{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "61593",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AL, CA, CO, FL, GA, IL, IN, IA, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX , UT, VA, WI and Internationally to the following countries:    VENEZUELA, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TUNISIA, THAILAND, TAIWAN, SWITZERLAND, SPAIN, SOUTH AFRICA, SINGAPORE, SERBIA, SAUDI ARABIA, RUSSIA, ROMANIA, REUNION, PORTUGAL, POLAND, PHILIPPINES, PERU, PANAMA, NORWAY, NIGERIA, NEW CALEDONIA, NEPAL, MEXICO, MALAYSIA, KOREA, JAPAN, ITALY, IRELAND, INDONESIA, INDIA, HONG KONG, GREECE, GERMANY, FRANCE, FINLAND, ESTONIA, EGYPT, ECUADOR, DENMARK, CZECH REPUBLIC, CHINA, CHILE, CANADA, BULGARIA, BRAZIL, BELGIUM, BELARUS, AUSTRIA, AUSTRALIA, ARGENTINA ALGERIA.",
      "recall_number": "Z-0560-2013",
      "product_description": "Advantage Workstation Volume Share 2 (versionAW4.4_04) with Volume Viewer 3 and 3.1 (version7.0 to version7.5 and version 8.1) and only when or more of these applications is installed:  \" VesselIQ Xpress or AVA Xpress  \" CardIQ Xpress Pro or Plus  \" CardEP  \" CardIQ Fusion PET or SPECT  Advantage Workstations Volume Share (v.AW4.3) and AW4.2 (version4.2-05 and above) provided with Volume Viewer 2(version6.0 to version6.11) and only when one or both of these applications is installed:  \" CardIQ Xpress Pro or Plus  \" CardEP    Advantage Workstations AW Volume Share 2 (AW version 4.4).  ONLY IF ONE OF FOLLOWING APPLICATIONS IS INSTALLED:  ¿ VesselIQ Xpress or AVA Xpress  ¿ CardIQ Xpress Pro or Plus  ¿ CardEP  ¿ CardIQ Fusion PET or SPECT  provided with Volume Viewer 3 and 3.1 from versions 7.0 to 7.5 and 8.1.  To verify the version installed on your system: Select the \u001cAdmin\u001d menu from the \u001cPatient list\u001d page. Then select \u001cDisplay Configuration\u001d. You will find the version of your system in \u001cInstalled Application(s)\u001d. This version will appear as vxtl_7.x or vxtl_8.x.      Intended use: K993792: Smart Vessel Analysis (Smart VA) is a software post-processing option for the Advantage Workstation (AW) platform, which can be used in the analysis of 3D angiography data. It provides a number of display, measurement and batch filming/archive features and will aid physicians in studying user-selected vessels for stenosis analysis, pre/post stent planning and directional vessel tortuosity visualization. K031261: Card EP is a post processing software option for the Advantage Workstation (AW) Platform. This product can be used for the analysis of CT angiographic images for the assessment of the heart to include the atria, pulmonary veins, and coronary sinus. It provides quantitative analysis tools which include a number of display, measurement and model export capabilities. This product can be used to aid trained physicians in the visualization and assessment of cardiac anatomy. K041267: CardIQ Analysis III is a CT image analysis software package, which allows the visualization of 2D and 3D medical image data of the heart derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular disease assessment. It provides functionally for 2D/3D rendering, assessment of calcified and non-calcified plaque to determine the densities of the plaque within a coronary artery, ventricular function of the heart, and measurement tools to detect coronary artery stenosis. This product can be used to aid a trained physician to process, render, review, archive, print and visualizing cardiac anatomy and coronary vessels. CardIQ Analysis II will run on the AW workstation, scanner operator