      "status": "Terminated",
      "city": "Miami Lakes",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81609",
      "recalling_firm": "Heartware, Inc.",
      "address_1": "14400 Nw 60th Ave",
      "address_2": "N/A",
      "postal_code": "33014-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US nationwide distribution in the state of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI. Countries of Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Estonia,   Finland, France, Germany,  Greece, Hong Kong, India, Italy, Latvia, Lebanon, Netherlands, Poland, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-0582-2019",
      "product_description": "Medtronic HeartWare Ventricular Assist HVAD System Battery Charger Model 1600US - Product Usage: The HeartWare Ventricular Assist System (HVAD) is indicated in the United States (US) and European Union (EU) for use as a bridge to cardiac transplantation (BTT) as well as an alternative to transplantation as destination therapy (DT) for patients who are at risk of death from refractory end-stage left ventricular heart failure. The HVAD Pump Controller is a microprocessor unit that controls and manages HeartWare System operation. It sends power and operating signals to the blood pump and collects information from the pump. The controller requires two power sources for safe operation: either two batteries, or one battery and an AC adapter or DC adapter. While active, patients will typically use two batteries. While relaxing or sleeping, patients should use power from an electrical outlet (AC adapter) because it provides power for an unlimited period of time. The HeartWare Battery Charger is part of the HeartWare Ventricular Assist System (HVAD). Patients receive 4 total batteries with their system along with a battery charger and are instructed to have spare, fully charged batteries always available. The battery charger can charge up to 4 batteries at a time using 4 charging bays. Each battery slides into a bay and is connected to the battery charger.",
      "product_quantity": "2031",
      "reason_for_recall": "HVAD Battery Charger units manufactured with wrong inductors.",
      "recall_initiation_date": "20181119",
      "center_classification_date": "20181207",
      "termination_date": "20200721",
      "report_date": "20181219",
      "code_info": "BCH000403, BCH000405, BCH000406, BCH000415, BCH000420, BCH000422, BCH000423, BCH000426, BCH000428, BCH000433,  BCH000435, BCH000440, BCH000442, BCH000446, BCH000451,  BCH000456, BCH000459, BCH000464, BCH000465, BCH000470,  BCH000471, BCH000472, BCH000475, BCH000484, BCH000485, BCH000486, BCH000490, BCH000493, BCH000495, BCH000500, BCH000502, BCH000503, BCH000507, BCH000514, BCH000523, BCH000524, BCH000525, BCH000526, BCH000527, BCH000528,  BCH000530, BCH000531, BCH000534, BCH000537, BCH000543, BCH000553, BCH000554, BCH000556, BCH000564, BCH000565, BCH000567, BCH000570, BCH000573, BCH000574, BCH000575, BCH000576, BCH000577, BCH000579, BCH000580, BCH000583,  BCH000586, BCH000590, BCH000592, BCH000593, BCH000594, BCH000595, BCH000598, BCH000601, BCH000602, BCH000603, BCH000607, BCH000610, BCH000612, BCH000616, BCH000628, BCH000631, BCH000633, BCH000635, BCH000637, BCH000640,  BCH000644, BCH000645, BCH000653, BCH000654, BCH000656, BCH000657, BCH000658, BCH000659, BCH000660, BCH000666, BCH000667, BCH000668, BCH000670, BCH000671, BCH000675, BCH000676, BCH000677, BCH000680, BCH000688, BCH000690,  BCH000691, BCH000693, BCH000694, BCH000695, BCH000696, BCH000697, BCH000698, BCH000702, BCH000703, BCH000706, BCH000713, BCH000715, BCH000716, BCH000721, BCH000722, BCH000728, BCH000733, BCH000734, BCH000735, BCH000739,  BCH000740, BCH000743, BCH000745, BCH000748, BCH000750, BCH000757, BCH000758, BCH000760, BCH000761, BCH000762, BCH000763, BCH000764, BCH000767, BCH000769, BCH000771, BCH000772, BCH000776, BCH000783, BCH000869, BCH000870,  