  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elyria",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69441",
      "recalling_firm": "Invacare Corporation",
      "address_1": "1200 Taylor St",
      "address_2": "N/A",
      "postal_code": "44035-6248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.",
      "recall_number": "Z-0577-2015",
      "product_description": "Invacare Perfecto 2 V Oxygen Concentrator    The finished device is packaged as a single unit.  The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device.      Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.",
      "product_quantity": "5,963",
      "reason_for_recall": "It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.",
      "recall_initiation_date": "20140923",
      "center_classification_date": "20141209",
      "termination_date": "20150618",
      "report_date": "20141217",
      "code_info": "The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099.      The serial number range is as follows for the finished device:   14GF053772-14HF032124 (Sequential serial numbers)"
    }
  ]
}