console and PACS system. K061370: CardIQ Fusion is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology, aid in the assessment of functional data e.g. PET perfusion, and aid in tailoring treatment plans based on the fused anatomical and functional information. Anatomical data could be from a set of Computed Tomography (CT) Angiographic images while functional data could be from PET, SPECT, or processed CT data. CardIQ Fusion is a software post-processing package for the Advantage Workstation (AW) platform, PET/CT and CT scanners and PACS reading stations. It is an additional tool for the analysis of 3D CT angiographic cardiac images/data providing a number of display, measurements and batch filming/archive features to study user-selected vessels. Also included is the capability to visualize reformatted CT/PET/SPECT perfusion and viability data. Finally, it provides different ways of visualizing CT anatomy fused with PET/SPECT functional information. With CardIQ Fusion, clinicians have the opportunity to overlay functional information over the (CT) anatomy of a patient's heart, and thus, they can potentially tailor their decision for that particular patient. CardIQ Fusion provides the visualization of the vessels in severa",
      "product_quantity": "548 (118 USA; 430 OUS)",
      "reason_for_recall": "GE Healthcare has became aware of two issues with the Advantage Workstation.    1) A possible mismatch between the label of the tracked vessel and the underlying image associated with the cardio- vascular applications of the Advantage workstation that may impact patient safety.    2) A refresh problem of the stenosis / aneurysm  measurement tools associated with the cardio-vascular applications of",
      "recall_initiation_date": "20120529",
      "center_classification_date": "20121219",
      "termination_date": "20130117",
      "report_date": "20121226",
      "code_info": "\"FMI# & Description :  FMI 25409  VESSEL LABELLING MISMATCH\"  Mfg Lot or Serial #  00000L0219A4C4  00000L02167897  00000L022D0776  00000L0240826C  00000L022924AF  00000L02408728  00000L020E7DEF  00000L02167D50  00000L0219CA0A  00000L025A16BA  00000L02408829  00000L02A1CED2  00000L02127C08  00000L021DFF09  00000L021DFE8C  00000L022925D3  00000L0246B1BB  00000L020E81FF  00000L027461C8  00000L02A13D92  00000L020E0003  00000L021653EE  00000L0213E474  00000L0213E7BA  00000L020E0A27  00000LEE1A74B4  00000L022EE8A6  00000L0235EED0  00000L022922EE  00000L026FFFE3  00000L02127CD6  00000L020E7E4E  00000L0215A47D  00000L6100B6EA  00000L0271A0AE  00000L020E0B95  00000L0276200E  00000L02762053  00000L027623AC  00000L0219C966  00000L021562BA  00000L0246B24D  00000L02408351  00000L02167DC8  00000L020E0AEA  00000L02408638  00000L0235EB60  00000L022D06CA  00000L022D06E2  00000L0246B17D  00000L02127B57  00000L022EE879  00000L027003FD  00000L0235EBFD  00000L02156750  00000L02A79998  00000L02165328  00000L0216769F  00000L02167E0C  00000L02167EC0  00000L02700194  00000L027004A8  00000L027004E6  00000L027452E6  00000L02408796  00000L022EE8B6  00000L02408432  00000L022923B7  00000L3ABB095C  00000L02A05D3B  00000L024085E7  00000L02127D54  00000L0235EB68  00000L025ADA7A  00000L024086A0  00000L02408505  00000L020E07D8  00000L0219C95D  00000L02292338  00000L0240826B  00000L02127C79  00000L022EE7E9  00000L0247FE35  00000LEE1B09F2  00000L02141CDF  00000L029FDC14  00000L0235EC8A  00000L0264CECC  00000L020D73A0  00000L0229247D  00000L022926B7  00000L0235EC49  00000L02A8C496  00000L02167E0F  00000LEE16BE1E  00000L0219A361  00000L027006B3  00000L0246B17C  00000L0246DE23  00000L0247E3DF  