BCH000872, BCH000873, BCH000876, BCH000877, BCH000881, BCH000885, BCH000887, BCH000888, BCH000890, BCH000894, BCH000895, BCH000902, BCH000906, BCH000914, BCH000915, BCH000921, BCH000924, BCH000925, BCH000926, BCH000930,  BCH000931, BCH000938, BCH000941, BCH000945, BCH000946, BCH000947, BCH000948, BCH000949, BCH000950, BCH000953,  BCH000954, BCH000959, BCH000964, BCH000966, BCH000967, BCH000976, BCH000977, BCH000979, BCH000985, BCH000987,  BCH000989, BCH000991, BCH000993, BCH000994, BCH000995, BCH001004, BCH001005, BCH001007, BCH001009, BCH001010, BCH001011, BCH001015, BCH001016, BCH001025, BCH001035, BCH001036, BCH001040, BCH001042, BCH001043, BCH001044,  BCH001045, BCH001047, BCH001049, BCH001056, BCH001063, BCH001065, BCH001072, BCH001086, BCH001091, BCH001103,  BCH001109, BCH001110, BCH001111, BCH001112, BCH001114, BCH001117, BCH001118, BCH001119, BCH001121, BCH001129,  BCH001134, BCH001136, BCH001142, BCH001152, BCH096768, BCH096770, BCH096771, BCH096773, BCH096775, BCH096779, BCH096782, BCH096785, BCH096787, BCH096789, BCH096790, BCH096791, BCH096794, BCH096796, BCH096799, BCH096800,  BCH096801, BCH096804, BCH096805, BCH096806, BCH096811, BCH096812, BCH096813, BCH096818, BCH096820, BCH096821, BCH096822, BCH096827, BCH096828, BCH096830, BCH096832, BCH096833, BCH096836, BCH096850, BCH096924, BCH096936,  BCH096947, BCH096954, BCH096956, BCH096959, BCH096962, BCH096963, BCH096964, BCH096967, BCH096968, BCH096969, BCH096972, BCH096973, BCH096974, BCH096975, BCH096976, BCH096977, BCH096978, BCH096979, BCH096981, BCH096983,  BCH096984, BCH096985, BCH096986, BCH096987, BCH096989, BCH096990, BCH096992, BCH096993, BCH096994, BCH096996, BCH096997, BCH096998, BCH096999, BCH097001, BCH097003, BCH097004, BCH097005, BCH097006, BCH097007, BCH097010,   BCH097012, BCH097013, BCH097014, BCH097015, BCH097016, BCH097017, BCH097020, BCH097021, BCH097022, BCH097051, BCH097052, BCH097053, BCH097054, BCH097055, BCH097056, BCH097060, BCH097061, BCH097066, BCH097067, BCH097068,  BCH097069, BCH097070, BCH097075, BCH097076, BCH097078, BCH097079, BCH097083, BCH097088, BCH097091, BCH097092, BCH097094, BCH097096, BCH097098, BCH097100, BCH097101, BCH097103, BCH097104, BCH097107, BCH097110, BCH097112,  BCH097113, BCH097114, BCH097117, BCH097120, BCH097122, BCH097123, BCH097125, BCH097129, BCH097131, BCH097132, BCH097134, BCH097135, BCH097137, BCH097199,  BCH097201, BCH097202, BCH097206, BCH097208, BCH097209, BCH097210,  BCH097211, BCH097212, BCH097213, BCH097214, BCH097215, BCH097217, BCH097218, BCH097219, BCH097222, BCH097224, BCH097225, BCH097226, BCH097227, BCH097228, BCH097230, BCH097232, BCH097233, BCH097235, BCH097236, BCH097237,  BCH097238, BCH097240, BCH097241, BCH097243, BCH097244, BCH097245, BCH097246, BCH097249, BCH097250, BCH097254, BCH097255, BCH097257, BCH097259, BCH097260, BCH097263, BCH097264, BCH097265, BCH097269, BCH097272, BCH097273,  BCH097275, BCH097279, BCH097312, BCH097328, BCH097359,   BCH097363, BCH097370, BCH097374, BCH097375, BCH097382,  BCH097383, BCH097385, BCH097387, BCH097388, BCH097392,  BCH097395, BCH097397, BCH097398, BCH097400, BCH097404,  BCH097405, BCH097406, BCH097407, BCH097408, BCH097409,  BCH097411, BCH097418, BCH097419, BCH097420, BCH097425,  BCH097426, BCH097428, BCH097429, BCH097430, BCH097431,  bch097432, BCH097433, BCH097434, BCH097437, BCH097438,  BCH097439, bch097440, BCH097441, BCH097443, bch097446,   bch097447, bch097448, bch097449, bch097450, bch097451,  bch097452, bch097453, BCH097454, bch097458, BCH097460,  BCH097461, BCH097462, BCH097464, BCH097466, bch097467,  bch097468, bch097470, BCH097472, BCH097473, BCH097474,  BCH097475, BCH097477, BCH097478, BCH097480, BCH097482,  BCH097484, BCH097485, BCH097488, BCH097489, BCH097490,  BCH097491, BCH097492, BCH097495, BCH097496, BCH097497,  bch097498  BCH097499  BCH097506  BCH097507  bch097509  BCH097511  BCH097516  BCH097519  BCH097520  bch097524  BCH097525  BCH097526  BCH097528  