{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "5643 Plymouth Rd",
      "reason_for_recall": "Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may  not discover the incompatibility issue until the",
      "address_2": "",
      "product_quantity": "sold in boxes of ten electrodes: 116 boxes; 1160 units total",
      "code_info": "All Lots.    Catalog / Part Number C100AC-PHILIPS",
      "center_classification_date": "20141223",
      "distribution_pattern": "Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of:  Thailand and South Korea.",
      "state": "MI",
      "product_description": "ADULT/CHILD Radiotransparent Electrode, Part number C100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED  \"Comparable to Philips Medical HeartStart Electrodes Plug Style Connector\", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.",
      "report_date": "20141231",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Heart Sync, Inc.",
      "recall_number": "Z-0584-2015",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "69875",
      "termination_date": "20151124",
      "more_code_info": "",
      "recall_initiation_date": "20141111",
      "postal_code": "48105-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}