00000L021DFE83  00000L022D0782  00000L022EE7E8  00000L022EE7FF  00000L02A8C644  00000L0270036E  00000L020CFA70  00000L02019944  00000L021279D4  00000L0240883C  00000L020E028F  00000L02A13F87  00000L0235EB43  00000L020B0B01  00000L0271A514  00000L0213E6C6  00000L02A79744  00000L02A76C57  00000L02547251  00000L026EAACE  00000L027463B5  00000L0264CEEA  00000L020D64E1  00000L020E07CB  00000L0238B0B2  00000L024083B9  00000L025B57DD  00000L02577C13  00000L0247DB77  00000L0247DBD3  00000L02A9DA71  00000L029C2FAF  00000L02A05ED3  00000L0246DB34  00000LEE1B09C7  00000L0247FE3F  00000L02127AA9  00000L02292354  00000L020B55AB  00000L02408261  00000L020E7CB2  00000L0213E35E  00000L02408598  00000L022EE8D1  00000L02700689  00000L02A79735  00000L02A05D4A  00000L02408628  00000L0246DE67  00000L02408537  00000L024085C6  00000L02A05CAB  00000L0213E5EF  00000L0219C88B  00000L027004E0  00000L02A05BC6  00000L02A9DA74  00000L0274A91F  00000L0274A957  00000L02A05CB0  00000L0219C8D0  00000L0229EA22  00000L02127C4A  00000L0219C84E  00000L0219C85C  00000L0274AC15  00000L020DB044  00000L02167DB6  00000L02156099  00000L0219CA08  00000L0284219B  00000L0235C524  00000L02A0D28B  00000L021676D7  00000L021654A6  00000L02A0D332  00000L02A0D430  00000L0219779A  00000L02135944  00000L02156644  00000L0219C786  00000L02150071  00000L022925C2  00000L021C4C62  00000L0219C885  00000L0213592B  00000L029FDDDA  00000L020B35CF  00000L0219C836  00000L0235EF2C  00000L0257B89B  00000L0216535E  00000L02292602  00000L020D65DF  00000L020E0AB6  00000L0235EB9E  00000L0271A1C5  00000L029FDB12  00000L02156646  00000L021DFDC7  00000L0235EB9B  00000L0229256A  00000L029FDDF5  00000L0219C93E  00000L0235EB53  00000L0271A3BB  00000L02A9592C  00000L02A95963  00000L0235ED94  00000L021C4BD4  00000L0235EC6E  00000L02746410  00000L02141B51  00000L02197752  00000L0202B678  00000L020D791F  00000L02156280  00000L0212FE0F  00000L0275FDB4  00000L02744F24  00000L027664A0  00000L02A9DBB8  00000L0202B668  00000L02745BDA  00000L02408365  00000L022924AA  00000L0235EEE1  00000L0235EEDF  00000L0257B462  00000L020E09F3  00000L0238AFBA  00000L020DB746  00000L024082B2  00000L02408606  00000L0257B7BB  00000L0257B882  00000L02AB1647  00000L02766434  00000L02711B9E  00000L02A05DCA  00000L02A05EB6  00000L020E2F37  00000L0246DE1C  00000L0246DE66  00000L027664C1  00000L0271A35D  00000L026EAA9A  00000L026EA87B  00000L026EA8F0  00000L025470C4  00000L02744BA6  00000L02746C51  00000L026EA854  00000L029FDB24  00000L02168069  00000L0247FDE3  00000L022D0615  00000L02AB15D8  00000L020DB21A  00000L0238AB2A  00000L026CC2CF  00000L026D3E08  00000L0271A051  00000L029C3007  00000L0274AB92  00000L02A05C96  00000L020DB1C8  00000L022D0628  00000L022D072D  00000L021562E5  00000L0271A1F0  00000L02A95AE6  00000L0216531B  00000L020E5EDB  00000L0238B09C  00000L0238B111  00000L0238B1B3  00000L022D4E94  00000L02AB15D2  00000L02AB165E  00000L02A8BE38  00000L020D7B1B  00000L0213E51E  00000L024087A1  00000L0240865B  00000L026EA7DD  00000L026EA950  00000L026EA88B  00000L02711BA1  00000L0257B423  00000L02370904  00000L020D790F  00000L0229E965  00000L021978B1  00000L0274A059  00000L0274ADA7  00000L0219C83D  00000L0238B0E6  00000L0229E9B6  00000L0247DC2E  00000L020D6975  00000L02A9DB8D  00000L0235C60A  00000L0271A453  00000L0274A7AC  00000L3A1E932B  00000L0238B195  00000L0257B88D  00000L02547277  00000L0202B5CC  00000L026D3FA0  00000L0219786B  00000L020DB335  00000L026EA928  00000L02765FED  00000L02765FF1  00000L021567F8  00000L0215034C  00000L3A4887F6  00000L02167E14  00000L027628DD  00000L029FDC25  00000L02156278  00000L0257B67B  00000LEE1FE90E  