BCH097529  BCH097530  BCH097534  BCH097538  BCH097545  BCH097548  BCH097560  BCH097568  BCH097569  BCH097578  BCH097582  BCH097586  BCH097587  BCH097588  BCH097591  BCH097593  BCH097595  BCH097596  BCH097607  BCH097610  BCH097611  BCH097617  BCH097621  BCH097623  BCH097624  BCH097625  BCH097627  BCH097628  BCH097629  BCH097630  BCH097631  BCH097633  BCH097634  BCH097635  BCH097636  BCH097638  BCH097639  BCH097640  BCH097643  BCH097644  BCH097645  BCH097646  BCH097647  BCH097649  BCH097650  BCH097651  BCH097652  BCH097655  BCH097656  BCH097657  BCH097658  BCH097659  BCH097660  BCH097661  BCH097663  BCH097665  BCH097666  BCH097667  BCH097668  BCH097669  BCH097671  BCH097672  BCH097673  BCH097674  BCH097675  BCH097677  BCH097678  BCH097679  BCH097680  BCH097682  BCH097683  BCH097696  BCH097707  BCH097708  BCH097728  BCH097738  BCH097748  BCH097749  BCH097751  BCH097752  BCH097762  BCH097763  BCH097770  BCH097782  BCH097787  BCH097790  BCH097791  BCH097806  BCH097817  bch097828  bch097830  bch097831  BCH097832  bch097840  bch097841  BCH097843  bch097844  bch097846  bch097848  bch097850  bch097855  bch097856  BCH097857  BCH097858  bch097861  BCH097868  BCH097869  BCH097870  BCH097871  BCH097872  BCH097873  BCH097874  BCH097875  BCH097876  BCH097877  BCH097878  BCH097881  BCH097883  BCH097885  BCH097886  BCH097888  bch097889  BCH097890  BCH097891  BCH097893  BCH097897  BCH097900  BCH097901  BCH097903  BCH097904  BCH097905  BCH097906  BCH097909  BCH097910  BCH097912  BCH097913  BCH097916  BCH097918  BCH097923  BCH097930  BCH097931  BCH097934  BCH097935  BCH097938  BCH097946  BCH097948  BCH097949  BCH097950  BCH097951  BCH097954  bch097958  BCH097959  BCH097960  BCH097961  BCH097962  BCH097963  BCH097964  BCH097965  BCH097967  BCH097973  BCH097974  BCH097981  BCH097991  BCH097992  BCH098014  BCH098025  BCH098060  BCH098061  BCH098062  bch098072  bch098073  BCH098075  BCH098076  BCH098077  BCH098078  BCH098079  BCH098080  bch098081  BCH098083  BCH098084  BCH098088  BCH098091  bch098095  BCH098097  BCH098100  bch098103  BCH098104  bch098105  BCH098106  BCH098107  bch098108  BCH098111  BCH098113  BCH098114  bch098115  bch098116  bch098119  bch098120  BCH098122  bch098123  bch098124  bch098125  bch098127  bch098128  bch098129  BCH098139  BCH098142  BCH098144  BCH098145  BCH098146  BCH098155  BCH098157  BCH098158  BCH098165  BCH098166  BCH098173  BCH098175  BCH098179  BCH098184  BCH098229  BCH098230  BCH098234  BCH098249  BCH098256  BCH098259  BCH098262  BCH098263  BCH098264  BCH098265  BCH098268  BCH098269  BCH098270  BCH098271  BCH098272  BCH098273  BCH098274  BCH098275  BCH098276  BCH098278  BCH098279  BCH098280  BCH098281  BCH098282  BCH098283  BCH098284  BCH098285  BCH098286  BCH098288  BCH098289  BCH098290  BCH098292  BCH098293  BCH098294  BCH098295  BCH098296  BCH098298  BCH098299  BCH098301  BCH098304  BCH098305  BCH098306  BCH098307  BCH098308  BCH098309  BCH098310  BCH098311  BCH098312  BCH098316  BCH098353  BCH098375  BCH098389  BCH098396  BCH098411  BCH098412  BCH098414  BCH098415  BCH098416  BCH098417  BCH098421  BCH098425  BCH098426  BCH098427  BCH098428  BCH098430  BCH098431  BCH098432  BCH098433  BCH098434  BCH098435  BCH098436  BCH098437  BCH098438  BCH098439  BCH098440  BCH098441  BCH098442  BCH098443  BCH098444  BCH098445  BCH098446  BCH098447  BCH098448  BCH098449  BCH098450  BCH098451  BCH098453  BCH098454  BCH098455  BCH098456  BCH098457  BCH098458  BCH098459  BCH098460  BCH098461  BCH098467  BCH098488  BCH098502  BCH098516  BCH098561  BCH098562  BCH098563  BCH098568  BCH098574  BCH098575  BCH098577  BCH098579  BCH098580  BCH098581  BCH098582  BCH098583  BCH098585  BCH098588  BCH098589  BCH098590  BCH098591  BCH098592  BCH098593  BCH098594  BCH098595  BCH098596  BCH098598  BCH098599  BCH098600  BCH098601  BCH098602  BCH098603  BCH098604  BCH098605  BCH098606  BCH098607  BCH098608  BCH098609  BCH098610  BCH098612  BCH098613  BCH098614  