{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79075",
      "recalling_firm": "Covidien LLC",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide including PR, Canada, China",
      "recall_number": "Z-0557-2018",
      "product_description": "Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 15cm (reinforced), Material CA15L2",
      "product_quantity": "N/A",
      "reason_for_recall": "The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.",
      "recall_initiation_date": "20170807",
      "center_classification_date": "20180208",
      "report_date": "20180214",
      "code_info": "UDI 10884521706583, Lot codes: S6MG005PX,S6MG008PX,S6MG012PX,S6MG019PX,S7AG001PX,S7AG006X,S7AG010X,S7AG013X,S7BG001X,S7BG003X,S7BG007X,S7BG013X,S7CG002X,S7CG005X,S7CG007X,S7CG013X,S7EG009X,S7FG003X,S7FG004X,S7FG007X"
    }
  ]
}