{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72684",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Australia, Austria, Bangladesh, Belgium, Brazil, Cambodia, Canada, China, Egypt, France, Germany, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Lebanon, Libya, Lithuania, Malaysia, Martinique, Myanmar, Mexico, Mozambique, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Vietnam.",
      "recall_number": "Z-0548-2016",
      "product_description": "Ingenuity Core 128 Computed Tomography X-ray system",
      "product_quantity": "177 Units",
      "reason_for_recall": "Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.",
      "recall_initiation_date": "20151029",
      "center_classification_date": "20151224",
      "termination_date": "20180626",
      "report_date": "20151230",
      "code_info": "Model No. 728323; S/N: 30016, 32012, 32013, 32016, 32024, 32028, 32033, 32036, 32041, 32046, 32048, 32050, 32051, 32052, 32053, 32054, 32055, 32056, 32058, 32059, 32062, 32063, 32064, 32068, 32069, 32070, 32072, 32073, 32074, 32075, 32077, 32078, 32079, 32081, 32082, 32083, 32084, 32086, 32087, 32089, 32092, 32093, 32095, 32097, 32098, 32099, 32100, 32102, 32103, 32104, 32105, 32106, 32108, 32110, 32113, 32114, 32115, 32116, 32119, 32120, 32122, 32123, 32124, 32125, 32128, 32129, 32131, 32132, 32133, 32134, 32135, 32136, 32137, 32138, 32140, 32141, 32145, 32146, 320003, 320006, 320018, 320024, 320029, 320030, 320033, 320037, 320054, 320058, 320069, 320073, 320078, 320079, 320090, 320092, 320099, 320114, 320115, 320123, 320125, 320126, 320128, 320133, 320134, 320138, 320146, 320200, 320201, 320202, 320203, 320204, 320205, 320208, 320212, 320217, 320219, 320220, 320222, 320223, 320224, 320225, 320226, 320228, 320231, 320232, 320239, 320241, 320243, 320246, 320247, 336013, 336015, 336016, 336018, 336027, 336029, 336034, 336046, 336059, 336060, 336064, 336066, 336068, 336071, 336072, 336073, 336075, 336078, 336079, 336080, 336081, 336083, 336084, 336085, 336086, 336087, 336088, 336089, 336091, 336092, 336095, 336097, 336098, 336099, 336105, 336107, 336108, 336110, 336111, 336112, 336113, 336115, 336116, 336117, 336119, 336120, 336121, 336122, 336123, 336126, 336128, 336131, 336132, 336133, 336135, 336137, 336138, 336139, 336140, 336143, 336144, 336145, 336147, 336148, 336150 & 336151."
    }
  ]
}