BCH098615  BCH098616  BCH098619  BCH098620  BCH098621  BCH098622  BCH098623  BCH098624  BCH098625  BCH098626  BCH098628  BCH098630  BCH098648  BCH098669  BCH098700  BCH098702  BCH098712  BCH098746  BCH098757  BCH098760  BCH098787  BCH098822  BCH098826  BCH098833  BCH098835  BCH098885  BCH098968  BCH098998  BCH098999  BCH001098  BCH000916  BCH001094  BCH000407  BCH000469  BCH000704  BCH001014  BCH001033  BCH001037  BCH000496  BCH000912  BCH001093  BCH000875  BCH000497  BCH000971  BCH000547  BCH000450  BCH000498  BCH000452  BCH000904  BCH000455  BCH000539  BCH001026  BCH000898  BCH000448  BCH000982  bch000683  BCH000973  BCH001147  BCH001082  BCH000439  BCH000544  BCH000669    BCH000974  bch000917  BCH000981  BCH001099  BCH001102  BCH000558  BCH000615  BCH000727  BCH000585  BCH000506  BCH000903  BCH000546  BCH000996  BCH000951  BCH000401  BCH000515  BCH000404  BCH001140  BCH001021  BCH001107  BCH000591  BCH000563  BCH001031  BCH000557  BCH098834  BCH000899  BCH000538  BCH000512  BCH001120  BCH001019  BCH001022  BCH000430  BCH001073  BCH000880  BCH000611  BCH000741  BCH000775  BCH000545  BCH001163    BCH001160  BCH000562  BCH000935  BCH000778  BCH001002  BCH000850  BCH000882  BCH000883  BCH001116  BCH000751  BCH000848  BCH000897  BCH000864  BCH001039  BCH000922  BCH000505  BCH000952  BCH000536  BCH001055  BCH000701  BCH000777  BCH000955  BCH000779  BCH000969  BCH000961  BCH001071  BCH000962  BCH000963  BCH000980  BCH000532  BCH000997  BCH000516  BCH000998    BCH000720  BCH001001  BCH000988  BCH001053  BCH000965  BCH000714  BCH000416  bch000742  BCH000724  BCH000936  BCH000896  BCH000766  BCH001146  BCH000684  BCH000908  BCH000891  BCH000482  BCH000481    bch000884  BCH000893  BCH000752  BCH000911  BCH000413  BCH000453  BCH000582  BCH000651  BCH000744  BCH000754  BCH000454  BCH000920  BCH000990  BCH001079  BCH000468  BCH000905  BCH000441  BCH001032  BCH000402  BCH000449  BCH000488  BCH000958  BCH000489  BCH096866  BCH000467  BCH000491  BCH001006  BCH000655  BCH001097  BCH001061  BCH000968  BCH001038  BCH000566  BCH000542  bch000918  BCH000473  BCH000597  BCH000517  BCH000584  BCH001076  BCH000596  BCH000940  BCH000943  BCH000460  BCH000956  BCH000672  BCH000957  BCH001127  BCH000443  BCH000738  BCH001100  BCH000561  BCH000755  BCH001106  BCH000541  BCH000520  bch000707  BCH000521  BCH001059  BCH001067  BCH000939  BCH000599  BCH000509  BCH000510  bch000774  BCH000878  BCH001069  BCH001060  BCH001145  BCH000642  BCH000909  BCH000555  BCH000552  BCH001078  BCH000913  BCH000986  BCH001077  BCH000901  BCH001068  BCH001070  BCH000535  BCH001075  BCH000589  BCH000606  BCH000540  BCH000629  BCH000975  BCH001041  BCH000569  BCH001153  BCH000708  BCH001080  BCH001113  BCH000447  BCH001096  BCH001108  BCH000550  BCH096856  BCH096858  BCH096862  BCH096865  bch097368  BCH096870  BCH096871  BCH096876  BCH096877  BCH096878  BCH096879  BCH096880  BCH096881  BCH096882  BCH096886  BCH096887  BCH096890  BCH096892  BCH096893  BCH097664  BCH096899  BCH096900  BCH096902  BCH096904  BCH096907  BCH096911  BCH096912  BCH096914  BCH096916  BCH098022  BCH096772  BCH096929  BCH096803  BCH096807  BCH096860  BCH097705  BCH097505  BCH097024  BCH097025  BCH097029  BCH097030  BCH097033  BCH097035  BCH097036  BCH097170  BCH097037  BCH097038  BCH097040  BCH097041  BCH097042  BCH097044  BCH097045  BCH097046  BCH097047  BCH097050  BCH097057  BCH097058  BCH097059  BCH097062  BCH097063  BCH097064  BCH097065  BCH097071  BCH097072  BCH097073  BCH097074  BCH097077  BCH097080  BCH097081  BCH097082  BCH097084  BCH097085  BCH097086  BCH097087  BCH097089  BCH097090  BCH097093  BCH098036  BCH097095  BCH097097  BCH097099  BCH097102  BCH097105  BCH097106  BCH097108  BCH097140  BCH097141  BCH097144  BCH097150  BCH097152  BCH097154  BCH097156  BCH097157  BCH097158  BCH097159  BCH097160  BCH097161  BCH097164  BCH097169  BCH097517  BCH097174  BCH097176  BCH097180  BCH097181  BCH097767  BCH097184  