  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Cloud",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89391",
      "recalling_firm": "Microbiologics Inc",
      "address_1": "200 Cooper Ave N",
      "address_2": "N/A",
      "postal_code": "56303-4440",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MS, MT, NC, ND, NE, NM, NY, OH, OK, OR, PR, SC, TN, TX, VA, VT, WA, WV and the countries of AUSTRALIA, AUSTRIA, BAHRAIN, BELARUS, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, FRANCE, GEORGIA, GERMANY, GUATAMALA, HONG KONG, INDIA, INDONESIA, IRELAND, ITALY, JAPAN, JORDAN, LATVIA, MEXICO, NETHERLANDS, PERU, PHILIPPINES, POLAND, ROMANIA, SAUDI ARABIA, SOUTH AFRICA, SWITZERLAND, UNITED ARAB EMIRATES, VIETNAM.",
      "recall_number": "Z-0582-2022",
      "product_description": "KWIK-STIK(TM), Streptococcus pneumoniae derived from ATCC(R) 49619(TM), packaged as follows:      a. KWIK-STIK(TM) 2 Pack;      b. KWIK-STIK(TM) 6 Pack.      Each KWIK-STIK unit contains a lyophilized microorganism pellet, an ampoule of hydrating fluid, and an inoculating swab. Each device is sealed within a laminated pouch that contains a desiccant to prevent adverse moisture accumulation. KWIK-STIK microorganisms are 3 passages or fewer from the reference culture and are guaranteed to recover when processed using the recommended media and incubation requirements. Available in packs of 2 or 6.",
      "product_quantity": "271 units total",
      "reason_for_recall": "Potential contamination with S. epidermidis E. coli and S. warneri.",
      "recall_initiation_date": "20201001",
      "center_classification_date": "20220208",
      "termination_date": "20220707",
      "report_date": "20220216",
      "code_info": "a. Catalog Number: 0947P; Lot Number: 947-126-2, 947-126-4; UDI: 20845357022947.    b. Catalog Number: 0947K; Lot Number: 947-126-3; UDI: 30845357022951"
    }
  ]
}