  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "South Burlington",
      "state": "VT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66754",
      "recalling_firm": "Data Innovations, Inc.",
      "address_1": "120 Kimball Ave Ste 100",
      "address_2": "N/A",
      "postal_code": "05403-6837",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide  Canada and Britain",
      "recall_number": "Z-0571-2014",
      "product_description": "Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer    Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004",
      "product_quantity": "24 systems",
      "reason_for_recall": "When connected to the Sysmex UF-100, the  Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003,  and 8.00.0004, and is configured with option \"Strip Zeros from Result\" set to \"Yes\"  may result in  sample error",
      "recall_initiation_date": "20131106",
      "center_classification_date": "20131224",
      "termination_date": "20141205",
      "report_date": "20140101",
      "code_info": "Software alll versions prior to: v7.00.0006 and v8.00.  Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004",
      "more_code_info": ""
    }
  ]
}