BCH097986  BCH097189  BCH097191  BCH097194  BCH097196  BCH097200  BCH097203  BCH097204  BCH097205  BCH098048  BCH097216  BCH097220  BCH097221  BCH097223  BCH097229  BCH097231  BCH097688  BCH097234  BCH097242  BCH097247  BCH097248  BCH097251  BCH097253  BCH097282  BCH097285  BCH097286  BCH097287  BCH001081  BCH097288  BCH097289  BCH097290  BCH097292  BCH097735  BCH097295  BCH097297  BCH097299  BCH097301  BCH097302  BCH097303  BCH097304  BCH097305  BCH097307  BCH097308  BCH097309  BCH097311  BCH097536  BCH097322  BCH097335  BCH097342  BCH097343  BCH097346  BCH097349  BCH097351  BCH097355  BCH097356  BCH097357  BCH097361  BCH097364  BCH097366  BCH097369  BCH097371  BCH097372  BCH097373  BCH097379  BCH097381  BCH097384  BCH097386  BCH097389  BCH097390  BCH097391  BCH097393  BCH097394  BCH096867  BCH096769  BCH098054  BCH097797  BCH097603  BCH097626  BCH097632  BCH097637  BCH097641  BCH097642  BCH097653  BCH097654  BCH096897  BCH097670  BCH097291  BCH097535  BCH097547  BCH097692  BCH097693  BCH097694  BCH097695  BCH097781  BCH097699  BCH097702  BCH097239  BCH097706  BCH097720  BCH097721  BCH097726  BCH096826  BCH097739  BCH096829  BCH097740  BCH098239  BCH096831  BCH097756  BCH096837  BCH096840  BCH096843  BCH096852  BCH096845  BCH096848  BCH096855  BCH096857  BCH096859  BCH096861  BCH096863  BCH096864  BCH096824  BCH096825  BCH097757  BCH096819  BCH097760  BCH096808  BCH096809  BCH097761  BCH096810  BCH096823  BCH098804  BCH098520  BCH097598  BCH098823  BCH096792  BCH096793  BCH096795  BCH096797  BCH097772  BCH096798  BCH096802  BCH098763  BCH000934  BCH098761  BCH096776  BCH097540  BCH097557  BCH097558  BCH096777  BCH096778  BCH096774  BCH096780  BCH096781  BCH097559  BCH096786  BCH096783  bch097620  BCH097773  BCH097353  BCH096788  BCH096784  BCH098796  bch097320  bch097613  BCH097774  bch097614  bch097321  bch097340  BCH097604  BCH097792  BCH097795  BCH098407  BCH097799  BCH097800  BCH097803  BCH097805  BCH097808  BCH097814  BCH097815  BCH098318  BCH098322  BCH097684  BCH097685  BCH097686  BCH097689  bch097601  BCH097690  BCH097691  BCH097697  BCH097700  BCH097701  BCH097703  BCH097704  BCH097710  BCH097711  BCH097712  BCH097713  BCH097714  BCH097715  BCH097716  BCH097717  BCH097718  BCH097719  BCH097722  BCH097723  BCH098043  BCH098044  BCH097724  BCH097725  BCH097727  BCH097729  BCH097730  BCH097731  BCH097732  BCH097733  BCH097734  BCH097736  BCH097737  BCH097741  BCH097742  BCH097743  BCH097744  BCH097745  BCH097746  BCH097753  BCH097754  BCH097465  BCH097471  BCH098844  BCH097755  BCH097758  BCH097759  BCH097476  BCH097765  BCH097777  BCH096814  BCH097486  BCH097130  BCH098845  BCH098021  BCH098045  BCH098046  BCH097648  BCH098050  BCH098051  bch098039  BCH098857  BCH097508  bch097615  BCH001133  BCH097537  BCH098797  BCH097608  BCH097501  BCH097502  bch097616  BCH097609  BCH097503  BCH097487  BCH097612  BCH097618  BCH097619  bch097605  bch097606  BCH098017  bch097709  BCH097376  BCH097481  BCH097969  BCH097377  BCH097971  BCH097972  BCH097975  BCH097976  BCH097978  BCH097979  BCH097980  BCH097983  BCH097988  BCH097990  BCH097993  BCH097995  BCH098001  bch097622  bch097676  BCH098011  bch097662  BCH098016  BCH098018  BCH098019  BCH098359  BCH098020  BCH098024  BCH098028  BCH098031  BCH098032  BCH098038  BCH098059  BCH098063  BCH098824  BCH098071  BCH098074  BCH098085  BCH098086  BCH098087  BCH098089  BCH098090  BCH098092  BCH098096  BCH098099  BCH098101  BCH098102  BCH098109  BCH098806  BCH098110  BCH098112  BCH098117  BCH098836  BCH098118  BCH098121  BCH098130  BCH098131  BCH098837  BCH096815  BCH098762  BCH097546  BCH098799  BCH098034  BCH098408  BCH098000  BCH098002  BCH098003  BCH097313  BCH097314  BCH001104  BCH098047  bch098035  BCH098026  BCH098029  BCH098030  BCH098033  BCH098023  BCH098013  BCH096816  BCH098015  BCH098037  BCH098377  BCH098379  BCH098053  BCH098764  BCH098055  BCH097315  BCH097331  BCH098056  BCH097360  BCH001123  