{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marlborough",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84150",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "300 Boston Scientific Way",
      "address_2": "",
      "postal_code": "01752-1291",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:   Austria  Australia  Belgium  Bulgaria  Croatia  Czech Rep  Denmark  Finland  France  Germany  Great Britain  Italy  Japan  Pakistan  Poland  Russian Fed.  Slovakia  South Africa  Spain  Sweden  Switzerland  Thailand  Turkey",
      "recall_number": "Z-0560-2020",
      "product_description": "Boston Scientific Advanix Pancreatic Stent, Straight Leading Barb, 10F x 18cm  UPN: M00536730",
      "product_quantity": "35 units",
      "reason_for_recall": "Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life",
      "recall_initiation_date": "20191021",
      "center_classification_date": "20191125",
      "termination_date": "20231220",
      "report_date": "20191204",
      "code_info": "Lot Numbers:  21542436       21897084       23230412",
      "more_code_info": ""
    }
  ]
}