  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elyria",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81071",
      "recalling_firm": "Invacare Corporation",
      "address_1": "1200 Taylor St",
      "address_2": "N/A",
      "postal_code": "44035-6248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Canada, Colombia, Germany, Guam, India, Northern Mariana Islands, Philippines and South Africa.",
      "recall_number": "Z-0549-2019",
      "product_description": "Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P    The intended use of a Non-AC-Powered patient lift is to lift and transfer a patient from one surface to another, as from a bed to a wheelchair. The device includes straps and a sling to support the patient.",
      "product_quantity": "73,703 units",
      "reason_for_recall": "Invacare has identified via customer complaints, the potential for the mounting bolt connecting the caster to the base frame of the patient lifts to become loose. Loose hardware can cause wear to the hardware and housings. If left unresolved, the caster may separate from the base of the lift causing the Lift to tilt and a patient fall. Failure does not occur right away but happens over time if the lift is not properly maintained.  Patient falls are inherently dangerous and may result in a range of injuries that are dependent on physical condition of the patient and the nature of the fall. When injuries occur, they are typically non-serious injuries such as bruising, lacerations, or abrasions. In rare cases, more serious injuries such as bone fracture, head trauma, or death can occur.  For caregivers using the lift, attempts to manually stabilize a lift if it becomes unstable could result in sprains, serious back injuries, or bone fractures.",
      "recall_initiation_date": "20181116",
      "center_classification_date": "20181203",
      "termination_date": "20200415",
      "report_date": "20181212",
      "code_info": "Serial numbers ranging between 09FSZ590001 to 16ESZ104075. Specifically, serial numbers with the following starting values (see below). Each consignee will be receiving an impacted units list, which will include full serial numbers for the units that they have purchased:     10ASZ 11ASZ 12ASZ 13ASZ 14ASZ 15ASZ 16ASZ   10BSZ 11BSZ 12BSZ 13BSZ 14BSZ 15BSZ 16BSZ   10CSZ 11CSZ 12CSZ 13CSZ 14CSZ 15CSZ 16CSZ   10DSZ 11DSZ 12DSZ 13DSZ 14DSZ 15DSZ 16DSZ   10ESSZ 11ESSZ 12ESSZ 13ESSZ 14ESSZ 15ESSZ 16ESSZ  09FSZ 10FSZ 11FSZ 12FSZ 13FSZ 14FSZ 15FSZ   09GSZ 10GSZ 11GSZ 12GSZ 13GSZ 14GSZ 15GSZ   09HSZ 10HSZ 11HSZ 12HSZ 13HSZ 14HSZ 15HSZ   09ISZ 10ISZ 11ISZ 12ISZ 13ISZ 14ISZ 15ISZ   09JSZ 10JSZ 11JSZ 12JSZ 13JSZ 14JSZ 15JSZ   09KSZ 10KSZ 11KSZ 12KSZ 13KSZ 14KSZ 15KSZ   09LSZ 10LSZ 11LSZ 12LSZ 13LSZ 14LSZ 15LSZ"
    }
  ]
}