BCH098049  BCH098057  BCH097332  BCH098040  BCH098041  BCH098052  BCH098317  BCH098042  BCH098067  BCH098068  BCH098069  bch098070  BCH098064  BCH098065  BCH097698  BCH097252  BCH001162  BCH000560  BCH000753  BCH000932  BCH000933  BCH096817  BCH000970  BCH098409  BCH098410  BCH000972  BCH098777  BCH098789  BCH098778  BCH098240  BCH098241  BCH098242  BCH098211  BCH098220  BCH098212  BCH098213  BCH098214  BCH098215  BCH098216  BCH098221  BCH097310  BCH098206  BCH098223  BCH098224  BCH098243  BCH098225  BCH098226  BCH098227  BCH098228  BCH098207  BCH098244  BCH098208  BCH098245  BCH098413  BCH098217  BCH098218  BCH098209  BCH098210  BCH097818  BCH098253  BCH098231  BCH098250  BCH098252  BCH098254  BCH098082  BCH098137  BCH098233  BCH098236  BCH098219  BCH098237  BCH098255  BCH098257  BCH098258  BCH098261  BCH098385  BCH098759  BCH098734  BCH098790  BCH096834  BCH096835  BCH096838  BCH096839  BCH098380  BCH098381  BCH098550  BCH096841  BCH096842  BCH096844  BCH098386  BCH098387  BCH098388  BCH098390  BCH097333  BCH097358  BCH098741  BCH097750  BCH097764  BCH097768  BCH098302  BCH098303  BCH098405  BCH098406  BCH098402  BCH098403  BCH098404  BCH098394  BCH098395  BCH098397  BCH098398  BCH098399  BCH098400  BCH098401  BCH098418  BCH098419  BCH098420  BCH098422  BCH098423  BCH098424  BCH098429  BCH098452  BCH098287  BCH098846  BCH096846  BCH098821  BCH098721  BCH098500  BCH098505  BCH098508  BCH098531  BCH098540  BCH098541  BCH098542  BCH098543  BCH098544  BCH098545  BCH098546  BCH098547  BCH098549  BCH098551  BCH098476  BCH098552  BCH098553  BCH098554  BCH098477  BCH098495  BCH098555  BCH098744  BCH098745  BCH098557  BCH098558  BCH098559  BCH098560  BCH098360  BCH098362  BCH098522  BCH098523  BCH098524  BCH098525  BCH098526  BCH098527  BCH098528  BCH098529  BCH098530  BCH098803  BCH098807  BCH098564  BCH098565  BCH098566  BCH098567  BCH098569  BCH098570  BCH098825  BCH098571  BCH098572  BCH098576  BCH098862  BCH096847  BCH098720  BCH098722  BCH098792  BCH098791  BCH098805  BCH098723  BCH098587  BCH098793  BCH098747  BCH098749  BCH098490  BCH096849  BCH096851  BCH097300  BCH096854  BCH001087  BCH001150  BCH001154  BCH000639  bch001017  BCH001149  BCH000487  BCH001148  BCH001159  bch000746  BCH000765  BCH001157  BCH000492  BCH000618  BCH001128  BCH000928  BCH000620  BCH000632  bch000749  BCH000434  bch000759  BCH000444  BCH000411  BCH001000  BCH000678  BCH001125  bch000725  BCH001158  BCH001141  BCH000730  BCH000494  BCH001115  BCH000929    BCH000900  BCH000483  BCH001151  BCH001165  BCH001132  BCH001126  BCH001013  BCH001003  BCH000463  BCH000427  BCH000419  bch000711  BCH000522  BCH000551  BCH000424  BCH000664  BCH000770  BCH001058  BCH001131  BCH000782  BCH001101  BCH000617  BCH000992  bch000649  BCH098752  BCH000533  BCH000445  BCH000437  BCH000999  BCH000501  BCH000409  BCH000874  BCH001090  BCH000910  BCH000438  BCH000712  BCH000626  BCH000479  BCH000478    BCH000458  BCH000504  BCH000960  BCH000634  BCH000613  BCH000718  BCH000723  BCH000614  BCH000571  BCH000622  BCH000699  BCH000636  BCH000780  BCH000773  BCH000879  BCH000978  BCH000674  BCH000457  BCH000480  BCH000927  BCH000568  BCH000673  BCH000892  BCH001155  BCH000647  BCH001034  BCH000609  BCH000726  BCH000519  BCH001051  BCH000705  BCH000410  BCH000662    BCH001105  BCH000710  BCH001027  BCH001024  BCH000650  BCH001144  BCH001161  bch000719  BCH000499  BCH000425  BCH000737  BCH000732  BCH000700  bch000679  BCH000513  BCH000937  bch000681  BCH000663    BCH000731  BCH000942  BCH000984  BCH000508  BCH000630  BCH000648  BCH000983  BCH000624  BCH000661  BCH000687  BCH000643  BCH000619  BCH001089  BCH000652  BCH000638  BCH000646  BCH000709  BCH001138  bch000689  BCH000605  bch000682  BCH000641  BCH000781  BCH096952  bch000692  BCH001135  BCH000418  BCH000429  BCH001052  BCH001029  bch000608  BCH001054  BCH000886  BCH000549  BCH000431  BCH000548  BCH000476  BCH000417  BCH000600  BCH000588  