00000L02A958FD  00000L021678D0  00000L02141CC9  00000L0257B806  00000L0238B305  00000L027626C2  00000L02694541  00000L0202B66A  00000L020E098D  00000L02A9DBB3  00000L02155F8F  00000L0240865D  00000L020E0865  00000L02141AC9  00000L022D4EEC  00000L0247DCDB  00000L0229E9B1  00000L0247FD80  00000L0202B4D2  00000L0238B0DE  00000L02A0D232  00000L0247DBF3  00000L0257B62E  00000L02133F39  00000L0257B763  00000L0246DA25  00000L0274A60E  00000L02A959A6  00000L3A420C0B  00000L0257B654  00000L0235B392  00000L0235B918  00000L0219C802  00000L0219C93B  00000L0257B71D  00000L026EA856  00000L0215688B  00000L022D4E8A  00000L026EA864  00000L0257B8D5  00000L0240856F  00000L026D3F83  00000L023A0E8E  00000L020E5E03  00000L0235BD3E  00000L0215A428  00000L0247DC9A  00000L0246DA77  00000L0257B745  00000L02AB143D  00000L0257B75B  00000L0257B7B5  00000L0238AA6C  00000L02A8BEF0  00000L02A05D2E  00000L021568E7  00000L0257B4E2  00000L02167BFB  00000L02168011  00000L02408378  00000L021C49A9  00000L02A05C1D  00000L020E0A12  00000L020E07B4  00000LEE170A5A  00000L02167D8F  00000L0213E4AD  00000L0235ECCD  00000L0257B77F  00000L0257B7AA  00000L0238B0BA  00000L0246DDD7  00000L0271A4C2  00000L0219CA36  00000L0264CEC7  00000L02408516  00000L02A05C46  00000L02746C7B  00000L021502CC  00000L0215630D  00000L0212FEE4  00000L025472AD  00000L02711B16  00000L02A95BC9  00000L0257B4D2  00000L02577D71  00000L0257B765  00000L0213E6A7  00000L02AB15AF  00000LEE1FEA64  00000L0215005A  00000L029FDBD4  00000L02A9593A  00000L0247DB00  00000L02711B2C  00000L02711BF2  00000L02719FCA  00000L0215A334  00000L022E5DF7  00000L0238B2BB  00000L026F4B1B  00000L02135914  00000L0257B4C8  00000L025474AD  00000L020D750A  00000L0252CDBC  00000L020DB73E  00000L020DB4AC  00000LEE1BC514  00000L0257B7ED  00000L0240872F  00000L02141B3C  00000L026F4B58  00000L026D3F25  00000L029FDD6D  00000L0252CE40  00000L0257B44B  00000L021C4AB8  00000L0212FE4F  00000L0246DC77  00000L021560F3  00000L022D0734  00000L020E0943  00000L0257AAEE  00000L0257B859  00000L025471B8  00000L02167B0E  00000L022D4F6A  00000L0252CDD4  00000L02547026  00000L02577B3C  00000L02577C11  00000L020DB39C  00000L02546FDE  00000L02546FE3  00000L02577B30  00000L02577B72  00000L0202B786  00000L0202B84D  00000L0212FDE9  00000L02133F45  00000L02B52121  00000L0214FF0C  00000L0214FF1F  00000L02197781  00000L02197786  00000L0257B75C  00000L0247DABB  00000L0247DC1C  00000L0274AD52  00000L02A8BE62  00000L02577B79  00000L0246DB42  00000L0246DC08  00000L0246DDC4  00000L02547175  00000L0257B4B4  00000L026EA86D  00000L0257B47F  00000L026EA8B1  00000L0246DC90  00000L0247DA15  00000L0247DB85  00000L020E0B08  00000L0238B07F  00000L02156011  00000L0247D9E9  00000L0247DBD2  00000L02547044  00000L0254704F  00000L0254705F  00000L025474AA  00000L020E0A79  00000L020E0B70  00000L0238B1AB  00000L02A9D99C  00000L6100C4DB  00000L020D69C0  00000L020E5D23  00000L023FB2FB  00000L025B5438  00000L6100C329  00000L0219775A  00000L02577B04  00000L02744A62  00000L0202B53F  00000L0229E9BC  00000L02765EB2  00000L026EA7F1  00000L026EA8DD  00000L0246DD11  00000L0246DDB6  00000L0246DDCB  00000L02165369  00000L0219C8DC  00000L0215002D  00000L020E0C2A  00000L0202B84F  00000L02A05EFE  00000L021567FD  00000L0257B3F0  00000L020D6726  00000L02156345  00000L021653A5  00000L021653B3  00000L022D4EE4  00000L0238B1F5  00000L0238B236  00000L0257B4DA  00000L0257B7BD  00000L02227484  00000L0238A91A  00000L0238B30A  00000L0257B51D  00000L02165434  00000L025AD974  00000L027664F0  00000L0276656A  00000L02408479  00000L0202B5B9  00000L0247DB01  00000L0202B6A1  00000L02167E1E",
      "more_code_info": ""
    }
  ]
}