  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65131",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr Ms 514",
      "address_2": "PO BOX 6101",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide and OUS.",
      "recall_number": "Z-0561-2014",
      "product_description": "Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052      Product Usage:  Used in the Dimension(R) system with the heterogeneous module",
      "product_quantity": "17,719",
      "reason_for_recall": "Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461.  These lots shipped between 12/12/2012 and 3/22/2013.  Internal investigation has confirmed shifts can be encountered in two scenarios:  1 - When an affected l",
      "recall_initiation_date": "20130506",
      "center_classification_date": "20131223",
      "termination_date": "20140827",
      "report_date": "20140101",
      "code_info": "Lot numbers RD23031, exp 2013-04-29; RD23111, exp 2013-05-07; RD23241, exp 2013-05-19; RD23311, exp 2013-06-04; and RD23461, exp 2013-06-11",
      "more_code_info": ""
    }
  ]
}