  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Garland",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86795",
      "recalling_firm": "Galt Medical Corporation",
      "address_1": "2220 Merritt Dr",
      "address_2": "N/A",
      "postal_code": "75041-6137",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: FL, GA, MA, NC, NJ, OH, PA, Puerto Rico and UT.    O.U.S: Canada, Netherlands, Turkey, United Kingdom, and UAE",
      "recall_number": "Z-0568-2021",
      "product_description": "Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-002-06    Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-006-16",
      "product_quantity": "9,920 catheters (updated on 03/17/2021)",
      "reason_for_recall": "Due to a manufacturing issue, the catheter tip geometry is incorrect (thicker wall) resulting in a more obtuse tip angle.",
      "recall_initiation_date": "20201116",
      "center_classification_date": "20201214",
      "termination_date": "20221107",
      "report_date": "20201223",
      "code_info": "Catalog Number: DRC-002-06    Lot Numbers/UDI Code:  20108509/(01)0084 1268 100398(17)240428(10)20108509  20139676/(01)0084 1268 100398(11)200520(17)240519(10)20139676  20148911/(01)0084 1268 100398(11)200609(17)240608(10)20148911  20160253/(01)0084 1268 100398(11)200707(17)240706(10)20160253  20189100/(01)0084 1268 100398(11)200821(17)240820(10)20189100    Updated Information provided 03/16/2021:  For Catalog Number: DRC-002-06 additional Lot Numbers:  -20080351  -20234573  -20297230  -21008452    Additional Products:  Catalog Number: DRC-006-16  Lot Numbers:  -20217235  -20217236  -20217237"
    }
  ]
}