  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72712",
      "recalling_firm": "Aesculap, Inc.",
      "address_1": "3773 Corporate Pkwy",
      "address_2": "N/A",
      "postal_code": "18034-8217",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to state of: NY.",
      "recall_number": "Z-0566-2016",
      "product_description": "Valve XS Atrium Retractor  FC429R  52154756    The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery.",
      "product_quantity": "41",
      "reason_for_recall": "AAG received a customer complaint that corrosion had been found by a hospital\u0019s central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier",
      "recall_initiation_date": "20151117",
      "center_classification_date": "20151231",
      "termination_date": "20161013",
      "report_date": "20160106",
      "code_info": "Product Information All lots affected   Product Description  FC424R Valve XS Atrium Lift Ret. Blade50X35  FC425R Valve XS Atrium Lift Ret. Blade60X35  FC426R Valve XS Atrium Lift Ret. Blade70X35  FC428R Valve XS Atrium Lift Ret. Blade50X45  FC429R Valve XS Atrium Lift Ret. Blade60X45  FC430R Valve XS Atrium Lift Ret. Blade70X45  FC431R Valve XS Atrium Lift Ret. Blade25X20  FC432R Valve XS Atrium Lift Ret. Blade25X25  FC433R Valve XS Atrium Lift Ret. Blade40X25  FC434R Valve XS Atrium Lift Ret. Blade50X25  FC435R Valve XS Atrium Lift Ret. Blade25X40  FC436R Valve XS Atrium Lift Ret. Blade25X50  FC420R Valve XS Atrium Lift Ret. Blade Complete Set"
    }
  ]
}