{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "La Ricamarie",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81582",
      "recalling_firm": "Stephanix",
      "address_1": "Zone Industrielle Du Bayon",
      "address_2": "10 rue Jean Moulin",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution in the states  of FL, IN, WV, TN, MO, WI, CA, FL, GA, PA",
      "recall_number": "Z-0557-2019",
      "product_description": "STEPHANIX D2RS Digital Dynamic Remote Systemhttp://cts.fda.gov/division-tracking/images/trash.png    Product The D2RS Digital Dynamic Remote System is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.",
      "product_quantity": "17 units",
      "reason_for_recall": "The firm has detected a potential risk using the command.  After releasing the command, the movement of the table may continue instead of stopping.",
      "recall_initiation_date": "20181102",
      "center_classification_date": "20181204",
      "termination_date": "20210725",
      "report_date": "20181212",
      "code_info": "DHJ008, DIG010, DIH012, DIH014, DIK022, DIK024, DIL025, DIL030, DJK051, DJK052, DKI032, DKI033, DKJ047, DKK051, DKK052, DLA001, DML062"
    }
  ]
}