{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89327",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-0570-2022",
      "product_description": "Baxter MiniCap Extended Life PD Transfer Set (Easy-Lock), 5C4449",
      "product_quantity": "23,940 units",
      "reason_for_recall": "Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.",
      "recall_initiation_date": "20220107",
      "center_classification_date": "20220208",
      "report_date": "20220216",
      "code_info": "All lot numbers within expiry"
    }
  ]
}