  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81569",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0547-2019",
      "product_description": "Brilliance 10 Slice (Air), Model 728251    Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories",
      "product_quantity": "N/A",
      "reason_for_recall": "There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall",
      "recall_initiation_date": "20181018",
      "center_classification_date": "20181130",
      "termination_date": "20210803",
      "report_date": "20181212",
      "code_info": "System Serial Number:  2598  3008  3098  4001  4002  4003  4005  4006  4007  4008  4009  4010  4011  4012  4014  4015  4016  4020  4021  4022  4023  4026  4027  4028  4029  4030  4031  4032  4033  4034  4035  4039  4040  4044  4045  4047  4048  4049  4050  4052  4053  4054  4056  4058  4059  4061  4062  4063  4064  4065  4066  4067  30099  40004  40005  40006  40007  40008  40009  40010  40011  40012  40014  40015  40016  40018  40019  40020  40021  40023  40025  40028  40029  40031  40032  40034  62986554  4008/6086  4039/7128  728251_859-1766"
    }
  ]
}