BCH000768  BCH000717  BCH000518  bch000919  BCH000511  BCH000466  BCH000729  BCH001139  BCH000756  BCH000604  BCH000578  bch000747  BCH000736  BCH001084  BCH000559  BCH000462  BCH000421  BCH001030  bch000889  BCH001046  BCH000529  BCH000923  BCH000432  BCH000436  BCH000685  BCH000625  BCH001048  BCH001023  BCH001062  BCH000581  BCH000871  BCH001020  BCH000665  BCH001064  BCH000623  BCH001092  BCH001156  BCH000412  BCH001057  BCH000587  BCH000474  BCH000907  BCH001166  BCH000414  BCH000627  BCH000621  BCH000408  BCH001088  BCH001130  BCH000461  BCH001018  BCH001066  BCH001095  BCH000944  BCH001085  BCH001164  BCH000686  BCH096957  BCH096958  BCH096961  BCH096965  BCH097561  BCH097575  BCH098167  BCH096971  BCH096980  BCH096982  BCH096988  BCH096991  BCH096995  BCH097000  BCH097002  BCH097008  BCH097009  BCH097011  BCH097687  BCH097018  BCH097019  BCH097023  BCH097026  BCH097027  BCH097028  BCH097031  BCH097032  BCH097034  BCH097039  BCH097043  BCH097048  BCH097049  BCH096905  BCH096906  BCH096960  BCH097927  BCH097510  BCH097109  BCH097111  BCH097115  BCH097116  BCH097118  BCH097119  BCH097121  BCH097281  BCH097124  BCH097126  BCH097127  BCH097128  BCH097133  BCH097138  BCH097139  BCH097142  BCH097143  BCH097145  BCH097146  BCH097147  BCH097148  BCH097149  BCH097151  BCH097153  BCH097155  BCH097162  BCH097163  BCH097165  BCH097166  BCH097167  BCH097168  BCH097171  BCH097172  BCH097173  BCH097175  BCH097177  BCH097178  BCH097179  BCH097182  BCH097185  bch097922  BCH097186  BCH097187  BCH097190  BCH097192  BCH097193  BCH097195  BCH097197  BCH097256  BCH097258  BCH097261  BCH097262  BCH097266  BCH097267  BCH097268  BCH097270  BCH097271  BCH097274  BCH097276  BCH097277  BCH097278  BCH097280  bch097574  BCH097283  BCH097284  BCH097327  BCH097293  BCH097294  BCH097296  BCH097298  BCH096970  BCH097188  BCH097427  BCH097316  BCH097317  BCH098093  BCH097319  BCH097323  BCH097325  BCH097326  BCH097681  BCH097329  BCH097330  BCH097334  BCH097336  BCH097339  BCH097341  BCH097347  BCH097348  BCH097350  BCH097362  BCH097365  BCH097396  BCH097399  BCH097401  BCH097402  BCH000477  BCH097403  BCH097410  BCH097412  BCH097413  BCH097941  BCH097414  BCH097415  BCH097416  BCH097417  BCH097421  BCH097422  BCH097423  BCH097424  BCH097436  BCH097442  BCH097444  BCH097445  BCH097522  BCH097455  BCH097456  BCH097459  BCH097463  BCH097469  BCH097479  BCH097513  BCH097514  BCH097515  BCH097521  BCH097523  BCH097542  BCH097550  BCH097552  BCH097554  BCH097555  BCH097563  BCH097564  BCH097566  BCH097567  BCH097576  BCH097577  BCH097580  BCH097581  BCH097589  BCH097602  BCH096966  BCH096868  bch097926  BCH097987  BCH097819  BCH097822  BCH097824  BCH097826  BCH097836  BCH097837  BCH097845  BCH097847  BCH097859  BCH097864  BCH097865  BCH097866  BCH097867  BCH097880  BCH097882  BCH097884  BCH097887  BCH097899  BCH097915  BCH097917  BCH097345  BCH097928  BCH097929  BCH097932  BCH097936  BCH096925  BCH097942  BCH096926  BCH097943  BCH098188  BCH096927  BCH097944  BCH096928  BCH096931  BCH096932  BCH096939  BCH096934  BCH096938  BCH096940  BCH096942  BCH096943  BCH096944  BCH096946  BCH096948  BCH096922  BCH096923  BCH097945  BCH096920  BCH097947  BCH096908  BCH096909  BCH097953  BCH096910  BCH096921  BCH098719  BCH098640  BCH097966  BCH098694  BCH096894  BCH096895  BCH096896  BCH096898  BCH097599  BCH096901  BCH096903  BCH098539  BCH001008  BCH098537  BCH096872  bch097584  bch097590  BCH097527  BCH096873  BCH096874  BCH096869  BCH096875  BCH096883  BCH097533  BCH096889  BCH096884  BCH097539  BCH097970  BCH097543  BCH096891  BCH096885  BCH098651  BCH097544  BCH097207  BCH097977  BCH097306  BCH097551  BCH097553  BCH097562  BCH097985  BCH097378  BCH098352  BCH097989  BCH098006  BCH098066  BCH098094  BCH098027  BCH098222  BCH098058  BCH098323  BCH098324  BCH097766  BCH097771  BCH097775  BCH097776  BCH097198  BCH097778  BCH097779  BCH097780  BCH097783  BCH097784  BCH097785  BCH097786  