{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72868",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.",
      "recall_number": "Z-0571-2016",
      "product_description": "BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x  4.0 mL (Lavender) BD Hemogard  Additive: K2EDTA (spray dried), 7.2mg Sterile IVD BD, Made in USA     BD Vacutainer¿ tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory. The BD Vacutainer¿ Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening.",
      "product_quantity": "29,815,600 units in total",
      "reason_for_recall": "Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.",
      "recall_initiation_date": "20151203",
      "center_classification_date": "20151231",
      "termination_date": "20181019",
      "report_date": "20160106",
      "code_info": "Catalog Reference #367844; Lot number 5154624    Catalog Reference #367862; Lot numbers 5089771 and 5187753    Catalog Reference #368021; Lot number 5089820"
    }
  ]
}