4 (03/31/2021), 959-97-5 (03/31/2021);    b. Catalog Number 0959K, UDI: 30845357023224, Lot Number (Expiration Date): 959-94-2 (04/30/2020),  959-95-1 (09/30/2020), 959-95-5 (09/30/2020), 959-96-3 (12/31/2020), 959-97-1 (03/31/2021), 959-97-6 (03/31/2021);"
    }
  ]
}