BCH097788  BCH097789  BCH097793  BCH097794  BCH097796  BCH097798  BCH097801  BCH097802  BCH097804  BCH097807  BCH097809  BCH097810  BCH097921  BCH097924  BCH097811  BCH097812  BCH097813  BCH097816  BCH097820  BCH097821  BCH097823  BCH097825  BCH097827  BCH097829  BCH097833  BCH097834  BCH097835  BCH097838  BCH097839  BCH097842  BCH097849  BCH097851  BCH097852  BCH096919  BCH097338  BCH098659  BCH097853  BCH097860  BCH097862  BCH097483  BCH097863  BCH096853  BCH096913  BCH097531  BCH097532  BCH098667  BCH097493  BCH097933  BCH097937  BCH098134  BCH097952  bch097919  bch097920  BCH098674  BCH097512  BCH097324  BCH001050  bch097367  BCH097541  BCH097565  BCH097457  BCH097494  BCH097337  BCH097570  BCH097500  BCH097344  BCH097571  BCH097854  bch097549  bch097556  bch097572  BCH097597  BCH097585  BCH097592  bch097573  BCH098135  BCH097594  BCH098136  BCH098138  BCH098140  BCH098143  BCH098147  BCH098151  BCH098154  BCH098156  BCH098159  BCH098160  BCH098162  BCH098164  BCH098168  BCH097600  BCH097583  BCH098182  BCH097579  BCH098185  BCH098232  BCH098235  BCH098326  BCH098238  BCH098246  BCH098247  BCH098248  BCH098251  BCH098260  BCH098266  BCH098267  BCH098696  BCH098291  BCH098297  BCH098325  BCH098327  BCH098333  BCH098335  BCH098337  BCH098340  BCH098343  BCH098344  BCH098346  BCH098347  BCH098351  BCH098682  BCH098356  BCH098361  BCH098363  BCH098652  BCH098378  BCH098391  BCH098392  BCH098393  BCH098653  BCH096915  BCH098538  BCH098697  BCH098699  bch097895  BCH098354  bch097914  BCH097352  BCH097354  bch097907  bch097911  BCH000572  bch097896  bch097898  BCH097892  BCH097902  BCH097908  bch097894  BCH097504  BCH097518  BCH096917  BCH097183  bch097925  BCH098328  BCH098329  bch097999  BCH098631  BCH097955  bch097984  bch097994  BCH097956  bch097996  BCH001083  bch097997  BCH097957  bch097982  BCH097939  BCH097940  bch097998  BCH098321  BCH097435  BCH098008  BCH098009  BCH098010  BCH098012  BCH098004  BCH098005  BCH098007  BCH097968  BCH001143  BCH001074  BCH001012  BCH001028  BCH001137  BCH096918  BCH001122  BCH098355  BCH098357  BCH001124  BCH098632  BCH098641  BCH098633  BCH098189  BCH098190  BCH098191  BCH098170  BCH098198  BCH098171  BCH098172  BCH096888  BCH098126  BCH098132  BCH098199  BCH098200  BCH098133  BCH098201  BCH098202  BCH098192  BCH098148  BCH098149  BCH098150  BCH098152  BCH098153  BCH098193  BCH098161  BCH098194  BCH098374  BCH098195  BCH098196  BCH098163  BCH098169  BCH098177  BCH098178  BCH098180  BCH098181  BCH098183  BCH098186  BCH097747  BCH097769  BCH097879  BCH098203  BCH098197  BCH098187  BCH098204  BCH098205  BCH098174  BCH098176  BCH098331  BCH098536  BCH098637  BCH098642  BCH096930  BCH096933  BCH096935  BCH097136  BCH098330  BCH098300  BCH098518  BCH096937  BCH096941  BCH096945  BCH098332  BCH098334  BCH098336  BCH098338  BCH098339  BCH098341  BCH098638  BCH098313  BCH098314  BCH098315  BCH098319  BCH098320  BCH098349  BCH098350  BCH098342  BCH098345  BCH098348  BCH098371  BCH098372  BCH098373  BCH098376  BCH098358  BCH098364  BCH098141  BCH098382  BCH098383  BCH098384  BCH098365  BCH098366  BCH098367  BCH098368  BCH098369  BCH098370  BCH098673  BCH097318  BCH098684  BCH098535  BCH098472  BCH098474  BCH098475  BCH098486  BCH098487  BCH098489  BCH098277  BCH098462  BCH098463  BCH098464  BCH098465  BCH098466  BCH098468  BCH098519  BCH098469  BCH098499  BCH098504  BCH098506  BCH098470  BCH098471  BCH098507  BCH098639  BCH098586  BCH098491  BCH098492  BCH098496  BCH098497  BCH098498  BCH098501  BCH098503  BCH098478  BCH098479  BCH098480  BCH098481  BCH098482  BCH098483  BCH098484  BCH098485  BCH098713  BCH098683  BCH098513  BCH098521  BCH098509  BCH098510  BCH098511  BCH098512  BCH098532  BCH098514  BCH098515  BCH098517  BCH098675  BCH096949  BCH098634  BCH098635  BCH098649  BCH098644  BCH098677  BCH098636  BCH098534  BCH098650  BCH098617  BCH098618  BCH098533  BCH096950  BCH096951  BCH096953  BCH096955"
    }
  ]
}