  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86846",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution including in the states of KY, CA, MI, OH, CT, MI, NY, SC, NC,IN, IL, TN, LA, CT, MT.",
      "recall_number": "Z-0554-2021",
      "product_description": "Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designed to provide fluoroscopic images of the patient during diagnostic surgical and interventional angiography and cardiology procedures.",
      "product_quantity": "7 units",
      "reason_for_recall": "The firm has identified a problem with the control software for the celling arm of the X-Ray System .  This software issue could cause the  ceiling arm of the device to move after the operator has released the  movement lever. This could result in possible collision or harm to the patient or system operator.",
      "recall_initiation_date": "20201118",
      "center_classification_date": "20201210",
      "report_date": "20201216",
      "code_info": "Serial Numbers: 40AD58C55001, 40AD58C4B001, 40AD58C24001, 40AD58522001, 40AD58C1C001,  101X968101, 091X955901"
    }
  ]
}