{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "345 E Main St",
      "reason_for_recall": "All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening",
      "address_2": "",
      "product_quantity": "188",
      "code_info": "Part Number 42-5399-071-01(left) Lot 62052718, 62121498 Part Number 42-5399-071-02 (right) Lot 62055082, 62121499",
      "center_classification_date": "20121218",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.",
      "state": "IN",
      "product_description": "Persona, The Personalized Knee System,   Tibial Sizing Plate, Cemented, Nonsterile, Size E Left; Size E Right    This device is used with the Persona Knee System for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities",
      "report_date": "20121226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-0552-2013",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "63690",
      "termination_date": "20140416",
      "more_code_info": "",
      "recall_initiation_date": "20121114",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}